Brand Name: Catalyst+ recallZ-1092-2026 · C-Rad Positioning AB · initiated Oct 9, 2025
Class II Open, Classified All recalls posted in January 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from C-Rad Positioning AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
FDA root cause category: Under Investigation by firm
What the company is doing
On 10/14/2025, the firm email an Urgent Field Safety Notice to customers informing them that C-RAD has noted a potential issue affecting the cSRS mode of the Catalyst+ HD and Catalyst HD systems. All software version above c4D 5.3.2 are affected by this issue. The issue is that the stated precision of the Catalyst system for stereotactic radiosurgery (SRS) treatments cannot be guaranteed for all couch angles and system setups. In certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. This could in the worst-case result in erroneous dose delivery. This issue affects Catalyst+ HD and Catalyst HD systems, configured for cSRS and cMotion. Customers are instructed to: 1. Ensure that all the staff operating the Catalyst+ HD and Catalyst HD, and the c4D software application are informed of this information. 2. As an immediate action, affected users are advised to not use Catalyst+ HD and Catalyst HD in combination with cSRS and cMotion. C-RAD will release a new version of the software application c4D resolving the issue and/or update IFU/labelling. For questions please contact C-RAD Quality Department as [email protected]
Recall details
- Recall number
- Z-1092-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 9, 2025
- Posted by the FDA
- Jan 14, 2026
- Recalling firm
- C-Rad Positioning AB
- Quantity
- 189
- Product code
- IYE
- FDA submissions
- K200435
- Distribution
- The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
Product description
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable
Code information (models, lots, serial numbers)
Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module
Other devices under the same FDA submission
4 devices cleared under K200435 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Catalyst+ SP003-0001C-Rad Positioning AB · Accelerator, Linear, Medical1 recall0 reports
- Catalyst+ HD SP003-0002C-Rad Positioning AB · Accelerator, Linear, Medical1 recall0 reports
- Catalyst+ PT SP003-0008C-Rad Positioning AB · Accelerator, Linear, Medical1 recall0 reports
- Catalyst+ PT X4 SP003-0009C-Rad Positioning AB · Accelerator, Linear, Medical1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.