Merit Medical , Custom Waste Management Kit Vascular Tray recallZ-1125-2026 · Merit Medical Systems, Inc. · initiated Dec 4, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inflation device handle may detach from the syringe during procedure.
FDA root cause category: Process control
What the company is doing
On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at [email protected] within seven (7) calendar days
Recall details
- Recall number
- Z-1125-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 4, 2025
- Posted by the FDA
- Jan 16, 2026
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 864 units
- Product code
- PPD
- Distribution
- Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Korea, Republic of Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Madagascar, Mexico, Monaco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Trinidad and Tobago, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian, Republic of Viet Nam.
Product description
Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261
Code information (models, lots, serial numbers)
UDI: 00884450320297/ Lot: T2736369 T2781930 T2898771S T3047503 T3090431 T3121937 T3146114 T3158708 T3204129 T3232415
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Merit Medical® 00884450320297Merit Medical Systems, Inc. · Universal drainage tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98139).
- Z-1123-2026 · Open, ClassifiedMerit Medical Inflation Device.basixCOMPAK.30 atm/bar.20 mL Reference Numbers: ENDO-AN2030/B IN4130/CNK IN4130/H IN4130/JPH IN4130/K IN4130/KRK IN4130/L IN4130/T IN4230/JPH IN4230…
- Z-1126-2026 · Open, ClassifiedMerit Medical , Custom Inflation Kit REF: K05-00761F K05-01717B K05-02080D K05-02090C K05-02221 K05-02338 K05-02470 K05-02954 K05-03094A K05T-00380C K05T-01674B K05T-01863 K05T-019…
- Z-1128-2026 · Open, ClassifiedMerit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.