Skip to content

SQ40S Blood Transfusion Filter recallZ-1157-2026 · GVS TM, Inc · initiated Nov 3, 2025

Class II Open, Classified All recalls posted in January 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GVS TM, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to products being released for distribution prior to completion of the required quality control release process the sterility assurance cannot be confirmed.

FDA root cause category: Process control

What the company is doing

On 11/03/2025, the firm sent via email a "GVS TM Inc Customer Statement" informing customers that GVS TM Inc discovered that some of the products in the affected lot may have been damaged during the transportation process. To avoid any potential risk this damage may pose to the downstream user, the firm is requesting return of the affected products. On 11/18/2025, the firm sent via email a "GVS TM Inc Product Recall Statements" informing customers that the products were distributed before completion of the required quality control release process, including final review of sterilization documentation and Certificate of Analysis. As a result the sterility assurance cannot be confirmed. Customers are instructed to: 1) Return all units of Lot 7536142 to GVS TM Inc - Customer Service using GVS TM FedEx Account 203742698, or 2) Destroy the affected units at their facility according to their internal disposal procedures for medical devices. 3) Confirmation Required - If Returning Product --Quantity returned --Date Shipped Back --Contact Person - If Destroying Product (Proof of Destruction Required) --Signed destruction confirmation --Statement on facility letterhead indicating: ---Quantity destroyed ---Date of Destruction ---Method of disposal ---Name/title of the person performing destruction Upon receipt of the above information, GVS TM Inc will issue replacements or credit depending on customer's preference. Questions - Contact GVS TM Customer Service at customerservice [email protected]

Recall details

Recall number
Z-1157-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Nov 3, 2025
Posted by the FDA
Jan 22, 2026
Recalling firm
GVS TM, Inc
Quantity
2,720 filters
Product code
CAK
FDA submissions
K960993
Distribution
U.S.: ID, NV, SC, MO, TX, CO, MD, CA, VA, DC, PA, IL, OH, AZ, FL, OK, MN, KY, OR, NE, NC, LA, MS, MI, NY, GA, AL, MA

Product description

SQ40S Blood Transfusion Filter

Code information (models, lots, serial numbers)

Product Number: SQ40S UDI-DI code: 10887691000666 Lot Number: 7536142

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page