Draeger Vapor 2000. Unheated, calibrated anesthetic… recallZ-0933-2026 · Draeger, Inc. · initiated Nov 24, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A certain component of affected devices was not delivered within specification and contained impurities.
FDA root cause category: Under Investigation by firm
What the company is doing
Consignees were mailed an Urgent Medical Device Recall notification dated November 2025. The notification recommends that consignees do not use affected vaporizers and return the provided reply card to confirm this. A Draeger representative will contact consignees to arrange replacement of affected vaporizers. If affected vaporizers have been placed into clinical use, Draeger asks that consignees clean the breathing system of the anesthesia device according to the IFU and use of inspiratory filters is recommended. Consignees are to share the recall notification with all users of affected devices to ensure awareness and forward the notice if product was further distributed. Questions regarding operation of devices are to be directed to Drager Service Technical Support at 1-800-437-2437 (Option 2, 2, 2). Questions about the recall notice are to be directed to Michael Kelhart via email at [email protected] or phone at 267-664-1131.
Recall details
- Recall number
- Z-0933-2026
- Status
- Open, Classified
- Initiated by the company
- Nov 24, 2025
- Posted by the FDA
- Dec 23, 2025
- Recalling firm
- Draeger, Inc.
- Quantity
- 431 units
- Product code
- CAD
- FDA submissions
- K971923
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, CA, FL, IA, KS, MD, ME, MN, NY, OH, PA, RI, SD, TN, TX, UT, W and the countries of Argentina, Australia, Belgium, Brazil, Cambodia, China, Columbia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Hungary, Indonesia, Israel, Japan, Kenya, Mexico, Nepal, Norway, Panama, Peru, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates, Vietnam.
Product description
Draeger Vapor 2000. Unheated, calibrated anesthetic vaporizer.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Model No. M35054; UDI-DI: 04048675228059; Serial No. ASUK-0421,ASUK-0422,ASUK-0423,ASUK-0424,ASUK-0425,ASUK-0426,ASUK-0602,ASUK-0603, ASUK-0705,ASUK-0706,ASUK-0707,ASUK-0708,ASUK-0721,ASUK-0722,ASUK-0723,ASUK-0724, ASUK-0725,ASUK-0726,ASUK-0727,ASUK-0728,ASUK-0729,ASUK-0730,ASUK-0731,ASUK-0732, ASUK-0733,ASUK-0734,ASUK-0735,ASUK-0736,ASUK-0737,ASUK-0738,ASUK-0739,ASUK-0740, ASUK-0749,ASUK-0761,ASUK-0762,ASUK-0763,ASUK-0764,ASUK-0765,ASUK-0766,ASUK-0767, ASUK-0768,ASUK-0769,ASUK-0770,ASUK-0835,ASUK-0836,ASUK-0837,ASUK-0972,ASUK-0973, ASUK-0974,ASUK-0975,ASUK-0993,ASUK-0994,ASUK-0995,ASUK-0996,ASUK-0997,ASUK-0998, ASUK-0999,ASUK-1000,ASUK-1011,ASUK-1015,ASUK-1016,ASUK-1017,ASUK-1018,ASUK-1221, ASUK-1222,ASUK-1232,ASUK-1233,ASUK-1234,ASUK-1235,ASUK-1236,ASUK-1237,ASUK-1239, ASUK-1240,ASUK-1246,ASUK-1306,ASUK-1307,ASUK-1308,ASUK-1309,ASUK-1539,ASUK-1540, ASUK-0428,ASUK-0429,ASUK-0430,ASUK-0431,ASUK-0432,ASUK-0433,ASUK-0434,ASUK-0435, ASUK-0436,ASUK-0437,ASUK-0438,ASUK-0439,ASUK-0440,ASUK-0441,ASUK-0442,ASUK-0443, ASUK-0444,ASUK-0445,ASUK-0446,ASUK-0447,ASUK-0448,ASUK-0449,ASUK-0450,ASUK-0,63, ASUK-0564,ASUK-0565,ASUK-0566,ASUK-0567,ASUK-0568,ASUK-0569,ASUK-0570,ASUK-0571, ASUK-0572,ASUK-0573,ASUK-0574,ASUK-0575,ASUK-0576,ASUK-0577,ASUK-0578,ASUK-0579, ASUK-0580,ASUK-0581,ASUK-0582,ASUK-0583,ASUK-0584,ASUK-0585,ASUK-0601,ASUK-0604, ASUK-0605,ASUK-0606,ASUK-0607,ASUK-0608,ASUK-0609,ASUK-0610,ASUK-0631,ASUK-0632, ASUK-0633,ASUK-0634,ASUK-0635,ASUK-0636,ASUK-0637,ASUK-0638,ASUK-0639,ASUK-0640, ASUK-0641,ASUK-0642,ASUK-0661,ASUK-0662,ASUK-0663,ASUK-0664,ASUK-0667,ASUK-0668, ASUK-0681,ASUK-0682,ASUK-0691,ASUK-0692,ASUK-0693,ASUK-0694,ASUK-0695,ASUK-0696, ASUK-0697,ASUK-0698,ASUK-0699,ASUK-0700,ASUK-0711,ASUK-0712,ASUK-0713,ASUK-0714, ASUK-0715,ASUK-0716,ASUK-0717,ASUK-0718,ASUK-0719,ASUK-0720,ASUK-0741,ASUK-0742, ASUK-0743,ASUK-0744,ASUK-0746,ASUK-0747,ASUK-0748,ASUK-0750,ASUK-0751,ASUK-0752, ASUK-0753,ASUK-0754,ASUK-0755,ASUK-0756,ASUK-0757,ASUK-0758,ASUK-0759,ASUK-0760, ASUK-0778,ASUK-0779,ASUK-0780,ASUK-0781,ASUK-0782,ASUK-0783,ASUK-0784,ASUK-0785, ASUK-0786,ASUK-0787,ASUK-0788,ASUK-0789,ASUK-0790,ASUK-0791,ASUK-0792,ASUK-0793, ASUK-0794,ASUK-0795,ASUK-0796,ASUK-0797,ASUK-0798,ASUK-0799,ASUK-0800,ASUK-0801, ASUK-0802,ASUK-0803,ASUK-0804,ASUK-0805,ASUK-0810,ASUK-0816,ASUK-0817,ASUK-0818, ASUK-0819,,ASUK-0820,ASUK-0821,ASUK-0822,ASUK-0823,ASUK-0824,ASUK-0825,ASUK-0826, ASUK-0827,ASUK-0828,ASUK-0829,ASUK-0830,ASUK-0831,ASUK-0832,ASUK-0833,ASUK-0834, ASUK-0838,ASUK-0839,ASUK-0840,ASUK-0842,ASUK-0843,ASUK-0844,ASUK-0845,ASUK-0846, ASUK-0847,ASUK-0848,ASUK-0849,ASUK-0850,ASUK-0851,ASUK-0852,ASUK-0853,ASUK-0854, ASUK-0855,ASUK-0856,ASUK-0857,ASUK-0858,ASUK-0859,ASUK-0860,ASUK-0871,ASUK-0872, ASUK-0873,ASUK-0874,ASUK-0875,ASUK-0876,ASUK-0877,ASUK-0878,ASUK-0879,ASUK-0880, ASUK-0891,ASUK-0892,ASUK-0893,ASUK-0894,ASUK-0916,ASUK-0917,ASUK-0918,ASUK-0919, ASUK-0920,ASUK-0921,ASUK-0922,ASUK-0923,ASUK-0924,ASUK-0925,ASUK-0930,ASUK-0931, ASUK-0964,ASUK-0965,ASUK-0971,ASUK-0976,ASUK-0977,ASUK-0978,ASUK-0979,ASUK-0980, ASUK-0981,ASUK-0982,ASUK-0983,ASUK-0984,ASUK-0985,ASUK-0986,ASUK-0987,ASUK-0988, ASUK-0989,ASUK-0990,ASUK-0991,ASUK-0992,ASUK-1001,ASUK-1002,ASUK-1003,ASUK-1005, ASUK-1006,ASUK-1007,ASUK-1008,ASUK-1009,ASUK-1010,ASUK-1012,ASUK-1014,ASUK-1019, ASUK-1020,ASUK-1061,ASUK-1062,ASUK-1063,ASUK-1064,ASUK-1065,ASUK-1066,ASUK-1070, ASUK-1071,ASUK-1073,ASUK-1074,ASUK-1075,ASUK-1076,ASUK-1077,ASUK-1078,ASUK-1079, ASUK-1080,ASUK-1081,ASUK-1082,ASUK-1084,ASUK-1085,ASUK-1086,ASUK-1087,ASUK-1088, ASUK-1089,ASUK-1090,ASUK-1091,ASUK-1092,ASUK-1093,ASUK-1094,ASUK-1095,ASUK-1096, ASUK-1097,ASUK-1098,ASUK-1099,ASUK-1100,ASUK-1121,ASUK-1122,ASUK-1123,ASUK-1124, ASUK-1125,ASUK-1126,ASUK-1127,ASUK-1128,ASUK-1129,ASUK-1130,ASUK-1230,ASUK-1231, ASUK-1241,ASUK-1242,ASUK-1243,ASUK-1244,ASUK-1245,ASUK-1247,ASUK-1261,ASUK-1262, ASUK-1263,ASUK-1264,ASUK-1265,ASUK-1266,ASUK-1267,ASUK-1268,ASUK-1269,ASUK-1270, ASUK-1271,ASUK-1272,ASUK-1273,ASUK-1274,ASUK-1275,ASUK-1276,ASUK-1277,ASUK-1278, ASUK-1279,ASUK-1280,ASUK-1288,ASUK-1289,ASUK-1291,ASUK-1292,ASUK-1293,ASUK-1294, ASUK-1295,ASUK-1296,ASUK-1297,ASUK-1298,ASUK-1299,ASUK-1300,ASUK-1301,ASUK-1302, ASUK-1305,ASUK-1311,ASUK-1312,ASUK-1313,ASUK-1314,ASUK-1315,ASUK-1352,ASUK-1353, ASUK-1357,ASUK-1358,ASUK-1360,ASUK-1463,ASUK-1464,ASUK-1465,ASUK-1466,ASUK-1492, ASUK-1493,ASUK-1496,ASUK-1497,ASUK-1498,ASUK-1499,ASUK-1500,ASUK-1532.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vapor 2000 M35054Drägerwerk AG & Co. KGaA · Vaporizer, Anesthesia, Non-Heated1 recall8 reports
Other devices under the same FDA submission
1 device cleared under K971923 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vapor 3000 M36500Drägerwerk AG & Co. KGaA · Vaporizer, Anesthesia, Non-Heated1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98002).
- Z-0934-2026 · Open, ClassifiedDraeger Vapor 3000. Unheated, calibrated anesthetic vaporizer.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.