IV Administration Set recallZ-0670-2026 · B Braun Medical Inc · initiated Oct 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

FDA root cause category: Process control

What the company is doing

B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."

Recall details

Recall number
Z-0670-2026
Status
Open, Classified
Initiated by the company
Oct 29, 2025
Posted by the FDA
Nov 26, 2025
Recalling firm
B Braun Medical Inc
Quantity
20,694,734 units
Product code
FPA
FDA submissions
K170595
Distribution
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Product description

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DR W/2 SAFEDAY; Catalog Number: 352636. 2. IV ADMIN SET 15 DR W/2 SAFEDAY 84 IN.; Catalog Number: 352639. 3. IV ADMIN SET 15 DROP W/3 SAFEDAY; Catalog Number: 352640. 4. IV ADMIN 15D W/2 SAFEDAY & 0.2 FILTER; Catalog Number: 352643. 5. IV ADMIN SET, 105 IN.; Catalog Number: 477003. 6. IV ADMIN SET 15D, 4 CARESTIE LAD, 145 IN; Catalog Number: 477006. 7. IV ADMIN SET, 121 IN.; Catalog Number: 477008. 8. IV ADMIN SET 15D, CARESTIE LAD, 90 IN; Catalog Number: 477012. 9. IV ADMIN SET, 2 CARESTIE LAD, 15D, 117IN; Catalog Number: 477019. 10. IV ADMIN SET, 15D, 5CARESITE LAD, 136 IN; Catalog Number: 477021. 11. IV ADMIN SET 114 IN.; Catalog Number: 477022. 12. IV ADMIN SET 5 CARESITE LADS, 126 IN.; Catalog Number: 477023. 13. IV SET 15DR 3 SAFEDAY, 106 IN.; Catalog Number: 477053. 14. IV ADMIN SET, 15DR, CARESITE LAD, 121 IN; Catalog Number: 490513. 15. IV ADMIN SET, 15D, 120 IN.; Catalog Number: 490521. 16. IV ADMIN SET 15D 2 CARESITE LADS, 112IN.; Catalog Number: 490527. 17. IV ADMIN SET 15 DRP CARESITE SAFEPORT; Catalog Number: 490529. 18. IV ADMIN SET, 15D, 139 IN.; Catalog Number: 490535. 19. IV ADMIN SET 15D, CARESITE LAD, 122IN.; Catalog Number: 490543. 20. IV ADMIN SET 15D 3 CARESITE LADS, 150IN.; Catalog Number: 490546. 21. IV ADMIN SET, 125 IN.; Catalog Number: 490547. 22. IV ADMIN SET 15D, 5 CARESITE INJ, 150in.; Catalog Number: 490552. 23. IV ADMIN SET, 120IN.; Catalog Number: 490553. 24. IV ADMIN SETW/ B/C VLV 3 CARESITE 134IN.; Catalog Number: 490554. 25. IV ADMIN SET 15DR w/3 B/C VLV, 124 IN.; Catalog Number: 490555. 26. IV ADMIN SET 15D w/B/C VLV, 139 IN.; Catalog Number: 490558. 27. IV ADMINISTRATION SET, 117 IN.; Catalog Number: 490559. 28. IV ADMINISTRATION SET, 108 IN.; Catalog Number: 490560. 29. IV ADMINISTRATION SET, 134 IN.; Catalog Number: 490561. 30. IV ADMIN SET 15D w/B/C VLV, 140 IN.; Catalog Number: 490562. 31. IV ADMIN SET, 111 IN.; Catalog Number: 490563. 32. IV ADMIN SET 15D 2 CARESITE LADS, 122 IN; Catalog Number: 490569. 33. IV ADMIN SET 15D STOPCOCK, DUAL MANIFOLD; Catalog Number: 490573. 34. IV ADMIN SET 15D w/B/C VLV, 110 IN.; Catalog Number: 490574. 35. IV ADMIN 15D 5 CARESITE LADS STPCK 120in; Catalog Number: 490575. 36. IV ADMIN SET 15D 5 CARESITE LADS, 140 IN; Catalog Number: 490578. 37. IV ADMIN SET W/15D, 4 CARESITE LADS; Catalog Number: 490581. 38. IV ADMIN 15D CARESITE LADS, STPCK 134 IN; Catalog Number: 490582. 39. IV ADMIN SET, 15D, 108 IN.; Catalog Number: 490583. 40. IV ADMIN SET, 10 DROP, 130 IN.; Catalog Number: 490588. 41. IV ADMIN SET, 132 IN.; Catalog Number: 490589. 42. IV ADMIN SET, 15DR, 105 IN.; Catalog Number: 490600. 43. IV ADMIN SET,3 CARESITE,3 HF SPCK, 129IN; Catalog Number: 490607. 44. IV SET 15DR 3CARESITE 2STOPCOCK, 126IN; Catalog Number: 490608. 45. IV ADMIN SET 15D 2 CARESITE LADS, 138 IN; Catalog Number: 490615. 46. IV SET 15DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490618. 47. IV SET 60DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490619. 48. IV ADMIN SET 15 DRPS 4W STOPCOCK W/ C&S; Catalog Number: 490644. 49. IV ADMIN SET 15 DRPS ULTRAP. ZERO 90.6IN; Catalog Number: 490657. 50. IV ADMIN SET 15 DRPS STOPCOCK MALE 117IN; Catalog Number: 490658.

Code information (models, lots, serial numbers)

Show the full code information (7k characters)

1. Catalog Number: 352636; Primary UDI-DI: 04046955594597; Unit of Dose UDI-DI: 04046955594580; Expiration Date: 36mths from Start Date. 2. Catalog Number: 352639; Primary UDI-DI: 04046955594672; Unit of Dose UDI-DI: 04046955594665; Expiration Date: 36mths from Start Date. 3. Catalog Number: 352640; Primary UDI-DI: 04046955594696; Unit of Dose UDI-DI: 04046955594689; Expiration Date: 36mths from Start Date. 4. Catalog Number: 352643; Primary UDI-DI: 04046955594757; Unit of Dose UDI-DI: 04046955594740; Expiration Date: 36mths from Start Date. 5. Catalog Number: 477003; Primary UDI-DI: 04046955202836; Unit of Dose UDI-DI: 04046955202829; Expiration Date: Earliest Exp of Comp or 36mths. 6. Catalog Number: 477006; Primary UDI-DI: 04046955202751; Unit of Dose UDI-DI: 04046955202744; Expiration Date: Earliest Exp of Comp or 36mths. 7. Catalog Number: 477008; Primary UDI-DI: 04046955202850; Unit of Dose UDI-DI: 04046955202843; Expiration Date: Earliest Exp of Comp or 36mths. 8. Catalog Number: 477012; Primary UDI-DI: 04046955202874; Unit of Dose UDI-DI: 04046955202867; Expiration Date: Earliest Exp of Comp or 36mths. 9. Catalog Number: 477019; Primary UDI-DI: 04046955203147; Unit of Dose UDI-DI: 04046955203130; Expiration Date: Earliest Exp of Comp or 36mths. 10. Catalog Number: 477021; Primary UDI-DI: 04046955269983; Unit of Dose UDI-DI: 04046955269976; Expiration Date: Earliest Exp of Comp or 36mths. 11. Catalog Number: 477022; Primary UDI-DI: 04046955270002; Unit of Dose UDI-DI: 04046955269990; Expiration Date: Earliest Exp of Comp or 36mths. 12. Catalog Number: 477023; Primary UDI-DI: 04046955270026; Unit of Dose UDI-DI: 04046955270019; Expiration Date: Earliest Exp of Comp or 36mths. 13. Catalog Number: 477053; Primary UDI-DI: 04046955672431; Unit of Dose UDI-DI: 04046955672424; Expiration Date: Earliest Exp of Comp or 36mths. 14. Catalog Number: 490513; Primary UDI-DI: 04046955137237; Unit of Dose UDI-DI: 04046955137220; Expiration Date: Earliest Exp of Comp or 36mths. 15. Catalog Number: 490521; Primary UDI-DI: 04046955185252; Unit of Dose UDI-DI: 04046955186563; Expiration Date: Earliest Exp of Comp or 24mths. 16. Catalog Number: 490527; Primary UDI-DI: 04046955203222; Unit of Dose UDI-DI: 04046955203215; Expiration Date: Earliest Exp of Comp or 36mths. 17. Catalog Number: 490529; Primary UDI-DI: 04046955203307; Unit of Dose UDI-DI: 04046955203291; Expiration Date: Earliest Exp of Comp or 24mths. 18. Catalog Number: 490535; Primary UDI-DI: 04046955203406; Unit of Dose UDI-DI: 04046955203390; Expiration Date: Earliest Exp of Comp or 36mths. 19. Catalog Number: 490543; Primary UDI-DI: 04046955270200; Unit of Dose UDI-DI: 04046955270194; Expiration Date: Earliest Exp of Comp or 36mths. 20. Catalog Number: 490546; Primary UDI-DI: 04046955270262; Unit of Dose UDI-DI: 04046955270255; Expiration Date: Earliest Exp of Comp or 36mths. 21. Catalog Number: 490547; Primary UDI-DI: 04046955270286; Unit of Dose UDI-DI: 04046955270279; Expiration Date: Earliest Exp of Comp or 36mths. 22. Catalog Number: 490552; Primary UDI-DI: 04046955275342; Unit of Dose UDI-DI: 04046955275335; Expiration Date: Earliest Exp of Comp or 36mths. 23. Catalog Number: 490553; Primary UDI-DI: 04046955275366; Unit of Dose UDI-DI: 04046955275359; Expiration Date: Earliest Exp of Comp or 36mths. 24. Catalog Number: 490554; Primary UDI-DI: 04046955275380; Unit of Dose UDI-DI: 04046955275373; Expiration Date: Earliest Exp of Comp or 36mths. 25. Catalog Number: 490555; Primary UDI-DI: 04046955275403; Unit of Dose UDI-DI: 04046955275397; Expiration Date: Earliest Exp of Comp or 36mths. 26. Catalog Number: 