Draeger Vapor 2000. Unheated, calibrated anesthetic… recallZ-0933-2026 · Draeger, Inc. · initiated Nov 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A certain component of affected devices was not delivered within specification and contained impurities.

FDA root cause category: Under Investigation by firm

What the company is doing

Consignees were mailed an Urgent Medical Device Recall notification dated November 2025. The notification recommends that consignees do not use affected vaporizers and return the provided reply card to confirm this. A Draeger representative will contact consignees to arrange replacement of affected vaporizers. If affected vaporizers have been placed into clinical use, Draeger asks that consignees clean the breathing system of the anesthesia device according to the IFU and use of inspiratory filters is recommended. Consignees are to share the recall notification with all users of affected devices to ensure awareness and forward the notice if product was further distributed. Questions regarding operation of devices are to be directed to Drager Service Technical Support at 1-800-437-2437 (Option 2, 2, 2). Questions about the recall notice are to be directed to Michael Kelhart via email at [email protected] or phone at 267-664-1131.

Recall details

Recall number
Z-0933-2026
Status
Open, Classified
Initiated by the company
Nov 24, 2025
Posted by the FDA
Dec 23, 2025
Recalling firm
Draeger, Inc.
Quantity
431 units
Product code
CAD
FDA submissions
K971923
Distribution
Worldwide - US Nationwide distribution in the states of AK, CA, FL, IA, KS, MD, ME, MN, NY, OH, PA, RI, SD, TN, TX, UT, W and the countries of Argentina, Australia, Belgium, Brazil, Cambodia, China, Columbia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Hungary, Indonesia, Israel, Japan, Kenya, Mexico, Nepal, Norway, Panama, Peru, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates, Vietnam.

Product description

Draeger Vapor 2000. Unheated, calibrated anesthetic vaporizer.

