Skip to content

Cardiosave Rescue. Intra-Aortic Balloon Pump system recallZ-0917-2026 · Datascope Corp. · initiated Nov 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.

FDA root cause category: Component design/selection

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION NOTICE, dated October 2025, was sent via FEDEX to consignees. The notification notifies consignees of four addendum updates: 1) NVRAM Preventive Maintenance (New FSCA - reflected in this recall entry), 2) Vibration and Shock Table Specifications (New FSCA), 3) Helium O-Ring Preventive Maintenance (FSCA 2249723-01/24/2023-004-C), and 4) Battery Runtime Specifications (FSCA 2249723-09/10/2021-001-R). The notification instructs consignees regarding the NVRAM Preventive Maintenance to review the provided IFU addendum, update local preventive maintenance procedures as such, and ensure all NVRAM batteries are within their life expectancy. If a NVRAM is greater than 8 years old, until it can be replaced, it is advised a back-up Cardiosave IABP be available to provide therapy. Consignees with any questions can contact Datascope/Getinge Customer Support at 1-88-943-8872 (options 4, 2, 1) Monday through Friday, from 8:00 AM EST and 6:00 PM EST.

Recall details

Recall number
Z-0917-2026
Status
Open, Classified
Initiated by the company
Nov 4, 2025
Posted by the FDA
Dec 11, 2025
Recalling firm
Datascope Corp.
Quantity
47 units
Product code
DSP
FDA submissions
K163542
Distribution
Domestic: Nationwide Distribution; Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine Territory, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen, Zambia.

Product description

Cardiosave Rescue. Intra-Aortic Balloon Pump system.

Code information (models, lots, serial numbers)

Model No. (UDI): 0998-00-0800-83 (10607567108407); ALL SERIAL NO.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.