Revogene C. difficile, in vitro diagnostic assay intended… recallZ-1019-2026 · Meridian Bioscience Inc · initiated Dec 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Meridian Bioscience Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The affected lots show a decline in performance over time, which may lead to false-negative results.

FDA root cause category: Under Investigation by firm

What the company is doing

An "URGENT MEDICAL DEVICE CORRECTION" dated 12/1/25 was sent to customers. Required Actions: " Identify and segregate any remaining inventory of impacted product lots and suspend further use. " Destroy remaining inventory of impacted product lot in accordance with state, local, and/or federal regulations. o Contact Technical Service for replacement product. " Review previously obtained results from Revogene C. difficile listed Lots numbers. Any negative results obtained should be re-evaluated in conjunction with presented clinical findings to determine if retesting is required. " Complete and return the CONFIRMATION OF NOTIFICATION Form. Complete this form even if there is no remaining inventory. Actions to Be Taken By Meridian Bioscience, Inc.: Meridian Bioscience, Inc. will replace any unused inventory and provide support for any necessary retesting, including replacement test materials as needed. Contact Information: For any question regarding this notification please call Meridian Bioscience Technical Service at 1-800-343-3858, or email at [email protected].

Recall details

Recall number
Z-1019-2026
Status
Open, Classified
Initiated by the company
Dec 5, 2025
Posted by the FDA
Jan 5, 2026
Recalling firm
Meridian Bioscience Inc
Quantity
211 units
Product code
OZN
FDA submissions
K172569
Distribution
US Nationwide distribution and the OUS country of Italy.

Product description

Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300

Code information (models, lots, serial numbers)

UDI-DI: 00840733102349; Lot Numbers: 410300U195, 410300U196, 410300U197, 410300U201

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.