Brand Name: Molift recallZ-1308-2026 · Etac A/S · initiated Nov 17, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Etac A/S or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

FDA root cause category: Component design/selection

What the company is doing

On November 17, 2025 customers were notified vis a phone call. Notifying them that they will receive an email with the Field Safety Notice and Response Form. The customers were emailed Field Safety Notice letters on or around November 20, 2025. Corrective action: To reduce the risk of further incidents, replacement units are offered for all potentially affected products. Continued use is safe as long as the lifting bar can move freely and remains vertical during the lifting cycle with or without load. If this is not the case, or if there is any doubt about the product s functionality, the unit must be taken out of service immediately. The components in the defective units (Mover 205 and Mover 300) will gradually be replaced in the coming months. The affected lifting bars will likewise be replaced.

Recall details

Recall number
Z-1308-2026
Status
Open, Classified
Initiated by the company
Nov 17, 2025
Posted by the FDA
Feb 5, 2026
Recalling firm
Etac A/S
Quantity
95 units
Product code
FSA
Distribution
Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

Product description

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

(1) Model/Catalog Number: M14000; UDI-DI: 05707995032270; Serial Numbers: 616606, 616604, 591003, 591002, 590906, 590901, 590904, 590903, 590902, 591001, 604101, 604005, 591703, 591007, 591006, 591005, 591004, 615105, 615104, 615103, 615102, 615101, 604004, 604003, 604002, 604001, 614202, 614201, 591702, 591701, 616603, 615106, 590905; (2) Model/Catalog Number: M1401; UDI-DI: 05707995032270; Serial Numbers: 593302, 593301, 593303, 612507, 612506, 612508, 612505, 612503, 620702, 612504, 612502, 612501, 593304, 593305, 620703, 620701; (3) Model/Catalog Number: M1402; UDI-DI: 05707995032270; Serial Numbers: 612601; (4) Model/Catalog Number: M14050; UDI-DI: 05707995032270; Serial Numbers: 590801, 596701, 585701, 584501, 591801, 604405, 604406, 604402, 604404, 604403, 604401, 604303, 604302, 604301; (5) Model/Catalog Number: M14150; UDI-DI: 05707995032270; Serial Numbers: 585811, 585810, 585809, 585808, 585806, 585805, 585804, 585803, 585801, 596604, 596603, 596602, 596601, 596611, 596610, 596609, 596608, 596607, 596606, 596605, 585807, 585802; (6) Model/Catalog Number: M1402-01; UDI-DI: 05707995032270; Serial Numbers: 612701, 612703, 612702; (7) Model/Catalog Number: M14050-00; UDI-DI: 05707995032270; Serial Numbers: 584307; (8) Model/Catalog Number: M14001; UDI-DI: 05707995032270; Serial Numbers: 603501, 599801, 599802, 604202, 604201;