Skip to content

OC-Auto FOBT-CHEK Negative Controls is the negative… recallZ-0538-2026 · Polymedco, Inc. / Polymedco Cancer Diagnostic Products, LLC · initiated Oct 6, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Polymedco, Inc. / Polymedco Cancer Diagnostic Products, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A labeling error was identified on the affected product. The expiration date on the kit box label and on the vial labels incorrectly stated the expiration date as 2026-01-16. The correct expiration date should be 2027-01-17.

FDA root cause category: Process control

What the company is doing

On October 6, 2025 URGENT PRODUCT RECALL: INCORRECT EXPIRATION DATE ON PRODUCT LABELING letters were sent to customers. ACTIONS REQUIRED BY YOU 1. Please trace the OCQN kits, lot 250805 in your inventory and at the customer sites listed in appendix one. If the kits have already been used no further action is required. There is no risk to patient care. 2. For customers with remaining stock there are two options:1) Manually correct the date on the packaging and vial labels to 2027-01-17 and add this notification to the kit box. OR 2) Request replacement stock with corrected labeling from Polymedco. ADDITIONAL INFORMATION Polymedco s inventory of Lot 250805 has been updated with new labels with the correct shelf life. All kits provided from September 25, 2025, will have the correct expiration date.

Recall details

Recall number
Z-0538-2026
Status
Open, Classified
Initiated by the company
Oct 6, 2025
Posted by the FDA
Nov 17, 2025
Recalling firm
Polymedco, Inc. / Polymedco Cancer Diagnostic Products, LLC
Quantity
403 units
Product code
JJY
FDA submissions
K924061
Distribution
USA: AK AL,AR,AZ, CA,CO,CT,DC,DE,FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, ,WI, WV, WY, Canada: AB, ON

Product description

OC-Auto FOBT-CHEK Negative Controls is the negative hemoglobin control materials designed to monitor FOBT (Fecal Occult Blood Test)

Code information (models, lots, serial numbers)

UDI: 00855574005520, Catalog Number(s): OCQN Lot Number(s): Lot 250805

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page