490558; Primary UDI-DI: 04046955275465; Unit of Dose UDI-DI: 04046955275458; Expiration Date: Earliest Exp of Comp or 36mths. 27. Catalog Number: 490559; Primary UDI-DI: 04046955275489; Unit of Dose UDI-DI: 04046955275472; Expiration Date: Earliest Exp of Comp or 36mths. 28. Catalog Number: 490560; Primary UDI-DI: 04046955275502; Unit of Dose UDI-DI: 04046955275496; Expiration Date: Earliest Exp of Comp or 36mths. 29. Catalog Number: 490561; Primary UDI-DI: 04046955275526; Unit of Dose UDI-DI: 04046955275519; Expiration Date: Earliest Exp of Comp or 36mths. 30. Catalog Number: 490562; Primary UDI-DI: 04046955275540; Unit of Dose UDI-DI: 04046955275533; Expiration Date: Earliest Exp of Comp or 36mths. 31. Catalog Number: 490563; Primary UDI-DI: 04046955275564; Unit of Dose UDI-DI: 04046955275557; Expiration Date: Earliest Exp of Comp or 36mths. 32. Catalog Number: 490569; Primary UDI-DI: 04046955275687; Unit of Dose UDI-DI: 04046955275670; Expiration Date: Earliest Exp of Comp or 36mths. 33. Catalog Number: 490573; Primary UDI-DI: 04046955287451; Unit of Dose UDI-DI: 04046955287444; Expiration Date: Earliest Exp of Comp or 24mths. 34. Catalog Number: 490574; Primary UDI-DI: 04046955287475; Unit of Dose UDI-DI: 04046955287468; Expiration Date: Earliest Exp of Comp or 24mths. 35. Catalog Number: 490575; Primary UDI-DI: 04046955287499; Unit of Dose UDI-DI: 04046955287482; Expiration Date: Earliest Exp of Comp or 36mths. 36. Catalog Number: 490578; Primary UDI-DI: 04046955287550; Unit of Dose UDI-DI: 04046955287543; Expiration Date: Earliest Exp of Comp or 36mths. 37. Catalog Number: 490581; Primary UDI-DI: 04046955287611; Unit of Dose UDI-DI: 04046955287604; Expiration Date: Earliest Exp of Comp or 24mths. 38. Catalog Number: 490582; Primary UDI-DI: 04046955287635; Unit of Dose UDI-DI: 04046955287628; Expiration Date: Earliest Exp of Comp or 36mths. 39. Catalog Number: 490583; Primary UDI-DI: 04046955287659; Unit of Dose UDI-DI: 04046955287642; Expiration Date: Earliest Exp of Comp or 60mths. 40. Catalog Number: 490588; Primary UDI-DI: 04046955287758; Unit of Dose UDI-DI: 04046955287741; Expiration Date: Earliest Exp of Comp or 36mths. 41. Catalog Number: 490589; Primary UDI-DI: 04046955287772; Unit of Dose UDI-DI: 04046955287765; Expiration Date: Earliest Exp of Comp or 36mths. 42. Catalog Number: 490600; Primary UDI-DI: 04046955352777; Unit of Dose UDI-DI: 04046955352760; Expiration Date: Earliest Exp of Comp or 24mths. 43. Catalog Number: 490607; Primary UDI-DI: 04046955352913; Unit of Dose UDI-DI: 04046955352906; Expiration Date: Earliest Exp of Comp or 36mths. 44. Catalog Number: 490608; Primary UDI-DI: 04046955352937; Unit of Dose UDI-DI: 04046955352920; Expiration Date: Earliest Exp of Comp or 36mths. 45. Catalog Number: 490615; Primary UDI-DI: 04046955664672; Unit of Dose UDI-DI: 04046955664665; Expiration Date: Earliest Exp of Comp or 36mths. 46. Catalog Number: 490618; Primary UDI-DI: 04046955664733; Unit of Dose UDI-DI: 04046955664726; Expiration Date: Earliest Exp of Comp or 24mths. 47. Catalog Number: 490619; Primary UDI-DI: 04046955664757; Unit of Dose UDI-DI: 04046955664740; Expiration Date: Earliest Exp of Comp or 24mths. 48. Catalog Number: 490644; Primary UDI-DI: 04046955765003; Unit of Dose UDI-DI: 04046955764990; Expiration Date: Earliest Exp of Comp or 36mths. 49. Catalog Number: 490657; Primary UDI-DI: 04046955836420; Unit of Dose UDI-DI: 04046955836413; Expiration Date: Earliest Exp of Comp or 36mths. 50. Catalog Number: 490658; Primary UDI-DI: 04046955836697; Unit of Dose UDI-DI: 04046955836680; Expiration Date: Earliest Exp of Comp or 36mths.

Devices named in this recall

50 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97816).