LifeShield Drug Library Management (DLM), LifeShield… recallZ-0966-2026 · ICU Medical, Inc. · initiated Nov 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

FDA root cause category: Software Design Change

What the company is doing

On November 24, 2025, the firm notified customers via Urgent Medical Device Correction letters. Consignees were asked to take the following actions: 1. Ensure all users or potential users of LifeShield DLM 2.2.1 are immediately made aware of this notification and proposed mitigations. 2. Until a software update is released, limit any newly created or edited Concentration Limits to one decimal place of precision for limits less than 100. ICU Medical is developing a software update to correct this issue and will notify customers of the timing of the software update.

Recall details

Recall number
Z-0966-2026
Status
Open, Classified
Initiated by the company
Nov 24, 2025
Posted by the FDA
Dec 19, 2025
Recalling firm
ICU Medical, Inc.
Quantity
12 customers
Product code
PHC
FDA submissions
K242117
Distribution
US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

Product description

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Code information (models, lots, serial numbers)

LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227