Skip to content

Axios recallZ-1158-2026 · Boston Scientific Corporation · initiated Dec 19, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

FDA root cause category: Process control

What the company is doing

On December 19, 2025, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: - Immediately stop further distribution or use of any remaining affected devices. Remove them from your facility s inventory and segregate them in a secure location until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. " If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. " Boston Scientific reminds healthcare professionals that AXIOS Stents should only be used as indicated in Instructions For Use.

Recall details

Recall number
Z-1158-2026
Status
Open, Classified
Initiated by the company
Dec 19, 2025
Posted by the FDA
Feb 5, 2026
Recalling firm
Boston Scientific Corporation
Quantity
557 units
Product code
KNS
FDA submissions
K233318
Distribution
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.

Product description

AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

Material Number (UPN): M00553680; UDI-DI: 8714729951179; Lot Numbers: 35961006 35961007 35962524 35967379 35968520, 35968521, 35968597, 35968598, 35971077, 35971078, 35971079, 35972100, 35982272, 35982273, 35982274, 35982275, 35982276, 35982277, 35982278, 35982279, 35982420, 35982424, 35982425, 35985313, 35985314, 35993619, 35993720, 35993721, 36109761, 36109763, 36109764, 36109765, 36109766, 36111314, 36111315, 36118993, 36118994, 36122066, 36122067, 36122075, 36132801, 36132802, 36132803, 36132804, 36134972, 36144058, 36144059, 36144120, 36144121, 36174475, 36174476, 36178005, 36187588, 36187589, 36187687, 36187688, 36188887, 36188888, 36188889, 36194446, 36194447, 36197239, 36198580, 36217768, 36217769, 36217770, 36227341, 36227342, 36227343, 36228603, 36228604, 36228605, 36235934, 36236151, 36240589, 36240590, 36241715, 36241716, 36241717, 36251363, 36251364, 36253000, 36253008, 36253041, 36260860, 36260861, 36260862, 36261599, 36261780, 36261781, 36261834, 36261835, 36280395, 36280396, 36280397, 36281912, 36281913, 36281914, 36283860, 36292682, 36292683, 36292684, 36292689, 36292690, 36292691, 36294495, 36301621, 36301622, 36301623, 36301624, 36302290, 36302291, 36302292, 36319024, 36319025, 36319026, 36320016, 36320017, 36323789, 36323790, 36324720, 36325159, 36325560, 36325568, 36325569, 36332101, 36332102, 36341577, 36342114, 36342418, 36352008, 36352009, 36355049, 36355050, 36355752, 36359567, 36362263, 36362264, 36362265, 36363150 36363151 36364647 36372361 36372362, 36377757, 36378548, 36380549, 36388531, 36390612, 36396985, 36399492, 36403441, 36408983, 36408984, 36417820, 36417821, 36418721, 36418722, 36424650, 36424651, 36424652, 36425889, 36425890, 36425891, 36425892, 36462323, 36462324, 36469427, 36469428, 36469812, 36472250, 36472253, 36472254, 36475370, 36478478, 36478479, 36572740, 36572741, 36573582, 36573583, 36573587, 36573588, 36573589, 36573590, 36591073, 36591074, 36591075, 36603384, 36603385, 36603386, 36604188, 36604189, 36604190, 36604191, 36604748, 36604749, 36607249, 36607250, 36607251, 36607252, 36607253, 36610818, 36610819, 36611060, 36615703, 36615704, 36621589, 36621590, 36623324, 36623325, 36623326, 36693849, 36693850, 36693851, 36693852, 36694699, 36695020, 36695021, 36695023, 36698247, 36698249, 36698251, 36698527, 36698528, 36698529, 36698530, 