Medline medical procedure convenience kits labeled as: KIT… recallZ-1090-2026 · Medline Industries, LP · initiated Dec 16, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/16/2025 via email and USPS first class mail. The notice explained the issue, potential risk, and requested the following: "1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantityof affected product you have in inventory on the form. Even if you do not have any affected productin inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-25-145-FGX5 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]."
Recall details
- Recall number
- Z-1090-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 16, 2025
- Posted by the FDA
- Jan 14, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 12 units
- Product code
- OKV
- Distribution
- US Nationwide distribution.
Product description
Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
Code information (models, lots, serial numbers)
UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYKMBNDL116GMedline Industries, Inc. · Obstetrical kit4 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98197).
- Z-1085-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
- Z-1086-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAP…
- Z-1087-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) …
- Z-1088-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-…
- Z-1089-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS98534…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page