Skip to content

The Stryker CranialMask Tracker is a single-use device… recallZ-0963-2026 · Howmedica Osteonics Corp. · initiated Nov 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

FDA root cause category: Under Investigation by firm

What the company is doing

On November 12, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list in Table 1. Impacted product should be discarded at the facility. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form with quantity of devices discarded and email to [email protected]. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions.

Recall details

Recall number
Z-0963-2026
Status
Open, Classified
Initiated by the company
Nov 12, 2025
Posted by the FDA
Dec 19, 2025
Recalling firm
Howmedica Osteonics Corp.
Quantity
320 units
Product code
HAW
FDA submissions
K212194
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.

Product description

The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.

Code information (models, lots, serial numbers)

Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page