36698531, 36698532, 36706712, 36706713, 36706714, 36706715, 36709766, 36709767, 36709768, 36709769, 36713255, 36713256, 36713257, 36713258, 36717717, 36717718, 36717719, 36717800, 36717801, 36717802, 36718966, 36718967, 36723909, 36723910, 36728154, 36764237, 36764238, 36767449, 36767450, 36771884, 36771885, 36771886, 36785593, 36788793, 36788794, 36788795, 36795369, 36795370, 36796430, 36796431, 36804013, 36804375, 36839665, 36895362, 36895363, 36895364, 36958216, 36958217, 36962750, 36970218, 36970219, 36980309, 36980310 36981855 36981856 36982694 36983909, 36985933, 36988630, 36988631, 36988632, 36990382, 36990383, 36991168, 36991170, 36991683, 36992721, 37014102, 37014103, 37014104, 37027339, 37028180, 37028181, 37030756, 37032104, 37036344, 37036345, 37037505, 37037506, 37040169, 37041012, 37041435, 37041436, 37046449, 37046450, 37046928, 37046929, 37048665, 37049464, 37049465, 37049874, 37050321, 37050322, 37050323, 37055465, 37055466, 37055467, 37056013, 37056014, 37057032, 37059006, 37059010, 37094844, 37106164, 37106583, 37131143, 37131144, 37131145, 37131146, 37177253, 37177271, 37177272, 37179955, 37179956, 37180902, 37180903, 37182323, 37183490, 37188257, 37188258, 37197563, 37197578, 37198908, 37199019, 37199360, 37199361, 37199972, 37199973, 37209169, 37209170, 37209171, 37211282, 37211285, 37212243, 37212244, 37212245, 37212246, 37213246, 37213247, 37213248, 37213249, 37240012, 37240013, 37240236, 37240237, 37240249, 37240250, 37240252, 37240254, 37242247, 37242248, 37242249, 37243754, 37243755, 37243756, 37247304, 37247305, 37247306, 37283604, 37283605, 37291851, 37291852, 37292951, 37292952, 37296017, 37296018, 37296019, 37297941, 37300544, 37300545, 37300546, 37302814, 37308038, 37308039, 37325355, 37328835, 37328836, 37328838, 37331112, 37332012, 37333358, 37333362, 37337364, 37340128, 37340129, 37340130, 37340131, 37340270, 37340271, 37341078, 37341079, 37344296 37344297 37346443 37346445 37348330, 37348331, 37352140, 37352141, 37354943, 37354944, 37357958, 37357959, 37358120, 37359884, 37359885, 37365502, 37365503, 37365504, 37365505, 37375861, 37388591, 37388592, 37388593, 37400608, 37402464, 37402466, 37410374, 37411886, 37411887, 37415923, 37415924, 37415925, 37427250, 37429329, 37429330, 37438134, 37438135, 37439609, 37439610, 37439611, 37451308, 37451309, 37454250, 37454251, 37456595, 37456596, 37459573, 37567407, 37573089, 37573090, 37573091, 37575640, 37578876, 37578877, 37578879, 37579368, 37579369, 37579370, 37582027, 37585524, 37588798, 37588799, 37589260, 37592382, 37592383, 37592384, 37596205, 37596206, 37596207, 37596208, 37601263, 37602176, 37602177, 37602178, 37602179, 37603348, 37607457, 37607458, 37607459, 37607540, 37625485, 37625486, 37626455, 37626456, 37631100, 37631101, 37633190, 37633191, 37638120, 37638121, 37638122, 37639137, 37639138, 37644407, 37644408, 37644409, 37649491, 37649492, 37686664, 37686665, 37686666, 37700937, 37700938, 37706532, 37706533, 37706534, 37708203, 37708204, 37716332, 37716333, 37718549, 37720553, 37720554, 37720555, 37727831, 37730055, 37754383, 37758168, 37758169, 37758170, 37759554, 37761362, 37761363, 37785682, 37785683, 37787089, 37787090, 37789164, 37789168, 37789169, 37820578, 37823439, 37826531, 37826532, 37832473, 37832474, 37833962, 37833963, 37834751, 37844023;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98237).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page