Skip to content

Hemochron Whole Blood Microcoagulation System recallZ-0880-2026 · Accriva Diagnostics, Inc. · initiated Nov 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Accriva Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Test cuvettes, used with an assay for monitoring heparin anticoagulation, may have a foreign object that has sharp edges and could pose a risk of injury if not detected prior to handling.

FDA root cause category: Packaging process control

What the company is doing

On 11/5/2025, recall notices were mailed to Medical Directors, Lab Directors, Clinical Laboratory Managers, or Point of Care Coordinators who were asked to do the following: 1) Forward this communication to those within your organization who need to be aware and verify they understand the procedure and can complete the enclosed forms. 2) Forward the enclosed customer notification to all other organizations where affected devices may have been transferred. 3) Please check your inventory for affected devices. If you find any, please refrain from opening any pouches or using the product. Destroy all impacted cuvette pouches and/or boxes of the affected devices. 4) Complete and return the Customer Acknowledgment Form via email to [email protected] 5) If you have questions regarding replacement product, please contact the firm's Customer Service at 1-800-955-9525 or e-mail to [email protected]. 6) For any technical related questions, please contact the firm's Technical Services at 1-800-678-0710 Option No. 1 or email to: [email protected]

Recall details

Recall number
Z-0880-2026
Status
Open, Classified
Initiated by the company
Nov 5, 2025
Posted by the FDA
Dec 3, 2025
Recalling firm
Accriva Diagnostics, Inc.
Quantity
20250
Product code
JBP
FDA submissions
K960749
Distribution
Worldwide - US Nationwide distribution including in the states of OH, IL, NY, GA, CA, IN, NJ, TX, KS, NE, FL, SC, MO, WV, WA, MI, CO, TN, PA, UT, LA, MT, NC, KY, TX and the countries of CA.

Product description

Hemochron Activated Clotting Time Low-Range Test Cuvette, REF: ACT-LR, to be used with the Hemochron Signature ELITE Whole Blood Microcoagulation System

Code information (models, lots, serial numbers)

UDI-DI: 10711234103081. Lot: E5JLR132. Expiration: 08-31-2026

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates