Skip to content

Ct 5300 recallZ-0882-2026 · Philips North America Llc · initiated Oct 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.

FDA root cause category: Under Investigation by firm

What the company is doing

On October 23, 2025 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Please continue to use your system in accordance with its intended use and do not stay in the scan room when scanning. 2. Circulate this URGENT Medical Device Correction letter to all users of this device so that they are aware of the issue. 3. Place this URGENT Medical Device Correction with your system documentation and ensure the letter is in a place likely to be seen/viewed. 4. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction letter, understanding of the issue, and required actions to be taken. Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the tube heat exchanger screws and correct the loosen screw if necessary (reference FCO72800842). Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact your local Philips Customer Care Solutions center: For North America, contact the Customer Care Solutions Center {1-800-722-9377, 8AM - 8PM EST, Monday-Friday).

Recall details

Recall number
Z-0882-2026
Status
Open, Classified
Initiated by the company
Oct 29, 2025
Posted by the FDA
Dec 3, 2025
Recalling firm
Philips North America Llc
Quantity
6 units (OUS only)
Product code
JAK
FDA submissions
K232491
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Ethiopia, France, FrenchPolynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua Nw Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Serbia, Sint Maarten (D, Slovakia, South Africa, South Korea, Spain, St.Pier,Miquel., Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.

Product description

CT 5300; Software Version Number: 4.5, 5.0, 5.1;

Code information (models, lots, serial numbers)

1) Model Number: 728285; UDI-DI: 00884838113237; Serial Numbers: 35071, 800007, 800002, 800001, 800004, 800003;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97919).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page