Leksell GammaPlan recallZ-1339-2026 · Elekta, Inc. · initiated Dec 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
FDA root cause category: Use error
What the company is doing
On 12/12/2025, the firm sent an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" letter via email to customers informing them that when Defining a new stereotactic reference in an examination for a frame treatment while failing to first change an already defined obsolete stereotactic reference (representing a non-current frame fitting) to a pre-plan reference, may result in a treatment plan that irradiates the wrong patient location if the resulting inconsistent alignment between the stereotactic references is not detected by the user. Customers are instructed to: If the stereotactic reference for a frame treatment must be changed prior to plan approval (e.g. due to a change of frame fitting), make sure to first change the already defined obsolete stereotactic reference to a pre-plan reference before defining the new stereotactic reference. By so doing, the treatment examination is automatically updated according to the new stereotactic reference when it is defined. Do not combine the use of a stereotactic CBCT reference with other stereotactic references for treatments with a frame. This specific recommendation for CBCT is motivated by the increased likelihood of failure for users to detect an inconsistent stereotactic CBCT reference relative to other stereotactic references, compared to MR and CT image studies, which are typically displayed and explored throughout the planning process. Post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. " Advise the appropriate personnel, working with this product, on the content of this letter. For Questions - Contact Elekta Care Support at 855-693-5358 or email [email protected]
Recall details
- Recall number
- Z-1339-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2025
- Posted by the FDA
- Feb 10, 2026
- Recalling firm
- Elekta, Inc.
- Product code
- MUJ
- FDA submissions
- K151666
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
Product description
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.
Code information (models, lots, serial numbers)
Lot Code: 07340048311236 (LGP 11.1)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Leksell GammaPlan 1529980Elekta Instrument AB · System, radiation therapy, radionuclide1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98212).
- Z-1340-2026 · Open, ClassifiedBrand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan is a computer-based system…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page