ErgoStar CM 45 recallZ-0591-2026 · Draeger, Inc. · initiated Nov 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Multiple complaints were reported in which cracks formed in the hose of specific catheter mounts.
FDA root cause category: Under Investigation by firm
What the company is doing
On November 10, 2025, Urgent Medical Device Recall letters were sent to customers. Necessary Action: Identify if you have any affected product. Do not use any unused ErgoStar CM 40/CM 45/CM 55/CM 60 in your stock. Unused stock should be removed and returned to Draeger. Draeger recommends the alternative product MP01850 ErgoStar CM 50 as a suitable replacement. Please ensure that all potential users in your facility are made aware of this Urgent Medical Device Recall Notice and the information contained herein. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification.
Recall details
- Recall number
- Z-0591-2026
- Status
- Open, Classified
- Initiated by the company
- Nov 10, 2025
- Posted by the FDA
- Dec 8, 2025
- Recalling firm
- Draeger, Inc.
- Quantity
- 191470 units
- Product code
- BZA
- Distribution
- Worldwide distribution. US states of CA, FL, IL, NY, OH, OK, PA, SC, TX, and WI; Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Czech Republic, Ecuador, Finland, France, Germany, Hungary, Indonesia, Italy, Japan, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkiye, Ukraine, and United Kingdom.
Product description
ErgoStar CM 45, Model/Catalog Number: MP01845, Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.
Code information (models, lots, serial numbers)
Model/Catalog Number: MP01845; UDI Number: 04048675420729; Lot numbers: All lot numbers;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ErgoStar CM 45 MP01845Drägerwerk AG & Co. KGaA · Connector, Airway (Extension)1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 97923).
- Z-0590-2026 · Open, ClassifiedErgoStar CM 40 Model/Catalog Number: MP01840 Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube…
- Z-0592-2026 · Open, ClassifiedErgoStar CM 55, Model/Catalog Number: MP01855, Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal …
- Z-0593-2026 · Open, ClassifiedErgoStar CM 60, Model/Catalog Number: MP01860, Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page