Brand Name: Xpert BCR-ABL Ultra recallZ-0959-2026 · Cepheid · initiated Oct 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cepheid or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.
FDA root cause category: Process design
What the company is doing
On November 14, 2025, Cepheid issued a "Urgent Medical Device Recall" notification to affected consignees. Cepheid asked consignees to take the following actions: 1. Cepheid recommends that laboratories and clinical facilities utilizing the Xpert BCR-ABL Ultra assay for quantification of BCR::ABL1 and ABL1 mRNA transcript levels in peripheral blood specimens from t(9;22)-positive chronic myeloid leukemia (CML) patients undergoing tyrosine kinase inhibitor (TKI) therapy, assess white blood cell count (WBCC) prior to testing in cases where elevated white blood cell count is suspected. 2. If WBCCC exceeds 30,000 cells/uL, the specimen should be processed in the accordance with the Retest Procedure for ERROR (Code 2008) or INVALID (Type 2) as detailed in the IFU for the following scenarios: - Cases where INVALID test results are observed and are associated with an elevated WBCC. - Cases of valid results in which an elevated WBCC has been observed. 3. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. 4. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them with a copy of this letter. 5. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication.
Recall details
- Recall number
- Z-0959-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 30, 2025
- Posted by the FDA
- Dec 18, 2025
- Recalling firm
- Cepheid
- Quantity
- 6325 units
- Product code
- OYX
- FDA submissions
- K190076
- Distribution
- US: AL AR AZ CA CO FL GA IA IL KS KY LA MD ME MI MN MO NC NE NH NJ NY OH PA RI TX VA WA WI OUS: Algeria Armenia Australia Austria Bahrain Bangladesh Belgium Brazil Burkina Faso Cambodia Canada Canary islands Chile Colombia Cyprus Czech Republic Dem. Rep. Congo Djibouti Dominican Rep. Ecuador El Salvador Eswatini Finland France Frenc.Polynesia Gambia Georgia Germany Ghana Greece Guadeloupe Guatemala Honduras Hong Kong India Indonesia Iraq Israel Italy Jamaica Jordan Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Luxembourg Malawi Malaysia Mauritania Mauritius Mexico Morocco Mozambique Myanmar Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pap. New Guinea Peru Philippines Poland Portugal Qatar Rep.of Congo Reunion Rwanda Saudi Arabia Senegal Seychelles Singapore South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tajikistan Tanzania Thailand Togo Tunisia Uganda United Kingdom Uruguay Utd.Arab Emir. Venezuela Zambia Zimbabwe
Product description
Brand Name: Xpert BCR-ABL Ultra Model/Catalog Number: GXBCRABL-US-10
Code information (models, lots, serial numbers)
UDI: 07332940007102/ Lot Numbers: 1001456698, 1001463946, 1001472599, 1001477457, 1001480637, 1001483326, 1001500425
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Xpert BCR-ABL Ultra GXBCRABL-US-10Cepheid · Bcr/Abl1 Monitoring Test1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page