IV Administration Set recallZ-0691-2026 · B Braun Medical Inc · initiated Oct 29, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA root cause category: Process control
What the company is doing
B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."
Recall details
- Recall number
- Z-0691-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 29, 2025
- Posted by the FDA
- Nov 26, 2025
- Recalling firm
- B Braun Medical Inc
- Quantity
- 24,072 units
- Product code
- FPA
- FDA submissions
- K083723, K790062
- Distribution
- Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Product description
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 2 CARESITES, 125 IN.; Catalog Number: 490408.
Code information (models, lots, serial numbers)
1. Catalog Number: 490408; Primary UDI-DI: 04046964773266; Unit of Dose UDI-DI: 04046964773259; Expiration Date: Earliest Exp of Comp or 24mths.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IV Administration Set 490408B. Braun Medical Inc. · Set, administration, intravascular1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97816).
- Z-0603-2026 · Open, ClassifiedADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0604-2026 · Open, ClassifiedADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0605-2026 · Open, ClassifiedADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0606-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0607-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0608-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0609-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0610-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0611-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0612-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0613-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0614-2026 · Open, ClassifiedAnesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump…
- Z-0615-2026 · Open, ClassifiedB.BRAUN utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled…
- Z-0616-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0617-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0618-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0619-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0620-2026 · Open, ClassifiedBlood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Bas…
- Z-0621-2026 · Open, ClassifiedBurette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Lab…
- Z-0622-2026 · Open, ClassifiedBurette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Lab…
- Z-0623-2026 · Open, ClassifiedCARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0624-2026 · Open, ClassifiedCaresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0625-2026 · Open, ClassifiedCaresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0626-2026 · Open, ClassifiedCaresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labele…
- Z-0627-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0628-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0629-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0630-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0631-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
- Z-0632-2026 · Open, ClassifiedExtension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. L…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page