Asserachrom HPIA recallZ-1084-2026 · Diagnostica Stago, Inc. · initiated Dec 15, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.

FDA root cause category: Under Investigation by firm

What the company is doing

Diagnostica Stago notified consignees of the recall on about 12/15/2025 via email. Consignees were instructed to stop using Lot 271288 and dispose of all remaining product according to local regulations, and complete, sign, and return the attached Acknowledgement Form provided. Additionally, consignees were instructed that if they have distributed the affected lot to any other laboratory, please notify those recipients of this recall and ask them to complete and return an Acknowledgement Form.

Recall details

Recall number
Z-1084-2026
Status
Open, Classified
Initiated by the company
Dec 15, 2025
Posted by the FDA
Jan 14, 2026
Recalling firm
Diagnostica Stago, Inc.
Quantity
438 units
Product code
LCO
FDA submissions
K003767
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, HI, IA, MA, MI, NJ, NV, NY, OH, TN, WI and the countries of Belgium, Canada, Denmark, Algeria, Spain, Finland, France, Great Britain, Italy, Lebanon, Saudi Arabia, Uganda.

Product description

Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.

Code information (models, lots, serial numbers)

Reference Number: 00615. UDI: (01)03607450006155(11)240531(17)260531(10)271288(241)00615. Lot Number: 271288.