Skip to content

VOLT Wrist Treatment System recallZ-1131-2026 · Tyber Medical · initiated Dec 19, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tyber Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The supplier manufactured anatomical left plates with an incorrect thread orientation.

FDA root cause category: Process control

What the company is doing

On December 19, 2025, an URGENT MEDICAL DEVICE REMOVAL NOTIFICATION email was sent to their only customer. Actions to be taken: 1. Acknowledge the receipt of this communication via email correspondence to [email protected] 2. Further distribute this communication through notifying the accounts that have received the identified devices. 3. Identify via product label and quarantine the impacted devices under DePuy Synthes control all devices that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical. 4. Identify via product label and quarantine the impacted devices that have been distributed. An RMA will be provided for the return of goods to Tyber Medical. 5. Complete the customer response form following this notification, or provide written notification, and return it to Tyber Medical via email correspondence to [email protected]. A sub-notification letter was also provided to the distributor to provide to customers with the following actions: 1. Identify via product label and quarantine the impacted devices in your possession. 2. Complete the subsequent Acknowledgement and Receipt From and return it to DePuy Synthes via email correspondence, with Tyber Medical copied at [email protected] 3. Return the impacted product to DePuy Synthes for replacement.

Recall details

Recall number
Z-1131-2026
Status
Open, Classified
Initiated by the company
Dec 19, 2025
Posted by the FDA
Jan 20, 2026
Recalling firm
Tyber Medical
Quantity
100 units
Product code
HRS
FDA submissions
K232652
Distribution
US Nationwide distribution in the state of PA.

Product description

Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/10Shaft/Left/Std/139mm Model/Catalog Number: 02.426.681S Product Description: 2 Column Distal Radius Plate - 6 Head, 10 Shaft, Standard Left - 139mm

Code information (models, lots, serial numbers)

Model No: 02.426.681S; UDI-DI: 00196449020508; Lot Number: 291294;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98214).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page