Skip to content

Phantom Fibula Nail System recallZ-0871-2026 · Paragon 28, Inc. · initiated Oct 13, 2025

Class II Open, Classified All recalls posted in December 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Paragon 28, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.

FDA root cause category: Device Design

What the company is doing

On 10/13/2025, recall notices were emailed to distributors who were asked to do the following: 1) Return affected devices to the firm using return label. 2) Ensure that in instances where you have distributed this affected product, that you forward this Recall Notification to your customers. 3) Replacements are available by contacting [email protected] Complaints can be directed to [email protected]

Recall details

Recall number
Z-0871-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Oct 13, 2025
Posted by the FDA
Dec 2, 2025
Recalling firm
Paragon 28, Inc.
Quantity
20
Product code
HSB
FDA submissions
K234128
Distribution
US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.

Product description

Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S

Code information (models, lots, serial numbers)

UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K234128 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates