Skip to content

IntelliVue MP5. Product Number: M8105A recallZ-0850-2026 · Philips North America Llc · initiated Oct 31, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential issue where the IntelliVue monitors did not alarm.

FDA root cause category: Use error

What the company is doing

An URGENT Medical Device Correction notice, dated 10/31/25, was mailed to customers and distributors. Distributors were notified of this correction and instructed to forward the provided recall notification to all consignees. Additionally, distributors are asked to include their contact information for consignees in the provided response form and perform a good faith effort for each customer to obtain a completed response form. Consignees are instructed to notify all users of this recall notification and forward the notice to all who need to be aware within their organization or where affected devices were potentially transferred. Consignees are to re-evaluate and reconfigure device configuration per the provided instructions. Once recall actions have been completed, consignees are to complete the provided response form either via QR code, by email to [email protected], or by fax to 877-499-7223. Philips is planning to develop a software update which will enforce the additional confirmation step when the user enables infinite Alarms Off for IntelliVue MP5 models only. Consignees with any questions can contact Philips at 800-722-9377.

Recall details

Recall number
Z-0850-2026
Status
Open, Classified
Initiated by the company
Oct 31, 2025
Posted by the FDA
Dec 2, 2025
Recalling firm
Philips North America Llc
Quantity
1,913,441 units
Product code
DSI
FDA submissions
K150310
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Botswana, Brazil, Brunei, Bulgaria, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Equat Guinea, Estonia, Ethiopia, Faroe Islands, Finland, French Polynesia, France, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea Rep, Kuwait, Lao, Latvia, Lebanon, Lesotho, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Ant, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, , R¿union, Romania, Russia, Saint Vincent, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, St. Pierre Mqln, Suriname, Sweden, Switzerland, Syrian Arab Rep, Taiwan, Tajikistan, Thailand, Trinidad Tobago, Tunisia, T¿rkiye, Turkmenistan, Uganda, Ukraine, United Kingdom, United States, Uruguay, Utd Arab Emir, Uzbekistan, Venezuela, Viet Nam, Virgin Isl (Br), Virgin Isl (US), White Russia, Yemen, Zambia, Zimbabwe.

Product description

IntelliVue MP5. Product Number: M8105A.

Code information (models, lots, serial numbers)

Product Number: M8105A; UDI-DI: 00884838000230; All Serial No.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97917).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page