Affixus recallZ-0982-2026 · Zimmer, Inc. · initiated Dec 2, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA root cause category: Process design
What the company is doing
Firm began notifying customers on December 2, 2025 via Urgent Medical Device Recall letters. Distributors are to locate and quarantine affected product in their inventory. They must also return all affected product from their distributorship and from affected hospitals within their territory. Risk Managers are to assist Zimmer Biomet sales representatives in quarantining all affected product. Sales representatives will remove the product from customer facilities. Surgeons are to maintain awareness of the issue, but there are no patient-monitoring instructions related to this recall recommended beyond their existing follow-up schedules.
Recall details
- Recall number
- Z-0982-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 2, 2025
- Posted by the FDA
- Dec 22, 2025
- Recalling firm
- Zimmer, Inc.
- Quantity
- 3
- Product code
- HSB
- FDA submissions
- K241651
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
Product description
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815809280 Software Version: N/A Product Description: Affixus 9 mm, Length 280 mm, Left, Priformis Nail Component: N/A
Code information (models, lots, serial numbers)
Lot Code: Model No 815809280 ; UDI-DI (01)00887868584763(17)350416(10)67066287 ; Lot Number 67066287
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Affixus® 815809280Zimmer, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
Product lines
Other devices under the same FDA submission
16 devices cleared under K241651 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Affixus® 815509300Zimmer, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- Affixus® 815509320Zimmer, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- Affixus® 815509340Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815509360Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815509380Zimmer, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- Affixus® 815509400Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815509420Zimmer, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- Affixus® 815509440Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815609300Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815609320Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815609340Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
- Affixus® 815609360Zimmer, Inc. · Screw, fixation, bone1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98069).
- Z-0967-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509320 Software Version: N/A Product Description: Affixu…
- Z-0968-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509340 Software Version: N/A Product Description: Affixu…
- Z-0969-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509360 Software Version: N/A Product Description: Affixu…
- Z-0970-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509380 Software Version: N/A Product Description: Affixu…
- Z-0971-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509400 Software Version: N/A Product Description: Affixu…
- Z-0972-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420 Software Version: N/A Product Description: Affixu…
- Z-0973-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509440 Software Version: N/A Product Description: Affixu…
- Z-0974-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609300 Software Version: N/A Product Description: Affixu…
- Z-0975-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609320 Software Version: N/A Product Description: Affixu…
- Z-0976-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609340 Software Version: N/A Product Description: Affixu…
- Z-0977-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609360 Software Version: N/A Product Description: Affixu…
- Z-0978-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609380 Software Version: N/A Product Description: Affixu…
- Z-0979-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609400 Software Version: N/A Product Description: Affixu…
- Z-0980-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609420 Software Version: N/A Product Description: Affixu…
- Z-0981-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609440 Software Version: N/A Product Description: Affixu…
- Z-0983-2026 · Open, ClassifiedBrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509300 Software Version: N/A Product Description: Affixu…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page