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

Model No. M35054; UDI-DI: 04048675228059; Serial No. ASUK-0421,ASUK-0422,ASUK-0423,ASUK-0424,ASUK-0425,ASUK-0426,ASUK-0602,ASUK-0603, ASUK-0705,ASUK-0706,ASUK-0707,ASUK-0708,ASUK-0721,ASUK-0722,ASUK-0723,ASUK-0724, ASUK-0725,ASUK-0726,ASUK-0727,ASUK-0728,ASUK-0729,ASUK-0730,ASUK-0731,ASUK-0732, ASUK-0733,ASUK-0734,ASUK-0735,ASUK-0736,ASUK-0737,ASUK-0738,ASUK-0739,ASUK-0740, ASUK-0749,ASUK-0761,ASUK-0762,ASUK-0763,ASUK-0764,ASUK-0765,ASUK-0766,ASUK-0767, ASUK-0768,ASUK-0769,ASUK-0770,ASUK-0835,ASUK-0836,ASUK-0837,ASUK-0972,ASUK-0973, ASUK-0974,ASUK-0975,ASUK-0993,ASUK-0994,ASUK-0995,ASUK-0996,ASUK-0997,ASUK-0998, ASUK-0999,ASUK-1000,ASUK-1011,ASUK-1015,ASUK-1016,ASUK-1017,ASUK-1018,ASUK-1221, ASUK-1222,ASUK-1232,ASUK-1233,ASUK-1234,ASUK-1235,ASUK-1236,ASUK-1237,ASUK-1239, ASUK-1240,ASUK-1246,ASUK-1306,ASUK-1307,ASUK-1308,ASUK-1309,ASUK-1539,ASUK-1540, ASUK-0428,ASUK-0429,ASUK-0430,ASUK-0431,ASUK-0432,ASUK-0433,ASUK-0434,ASUK-0435, ASUK-0436,ASUK-0437,ASUK-0438,ASUK-0439,ASUK-0440,ASUK-0441,ASUK-0442,ASUK-0443, ASUK-0444,ASUK-0445,ASUK-0446,ASUK-0447,ASUK-0448,ASUK-0449,ASUK-0450,ASUK-0,63, ASUK-0564,ASUK-0565,ASUK-0566,ASUK-0567,ASUK-0568,ASUK-0569,ASUK-0570,ASUK-0571, ASUK-0572,ASUK-0573,ASUK-0574,ASUK-0575,ASUK-0576,ASUK-0577,ASUK-0578,ASUK-0579, ASUK-0580,ASUK-0581,ASUK-0582,ASUK-0583,ASUK-0584,ASUK-0585,ASUK-0601,ASUK-0604, ASUK-0605,ASUK-0606,ASUK-0607,ASUK-0608,ASUK-0609,ASUK-0610,ASUK-0631,ASUK-0632, ASUK-0633,ASUK-0634,ASUK-0635,ASUK-0636,ASUK-0637,ASUK-0638,ASUK-0639,ASUK-0640, ASUK-0641,ASUK-0642,ASUK-0661,ASUK-0662,ASUK-0663,ASUK-0664,ASUK-0667,ASUK-0668, ASUK-0681,ASUK-0682,ASUK-0691,ASUK-0692,ASUK-0693,ASUK-0694,ASUK-0695,ASUK-0696, ASUK-0697,ASUK-0698,ASUK-0699,ASUK-0700,ASUK-0711,ASUK-0712,ASUK-0713,ASUK-0714, ASUK-0715,ASUK-0716,ASUK-0717,ASUK-0718,ASUK-0719,ASUK-0720,ASUK-0741,ASUK-0742, ASUK-0743,ASUK-0744,ASUK-0746,ASUK-0747,ASUK-0748,ASUK-0750,ASUK-0751,ASUK-0752, ASUK-0753,ASUK-0754,ASUK-0755,ASUK-0756,ASUK-0757,ASUK-0758,ASUK-0759,ASUK-0760, ASUK-0778,ASUK-0779,ASUK-0780,ASUK-0781,ASUK-0782,ASUK-0783,ASUK-0784,ASUK-0785, ASUK-0786,ASUK-0787,ASUK-0788,ASUK-0789,ASUK-0790,ASUK-0791,ASUK-0792,ASUK-0793, ASUK-0794,ASUK-0795,ASUK-0796,ASUK-0797,ASUK-0798,ASUK-0799,ASUK-0800,ASUK-0801, ASUK-0802,ASUK-0803,ASUK-0804,ASUK-0805,ASUK-0810,ASUK-0816,ASUK-0817,ASUK-0818, ASUK-0819,,ASUK-0820,ASUK-0821,ASUK-0822,ASUK-0823,ASUK-0824,ASUK-0825,ASUK-0826, ASUK-0827,ASUK-0828,ASUK-0829,ASUK-0830,ASUK-0831,ASUK-0832,ASUK-0833,ASUK-0834, ASUK-0838,ASUK-0839,ASUK-0840,ASUK-0842,ASUK-0843,ASUK-0844,ASUK-0845,ASUK-0846, ASUK-0847,ASUK-0848,ASUK-0849,ASUK-0850,ASUK-0851,ASUK-0852,ASUK-0853,ASUK-0854, ASUK-0855,ASUK-0856,ASUK-0857,ASUK-0858,ASUK-0859,ASUK-0860,ASUK-0871,ASUK-0872, ASUK-0873,ASUK-0874,ASUK-0875,ASUK-0876,ASUK-0877,ASUK-0878,ASUK-0879,ASUK-0880, ASUK-0891,ASUK-0892,ASUK-0893,ASUK-0894,ASUK-0916,ASUK-0917,ASUK-0918,ASUK-0919, ASUK-0920,ASUK-0921,ASUK-0922,ASUK-0923,ASUK-0924,ASUK-0925,ASUK-0930,ASUK-0931, ASUK-0964,ASUK-0965,ASUK-0971,ASUK-0976,ASUK-0977,ASUK-0978,ASUK-0979,ASUK-0980, ASUK-0981,ASUK-0982,ASUK-0983,ASUK-0984,ASUK-0985,ASUK-0986,ASUK-0987,ASUK-0988, ASUK-0989,ASUK-0990,ASUK-0991,ASUK-0992,ASUK-1001,ASUK-1002,ASUK-1003,ASUK-1005, ASUK-1006,ASUK-1007,ASUK-1008,ASUK-1009,ASUK-1010,ASUK-1012,ASUK-1014,ASUK-1019, ASUK-1020,ASUK-1061,ASUK-1062,ASUK-1063,ASUK-1064,ASUK-1065,ASUK-1066,ASUK-1070, ASUK-1071,ASUK-1073,ASUK-1074,ASUK-1075,ASUK-1076,ASUK-1077,ASUK-1078,ASUK-1079, ASUK-1080,ASUK-1081,ASUK-1082,ASUK-1084,ASUK-1085,ASUK-1086,ASUK-1087,ASUK-1088, ASUK-1089,ASUK-1090,ASUK-1091,ASUK-1092,ASUK-1093,ASUK-1094,ASUK-1095,ASUK-1096, ASUK-1097,ASUK-1098,ASUK-1099,ASUK-1100,ASUK-1121,ASUK-1122,ASUK-1123,ASUK-1124, ASUK-1125,ASUK-1126,ASUK-1127,ASUK-1128,ASUK-1129,ASUK-1130,ASUK-1230,ASUK-1231, ASUK-1241,ASUK-1242,ASUK-1243,ASUK-1244,ASUK-1245,ASUK-1247,ASUK-1261,ASUK-1262, ASUK-1263,ASUK-1264,ASUK-1265,ASUK-1266,ASUK-1267,ASUK-1268,ASUK-1269,ASUK-1270, ASUK-1271,ASUK-1272,ASUK-1273,ASUK-1274,ASUK-1275,ASUK-1276,ASUK-1277,ASUK-1278, ASUK-1279,ASUK-1280,ASUK-1288,ASUK-1289,ASUK-1291,ASUK-1292,ASUK-1293,ASUK-1294, ASUK-1295,ASUK-1296,ASUK-1297,ASUK-1298,ASUK-1299,ASUK-1300,ASUK-1301,ASUK-1302, ASUK-1305,ASUK-1311,ASUK-1312,ASUK-1313,ASUK-1314,ASUK-1315,ASUK-1352,ASUK-1353, ASUK-1357,ASUK-1358,ASUK-1360,ASUK-1463,ASUK-1464,ASUK-1465,ASUK-1466,ASUK-1492, ASUK-1493,ASUK-1496,ASUK-1497,ASUK-1498,ASUK-1499,ASUK-1500,ASUK-1532.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K971923 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98002).