FlexLab (FLX) recallZ-0595-2026 · Inpeco S.A. · initiated Oct 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Inpeco S.A. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
FDA root cause category: Device Design
What the company is doing
On October 30, 2025, "Urgent: Medical Device Correction" letters were sent to customers. Action to be taken: To avoid the occurrence of the described issue Inpeco recommends removing all the secondary sample tubes in the printer tubes buffer (red arrow) each time the printer cover (blue arrow) is opened for error recoveries and for preventive maintenance procedures. Your service provider will contact you to schedule the service visit to implement the corrective actions which include the installation of a new sensor, the software and firmware upgrade, and the operations manual update. Transmission of this Customer Letter: Until the service visit, please maintain awareness on this notice and apply the action recommended above. Please transfer this notice to whom it might concern. Please complete and return the Customer Letter Receipt Confirmation and Implementation Check form attached to this letter within 30 days directly to the email address specified in the email communication. If you have any questions, contact the Regulatory Affairs Department via email: [email protected].
Recall details
- Recall number
- Z-0595-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 30, 2025
- Posted by the FDA
- Nov 26, 2025
- Recalling firm
- Inpeco S.A.
- Quantity
- 1 unit (OUS only)
- Product code
- CEM
- FDA submissions
- K121012
- Distribution
- US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
Product description
FlexLab (FLX); Version: FLX-217-10;
Code information (models, lots, serial numbers)
UDI-DI: 07640172340004 Serial number: FLX.0226;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- FlexLab FLXInpeco SA · Electrode, Ion Specific, Potassium8 recalls17 reports
Other devices under the same FDA submission
3 devices cleared under K121012 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Aptio Automation AP2Inpeco SA · Electrode, Ion Specific, Potassium5 recalls6 deaths16 reports
- Accelerator a3600 ACPInpeco SA · Electrode, Ion Specific, Potassium5 recalls13 reports
- FlexLab X FXXInpeco SA · Electrode, Ion Specific, Potassium2 recalls2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97887).
- Z-0596-2026 · Open, ClassifiedFlexLab (FLX); Version: FLX-217-40;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page