BVI Leos Laser and Endoscopy System. Model Number: OME6000U recallZ-1008-2026 · Beaver-Visitec International, Inc. · initiated Dec 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beaver-Visitec International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

FDA root cause category: Process control

What the company is doing

Beaver-Visitec International (BVI) notified consignees via emailed letter on 12/05/2025. Consignees were informed that the affected units can still be used. A BVI service technician will contact each consignee to schedule a visit. During this visit, the missing ferrites will be installed to restore the equipment to its fully compliant and validated configuration. BVI aims to complete the correction by December 31st, 2025.

Recall details

Recall number
Z-1008-2026
Status
Open, Classified
Initiated by the company
Dec 5, 2025
Posted by the FDA
Dec 31, 2025
Recalling firm
Beaver-Visitec International, Inc.
Quantity
5 units
Product code
HQF
FDA submissions
K240615
Distribution
US Nationwide distribution in the states of FL, MI, OK, TX, WI.

Product description

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Code information (models, lots, serial numbers)

Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.