Skip to content

Detachable 10in Pendant Upgrade component for the AIRO… recallZ-1041-2026 · Mobius Imaging, LLC · initiated Dec 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mobius Imaging, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.

FDA root cause category: Labeling Change Control

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION notice dated 12/4/25 instructs consignees to ensure the following mitigations are in place while the correction is deployed: the hand position warnings as indicated in the IFU, train operators, the AIRO is accessible through a physical key switch and password protected log in screen, and the Backing Up System Warning label is affixed on to the gantry of affected devices. Stryker will correct devices during an upcoming service visit. Consignees are to inform users of devices of this recall by sharing the recall notice, immediately check stock areas and operating rooms to identify affected devices, and inform all users of the potential harms and their risk mitigations. Consignees with any questions are to contact their local sales office or Stryker Representative.

Recall details

Recall number
Z-1041-2026
Status
Open, Classified
Initiated by the company
Dec 4, 2025
Posted by the FDA
Jan 9, 2026
Recalling firm
Mobius Imaging, LLC
Quantity
26 units
Product code
JAK
FDA submissions
K180393
Distribution
US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.

Product description

Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System.

Code information (models, lots, serial numbers)

Part No. MI-70-0128; UDI: (01) 00869346000200 (11)230920 (21) AIRO-XXXX; Serial No. 2232206183, 2232206193, 2232206203, 2232206213, 2232206223, 2232206233, 2235522143, 2235522153, 2235522163, 2236339153, 2236339163, 2236339173, 2236339183, 2303429483, 2303429503, 2313925663, 2313925673, 2316030773, 2316030783, 2316703503, 2316703523, 2317415473, 2320006923, 2320006933, 2323513413, 70-0128-13904392.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98099).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page