Skip to content

Mazor X robotic guidance system recallZ-1122-2026 · Mazor Robotics Ltd · initiated Dec 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mazor Robotics Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software errors that can result in incorrect surgical instrument positioning during spinal surgery.

FDA root cause category: Under Investigation by firm

What the company is doing

Beginning December 10, 2025, Medtronic (Mazor Robotics, Ltd) issued a Urgent Medical Device Recall Notification to affected consignees via E-Mail. Medtronic asked consignees to take the following actions: 1. Please review this information with all physician users and/or post a copy of this notification with your Mazor X system until the software update is completed. 2. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at [email protected]. 3. Provide this notification to those who need to be aware within your organization or to any organization where the affected systems have been transferred. 4. Maintain a copy of this notice in your records. 5. Continue to use the Mazor X system as instructed in the user manual and with the mitigations provided below in Appendix A.

Recall details

Recall number
Z-1122-2026
Status
Open, Classified
Initiated by the company
Dec 10, 2025
Posted by the FDA
Jan 16, 2026
Recalling firm
Mazor Robotics Ltd
Quantity
549 systems
Product code
OLO
FDA submissions
K230064
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, China, Croatia, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Vietnam.

Product description

Mazor X robotic guidance system REF: TPL0059

Code information (models, lots, serial numbers)

Software versions: 5.0.1, 5.1.2, 5.1.3/ UDI: 07290109180465, 07290109184524, 07290109181158, 00763000635169, 07290109183213, 07290109184517, 00763000419134, 00763000431761, 07290109184098, 07290109184838, 07290115751376, 07290115751895

Devices named in this recall

9 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page