Skip to content

The Pressio 2 ICP Monitoring System, Model Number PSO-4000 recallZ-1029-2026 · Sophysa · initiated Nov 17, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sophysa or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customer complaints of Pressio monitor rebooting.

FDA root cause category: Software design

What the company is doing

Sophysa issued a Field Safety notice to its consignees on 11/25/2025 via email. The notice explained the issue and requested monitoring of the patient should the unintended reboot of the monitor occur. The consignees were directed to notify their sales representatives to schedule the monitor update.

Recall details

Recall number
Z-1029-2026
Status
Open, Classified
Initiated by the company
Nov 17, 2025
Posted by the FDA
Jan 8, 2026
Recalling firm
Sophysa
Quantity
105 units
Product code
GWM
FDA submissions
K162108
Distribution
US Nationwide distribution in the states of AZ, WI, TX, OR, CA, OK, ID, FL, NJ.

Product description

The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor

Code information (models, lots, serial numbers)

UDI/DI 03760124132076, All monitors

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page