Skip to content

Focalyx Fusion recallZ-1243-2026 · Focalyx Technologies, LLC. · initiated Dec 23, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Focalyx Technologies, LLC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

FDA root cause category: Software change control

What the company is doing

On 12/23/2025, correction notices started being emailed to customers informing them of the following: You will be contacted to schedule installation of the new version. An IFU will be provided at time of installation. Complete and return the response form via email to [email protected] If there are any questions, contact the firm at: focalyxtechnologies.com/contact, [email protected], (888) 636 2259, (Sel. 3 Technical Support)

Recall details

Recall number
Z-1243-2026
Status
Open, Classified
Initiated by the company
Dec 23, 2025
Posted by the FDA
Feb 3, 2026
Recalling firm
Focalyx Technologies, LLC.
Quantity
18
Product code
LLZ
FDA submissions
K170250
Distribution
Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.

Product description

Focalyx Fusion

Code information (models, lots, serial numbers)

UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates