Patient Monitor recallZ-1154-2026 · Edan Diagnostics · initiated Oct 24, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edan Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
FDA root cause category: Device Design
What the company is doing
On 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructed to: 1. Immediately examine their inventory and quarantine product subject to correction. 2. If customers may have further distributed affected product, identify their customers and notify them at once of this product correction. The notification to their customers may be enhanced by including the enclosed URGENT MEDICAL DEVICE CORRECTION. 3. For affected product still in their inventory, place copy of the enclosed Notice Before Use and Corrigenda of User Manual into its package before distribution. 4. For affected product that has been distributed or is in use: (1) Confirm that they have received and acknowledged Corrigenda of User Manual. Print out a copy, attach to the User Manual. (2) Confirm the publicly routable IP address 202.114.4.* is NOT used in the User Maintain in the device. If 202.114.4.* is being used, change to appropriate IP settings of device in the Network Setup according to Instruction Instructions for IP Settings Configuration. 5. Complete the attached Acknowledgement and Receipt Form, sign it, and return a scanned copy to the sender of this notification by email. For questions - Contact: EDAN Instruments, Inc. Monday to Friday 09:00-17:00 (UTC+8) Tel:+86-0755-26898326 EDAN Diagnostics, Inc. Monday to Friday 09:00-17:00 (PST) Tel: 1-(858)-750-3066
Recall details
- Recall number
- Z-1154-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 24, 2025
- Posted by the FDA
- Jan 22, 2026
- Recalling firm
- Edan Diagnostics
- Quantity
- 3141
- Product code
- MHX
- FDA submissions
- K113653
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
Product description
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM8, iM8A, iM8B Product Description: iM8 Series Patient Monitor can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. The patient monitor supports software upgrade online and networking and build-in battery power is available for all the models. iM8 Series Patient Monitor can monitor physiological parameters including SpO2, NIBP, ECG, RESP, TEMP, CO2, IBP. The above is the maximum configuration, the user may select different monitoring parameters in according with the requirement. iM8 Series patient monitor includes three models iM8, iM8A, iM8B, and screen size is the primary difference for three models. Component: No
Code information (models, lots, serial numbers)
Show the full code information (63k characters)
EDAN Model: iM8 UDI-DI code: 06944413801592 Serial/Lot Numbers: 260645-M15C07240016 260645-M15C07240001 260645-M15C07240004 260645-M15C07240005 260645-M15C07240017 260645-M15C07240006 260645-M15C07240007 260645-M15C07240008 260645-M15C07240015 260645-M15C07240009 260645-M15C07240018 260645-M15C07240019 260645-M15C07240020 260645-M15C07240010 260645-M15C07240011 260645-M15C07240012 260645-M15C07240013 260645-M15C07240014 260645-M15C07240002 260645-M15C07240003 260645-M16200530008 260645-M16200530007 260645-M16200530006 260645-M16200530005 260645-M16200530001 260645-M16200530003 260645-M16200530002 260645-M16200530004 260605-M16304610008 260605-M16304610010 260605-M16304610007 260605-M16304610009 260605-M16304610006 260605-M16304610001 260605-M16304610004 260605-M16304610003 260605-M16304610002 260605-M16304610005 260605-K16300150005 260605-K16300150006 260605-K16300150007 260605-K16300150008 260605-K16300150009 260605-K16300150010 260605-K16300150004 260645-M16203940001 260645-M16203940003 260645-M16203940005 260645-M16203940006 260645-M16203940007 260645-M16203940008 260645-M16203940002 260605-K16300150003 260605-K16300150002 260605-K16300150001 260645-M16203940009 260645-M16203940010 260645-M16203940004 260645-M16203940012 260645-M16203940011 260605-M16313860011 260605-M16313860018 260605-M16313860017 260605-M16313860016 260605-M16313860019 260605-M16313860015 260605-M16313860014 260605-M16313860013 260605-M16313860012 260605-M16313860010 260605-M16313860005 260605-M16313860008 260605-M16313860007 260605-M16313860006 260605-M16313860020 260605-M16313860004 260605-M16313860003 260605-M16313860002 260605-M16313860001 260605-M16313860009 260605-M16314840022 260605-M16314840023 260605-M16314840024 260605-M16314840025 260605-M16314840026 260605-M16314840027 260605-M16314840030 260605-M16314840029 260605-M16314840021 260645-M16314850001 260645-M16314850002 260645-M16314850003 260645-M16314850004 260645-M16314850005 260605-M16314840028 260605-M16314840020 260605-M16314840016 260605-M16314840018 260605-M16314840001 260605-M16314840002 260605-M16314840003 260605-M16314840004 260605-M16314840005 260605-M16314840006 260605-M16314840019 260605-M16314840007 260605-M16314840009 260605-M16314840011 260605-M16314840013 260605-M16314840014 260605-M16314840015 260605-M16314840017 260605-M16314840008 260605-M16314840010 260605-M16314840012 260605-M16406990019 260605-M16406990018 260605-M16406990016 260605-M16406990013 260605-M16406990015 260605-M16406990014 260605-M16406990017 260605-M16406990020 260645-M16406980017 260645-M16406980002 260645-M16406980020 260645-M16406980019 260645-M16406980018 260605-M16406990012 260645-M16406980016 260645-M16406980015 260645-M16406980014 260645-M16406980013 260645-M16406980001 260645-M16406980012 260645-M16406980010 260645-M16406980009 260645-M16406980008 260645-M16406980007 260645-M16406980006 260645-M16406980005 260645-M16406980004 260645-M16406980003 260645-M16406980011 260605-M16406990011 260605-M16406990004 260605-M16406990009 260605-M16406990001 260605-M16406990002 260605-M16406990003 260605-M16406990005 260605-M16406990006 260605-M16406990007 260605-M16406990008 260605-M16406990010 260645-M16407130030 260645-M16407130028 260645-M16407130026 260645-M16407130025 260645-M16407130024 260645-M16407130023 260645-M16407130022 260645-M16407130021 260645-M16407130020 260645-M16407130019 260645-M16407130018 260645-M16407130027 260645-M16407130017 260645-M16407130016 260645-M16407130029 260645-M16407130014 260605-M16407120010 260605-M16407120009 260605-M16407120008 260605-M16407120007 260605-M16407120006 260605-M16407120005 260605-M16407120004 260605-M16407120003 260605-M16407120002 260645-M16407130015 260605-M16407120011 260605-M16407120012 260605-M16407120001 260605-M16407120014 260605-M16407120013 260645-M16407130013 260645-M16407130012 260645-M16407130011 260645-M16407130010 260645-M16407130009 260645-M16407130007 260645-M16407130006 260645-M16407130005 260645-M16407130004 260645-M16407130008 260645-M16407130003 260645-M16407130002 260645-M16407130001 260605-M16407120020 260605-M16407120019 260605-M16407120018 260605-M16407120017 260605-M16407120016 260605-M16407120015 260645-M16810740010 260645-M16810740009 260645-M16810740008 260645-M16810740007 260645-M16810740006 260645-M16810740005 260645-M16810740004 260645-M16810740001 260645-M16810740002 260645-M16810740003 260645-M16B08640010 260645-M16B08640009 260645-M16B08640008 260645-M16B08640007 260645-M16B08640006 260645-M16B08640005 260645-M16B08640004 260645-M16B08640003 260645-M16B08640002 260645-M16B08640001 260605-M16B08630005 260605-M16B08630004 260605-M16B08630003 260605-M16B08630002 260605-M16B08630001 260605-M16C02510002 260605-M16C02510001 260605-M16B08650001 260605-M16B08650002 260605-M16B08650003 260605-M16B08650004 260605-M16B08650005 260645-M16B08660001 260645-M16B08660002 260645-M16B08660003 260645-M16B08660004 260645-M16B08660005 260645-M16B08660007 260645-M16B08660008 260645-M16B08660009 260645-M16B08660010 260645-M16B08660006 260645-M17201890014 260605-M17201880005 260605-M17201880006 260605-M17201880009 260605-M17201880007 260605-M17201880014 260645-M17201890001 260605-M17201880002 260605-M17201880008 260645-M17201890003 260605-M17201880013 260605-M17201880010 260645-M17201890005 260605-M17201880012 260605-M17201880015 260645-M17201890002 260645-M17201890004 260645-M17201890013 260605-M17201880001 260605-M17201880011 260645-M17201890007 260605-M17201880003 260605-M17201880004 260645-M17201890010 260645-M17201890012 260645-M17201890006 260645-M17201890009 260645-M17201890011 260645-M17201890015 260645-M17201890008 260605-M17302780001 260605-M17302780010 260605-M17302780007 260605-M17302780003 260605-M17302780004 260605-M17302780008 260605-M17302780005 260605-M17302780002 260605-M17302780006 260605-M17302780009 260605-M17304380015 260605-M17304380008 260605-M17304380014 260605-M17304380007 260605-M17304380013 260605-M17304380010 260605-M17304380003 260605-M17304380001 260605-M17304380011 260605-M17304380002 260605-M17304380004 260605-M17304380009 260605-M17304380006 260605-M17304380005 260605-M17304380012 260605-M17313470012 260605-M17313470001 260605-M17313470006 260605-M17313470008 260605-M17313470010 260605-M17313470014 260605-M17313470020 260605-M17313470004 260605-M17313470011 260605-M17313470016 260605-M17313470018 260605-M17313470019 260605-M17313470007 260605-M17313470009 260605-M17313470013 260605-M17313470015 260605-M17313470017 260605-M17313470005 260605-M17313470003 260605-M17313470002 260605-M17604510020 260645-M17604520012 260605-M17604510015 260605-M17604510009 260645-M17604520023 260605-M17604510019 260605-M17604510005 260605-M17604510003 260645-M17604520018 260645-M17604520020 260605-M17604510011 260645-M17604520022 260605-M17604510008 260605-M17604510004 260645-M17604520019 260645-M17604520013 260605-M17604510013 260645-M17604520016 260645-M17604520017 260645-M17604520003 260645-M17604520001 260605-M17606930008 260645-M17604520009 260645-M17604520015 260645-M17606920001 260605-M17604510002 260645-M17606920009 260645-M17606920007 260605-M17604510012 260645-M17604520010 260645-M17604520007 260645-M17604520006 260645-M17604520005 260605-M17606930001 260605-M17604510017 260645-M17606920010 260645-M17606920008 260645-M17606920006 260645-M17606920004 260605-M17606930003 260605-M17604510001 260605-M17604510007 260605-M17606930007 260645-M17604520002 260645-M17604520021 260645-M17606920003 260605-M17604510016 260605-M17606930002 260605-M17606930005 260605-M17606930006 260645-M17604520004 260645-M17604520024 260645-M17604520014 260645-M17604520008 260605-M17606930010 260605-M17604510010 260605-M17604510014 260605-M17604510018 260645-M17606920002 260645-M17606920005 260605-M17606930004 260605-M17606930009 260645-M17604520011 260605-M17604510006 260605-M17615160004 260605-M17615160005 260605-M17615160003 260605-M17615160001 260605-M17615160002 260605-M17708740029 260605-M17708740021 260605-M17708740016 260605-M17708740012 260605-M17708740011 260605-M17708740026 260605-M17708740025 260605-M17708740024 260605-M17708740014 260605-M17708740009 260605-M17708740008 260605-M17708740006 260605-M17708740022 260605-M17708740002 260605-M17708740001 260605-M17708740015 260605-M17708740010 260605-M17708740003 260605-M17708740005 260605-M17708740004 260605-M17708740030 260605-M17708740028 260605-M17708740027 260605-M17708740018 260605-M17708740019 260605-M17708740020 260605-M17708740023 260605-M17708740007 260605-M17708740017 260605-M17708740013 260605-M17806510011 260605-M17806510009 260605-M17806510006 260605-M17806510003 260605-M17806510010 260605-M17806510008 260605-M17806510007 260605-M17806510013 260605-M17806510012 260605-M17806510001 260605-M17806510015 260605-M17806510014 260605-M17806510005 260605-M17806510004 260605-M17806510002 260605-M17807710001 260605-M17807710020 260605-M17807710023 260605-M17807710025 260605-M17807710013 260645-M17807720005 260605-M17807710011 260605-M17807710005 260605-M17807710006 260645-M17807720003 260605-M17807710015 260605-M17807710016 260605-M17807710010 260605-M17807710003 260605-M17807710007 260605-M17807710017 260605-M17807710018 260605-M17807710014 260605-M17807710009 260645-M17807720002 260645-M17807720001 260605-M17807710019 260605-M17807710008 260605-M17807710004 260605-M17807710022 260605-M17807710012 260605-M17807710002 260645-M17807720004 260605-M17807710024 260605-M17807710021 260605-M17811100004 260605-M17811100017 260645-M17811110010 260605-M17811100002 260605-M17811100001 260645-M17811110007 260605-M17811100020 260605-M17811100016 260605-M17811100010 260605-M17811100005 260645-M17811110002 260645-M17811110009 260605-M17811100007 260605-M17811100003 260605-M17811100008 260645-M17811110008 260645-M17811110005 260645-M17811110001 260605-M17811100012 260645-M17811110006 260645-M17811110004 260645-M17811110003 260605-M17811100019 260605-M17811100006 260605-M17811100009 260605-M17811100015 260605-M17811100011 260605-M17811100018 260605-M17811100014 260605-M17811100013 260645-M17A06010012 260645-M17A06010002 260645-M17A06010005 260645-M17A06010011 260645-M17A06010006 260645-M17A06010010 260645-M17A06010007 260645-M17A06010004 260645-M17A06010001 260645-M17A06010003 260645-M17A06010008 260645-M17A06010009 260645-M17B07050012 260605-M17B07040011 260645-M17B07050011 260605-M17B07040018 260605-M17B07040017 260605-M17B07040012 260605-M17B07040003 260645-M17B07050007 260645-M17B07050005 260645-M17B07050003 260605-M17B07040013 260605-M17B07040005 260645-M17B07050013 260645-M17B07050010 260605-M17B07040010 260605-M17B07040019 260605-M17B07040006 260645-M17B07050015 260645-M17B07050001 260605-M17B07040020 260605-M17B07040016 260605-M17B07040004 260605-M17B07040009 260645-M17B07050004 260645-M17B07050006 260605-M17B07040008 260605-M17B07040007 260605-M17B07040002 260645-M17B07050014 260645-M17B07050002 260605-M17B07040014 260605-M17B07040001 260645-M17B07050009 260645-M17B07050008 260605-M17B07040015 260645-M17C11480016 260645-M17C11480001 260645-M17C11480005 260645-M17C11480020 260645-M17C11480025 260645-M17C11480012 260645-M17C11480023 260605-M17C11490001 260605-M17C11490006 260605-M17C11490008 260605-M17C11490011 260605-M17C11490018 260645-M17C11480022 260645-M17C11480013 260605-M17C11490017 260645-M17C11480003 260645-M17C11480010 260645-M17C11480011 260645-M17C11480015 260645-M17C11480017 260645-M17C11480021 260645-M17C11480019 260605-M17C11490003 260605-M17C11490004 260605-M17C11490013 260645-M17C11480007 260645-M17C11480002 260645-M17C11480026 260605-M17C11490020 260645-M17C11480024 260605-M17C11490005 260605-M17C11490007 260605-M17C11490019 260645-M17C11480004 260645-M17C11480006 260645-M17C11480014 260605-M17C11490015 260645-M17C11480008 260645-M17C11480018 260645-M17C11480009 260605-M17C11490016 260605-M17C11490002 260605-M17C11490012 260605-M17C11490014 260605-M17C11490010 260605-M17C11490009 260605-M18108320014 260605-M18108320005 260605-M18108320004 260605-M18108320002 260605-M18108320015 260645-M18108330001 260605-M18108320019 260605-M18108320016 260605-M18108320008 260605-M18108320007 260605-M18108320006 260645-M18108330010 260605-M18108320011 260605-M18108320020 260645-M18108330009 260645-M18108330002 260605-M18108320009 260645-M18108330003 260605-M18108320013 260645-M18108330007 260605-M18108320003 260645-M18108330006 260645-M18108330005 260605-M18108320001 260605-M18108320010 260605-M18108320017 260605-M18108320018 260645-M18108330004 260605-M18108320012 260645-M18108330008 260605-M18306210013 260605-M18306210020 260605-M18306210030 260605-M18306210002 260605-M18306210006 260605-M18306210009 260605-M18306210039 260605-M18306210037 260605-M18306210032 260605-M18306210008 260605-M18306210004 260605-M18306210005 260605-M18306210010 260605-M18306210001 260605-M18306210014 260605-M18306210015 260605-M18306210024 260605-M18306210025 260605-M18306210031 260605-M18306210036 260605-M18306210038 260605-M18306210003 260605-M18306210007 260605-M18306210021 260605-M18306210028 260605-M18306210034 260605-M18306210033 260605-M18306210040 260605-M18306210018 260605-M18306210029 260605-M18306210035 260605-M18306210022 260605-M18306210016 260605-M18306210011 260605-M18306210027 260605-M18306210019 260605-M18306210023 260605-M18306210026 260605-M18306210012 260605-M18306210017 260645-M18307660007 260645-M18307660009 260645-M18307660004 260645-M18307660005 260645-M18307660001 260645-M18307660006 260645-M18307660008 260645-M18307660010 260645-M18307660002 260645-M18307660003 260605-M18314150005 260605-M18314150007 260605-M18314150008 260605-M18314150002 260605-M18314150004 260605-M18314150009 260605-M18314150012 260645-M18314160003 260645-M18314160002 260605-M18314150001 260605-M18314150011 260605-M18314150003 260605-M18314150006 260645-M18314160001 260645-M18314160004 260605-M18314150010 260605-M18407580011 260605-M18407580002 260605-M18407580014 260605-M18407580013 260605-M18407580016 260605-M18407580001 260605-M18407580012 260605-M18407580018 260605-M18407580017 260605-M18407580015 260605-M18407580008 260605-M18407580019 260605-M18407580020 260605-M18407580006 260605-M18407580007 260605-M18407580009 260605-M18407580004 260605-M18407580010 260605-M18407580003 260605-M18407580005 260605-M18508680003 260645-M18508690004 260605-M18508680010 260605-M18508680004 260645-M18508690005 260605-M18508680006 260605-M18508680002 260605-M18508680001 260605-M18508680007 260605-M18508680008 260605-M18508680009 260605-M18508680005 260645-M18508690001 260645-M18508690002 260645-M18508690003 260605-M18606370040 260605-M18606370010 260605-M18606370005 260645-M18606380009 260605-M18606370036 260605-M18606370029 260605-M18606370004 260605-M18606370003 260645-M18606380006 260605-M18606370016 260605-M18606370009 260605-M18606370034 260605-M18606370013 260605-M18606370023 260605-M18606370024 260605-M18606370031 260645-M18606380002 260605-M18606370007 260605-M18606370008 260605-M18606370014 260605-M18606370015 260605-M18606370026 260605-M18606370032 260605-M18606370022 260645-M18606380008 260605-M18606370020 260605-M18606370037 260605-M18606370038 260605-M18606370039 260645-M18606380007 260605-M18606370006 260605-M18606370011 260605-M18606370012 260605-M18606370027 260605-M18606370028 260605-M18606370033 260605-M18606370002 260605-M18606370035 260645-M18606380010 260605-M18606370030 260645-M18606380004 260645-M18606380003 260645-M18606380001 260605-M18606370021 260645-M18606380005 260605-M18606370018 260605-M18606370017 260605-M18606370001 260605-M18606370019 260605-M18606370025 260645-M18613440004 260645-M18613440007 260645-M18613440006 260645-M18613440002 260645-M18612870003 260645-M18612870006 260605-M18613430005 260605-M18613430010 260645-M18613440003 260645-M18613440010 260645-M18613440014 260645-M18613440013 260645-M18613440012 260645-M18613440001 260605-M18613430009 260605-M18613430003 260605-M18613430001 260645-M18612870001 260645-M18613440015 260645-M18612870008 260645-M18613440009 260645-M18613440008 260605-M18613430007 260645-M18612870007 260645-M18612870005 260645-M18612870002 260645-M18613440011 260605-M18613430006 260605-M18613430002 260645-M18613440005 260605-M18613430008 260645-M18612870004 260605-M18613430004 260605-M18800870007 260645-M18800860001 260645-M18800860003 260645-M18800860010 260605-M18800870004 260605-M18800870006 260605-M18800870008 260605-M18800870009 260645-M18800860004 260645-M18800860005 260645-M18800860019 260605-M18800870019 260605-M18800870016 260605-M18800870018 260605-M18800870010 260605-M18800870011 260605-M18800870012 260645-M18800860017 260605-M18800870015 260605-M18800870020 260645-M18800860002 260645-M18800860013 260645-M18800860014 260645-M18800860015 260605-M18800870017 260645-M18800860011 260645-M18800860016 260645-M18800860020 260645-M18800860006 260605-M18800870013 260605-M18800870014 260645-M18800860009 260645-M18800860018 260605-M18800870003 260605-M18800870001 260605-M18800870005 260645-M18800860008 260645-M18800860012 260605-M18800870002 260645-M18800860007 260605-M18901320002 260645-M18901330003 260645-M18901330004 260645-M18901330006 260645-M18901330017 260645-M18901330016 260645-M18901330014 260645-M18901330013 260605-M18901320001 260645-M18901330009 260645-M18901330010 260645-M18901330011 260645-M18901330019 260645-M18901330005 260645-M18901330018 260605-M18901320005 260645-M18901330007 260645-M18901330020 260605-M18901320003 260605-M18901320004 260605-M18901320006 260645-M18901330008 260645-M18901330015 260605-M18901320007 260605-M18901320008 260605-M18901320009 260645-M18901330001 260645-M18901330002 260645-M18901330012 260605-M18901320010 260605-M18A10580007 260605-M18A10580008 260645-M18A10590002 260645-M18A10590007 260605-M18A10580002 260645-M18A10590006 260645-M18A10590003 260645-M18A10590004 260605-M18A10580005 260645-M18A10590008 260605-M18A10580006 260645-M18A10590009 260605-M18A10580009 260605-M18A10580001 260645-M18A10590001 260605-M18A10580003 260605-M18A10580004 260645-M18A10590010 260645-M18A10590005 260605-M18A10580010 260605-M18C07410006 260605-M18C07410012 260645-M18C07420001 260645-M18C07420005 260605-M18C07410001 260645-M18C07420020 260645-M18C07420018 260645-M18C07420013 260645-M18C07420007 260605-M18C07410018 260605-M18C07410014 260605-M18C07410005 260645-M18C07420022 260645-M18C07420012 260645-M18C07420008 260645-M18C07420006 260605-M18C07410010 260645-M18C07420016 260645-M18C07420014 260645-M18C07420009 260605-M18C07410008 260645-M18C07420002 260605-M18C07410021 260605-M18C07410013 260605-M18C07410009 260605-M18C07410007 260645-M18C07420021 260645-M18C07420019 260645-M18C07420003 260605-M18C07410020 260605-M18C07410004 260645-M18C07420017 260645-M18C07420015 260645-M18C07420011 260605-M18C07410016 260605-M18C07410015 260605-M18C07410011 260605-M18C07410002 260645-M18C07420010 260605-M18C07410022 260605-M18C07410019 260605-M18C07410017 260605-M18C07410003 260645-M18C07420004 260605-M19302000016 260605-M19302000019 260605-M19302000003 260605-M19302000004 260605-M19302000008 260605-M19302000005 260605-M19302000007 260605-M19302000010 260605-M19302000014 260605-M19302000009 260605-M19302000015 260605-M19302000002 260605-M19302000018 260605-M19302000012 260605-M19302000013 260605-M19302000006 260605-M19302000001 260605-M19302000020 260605-M19302000011 260605-M19302000017 260605-M19403140010 260605-M19403140002 260605-M19403140012 260605-M19403140006 260605-M19403140003 260605-M19403140009 260645-M19403150004 260645-M19403150002 260645-M19403150001 260605-M19403140016 260605-M19403140018 260605-M19403140013 260645-M19403150008 260605-M19403140014 260605-M19403140008 260645-M19403150005 260645-M19403150003 260605-M19403140015 260605-M19403140001 260605-M19403140020 260605-M19403140019 260605-M19403140005 260605-M19403140007 260645-M19403150006 260645-M19403150009 260605-M19403140004 260645-M19403150010 260645-M19403150007 260605-M19403140017 260605-M19403140011 260645-M19508900009 260645-M19508900004 260645-M19508900010 260645-M19508900012 260645-M19508900003 260645-M19508900005 260645-M19508900007 260645-M19508900002 260645-M19508900006 260645-M19508900008 260645-M19508900011 260645-M19508900014 260645-M19508900013 260645-M19508900015 260645-M19508900001 260605-M19612590014 260605-M19612590022 260605-M19612590024 260605-M19612590030 260645-M19612600014 260605-M19612590012 260605-M19612590029 260645-M19612600006 260645-M19612600008 260645-M19612600013 260605-M19612590002 260605-M19612590004 260645-M19612600012 260645-M19612600004 260605-M19612590011 260645-M19612600010 260645-M19612600005 260605-M19612590025 260605-M19612590027 260645-M19612600001 260645-M19612600007 260645-M19612600009 260605-M19612590001 260605-M19612590005 260605-M19612590007 260605-M19612590026 260645-M19612600002 260645-M19612600011 260605-M19612590008 260605-M19612590016 260605-M19612590018 260605-M19612590028 260605-M19612590009 260645-M19612600003 260605-M19612590021 260605-M19612590020 260605-M19612590019 260605-M19612590015 260605-M19612590017 260605-M19612590003 260605-M19612590023 260645-M19612600015 260605-M19612590006 260605-M19612590010 260605-M19612590013 260645-M19809850013 260645-M19809850014 260645-M19809850017 260605-M19809840003 260605-M19809840007 260605-M19809840010 260605-M19809840013 260605-M19809840018 260605-M19809840004 260645-M19809850001 260645-M19809850002 260645-M19809850005 260645-M19809850010 260645-M19809850011 260645-M19809850015 260645-M19809850016 260645-M19809850019 260645-M19809850012 260605-M19809840005 260645-M19809850004 260605-M19809840014 260605-M19809840020 260605-M19809840012 260605-M19809840009 260605-M19809840006 260605-M19809840008 260605-M19809840016 260605-M19809840015 260605-M19809840002 260605-M19809840019 260605-M19809840017 260645-M19809850006 260645-M19809850008 260645-M19809850009 260605-M19809840011 260645-M19809850003 260645-M19809850007 260605-M19809840001 260645-M19809850020 260645-M19809850018 260605-M19C01620028 260605-M19C01620023 260605-M19C01620021 260605-M19C01620016 260605-M19C01620005 260605-M19C01620039 260605-M19C01620020 260605-M19C01620031 260605-M19C01620018 260605-M19C01620025 260605-M19C01620019 260605-M19C01620007 260605-M19C01620006 260605-M19C01620034 260605-M19C01620032 260605-M19C01620038 260605-M19C01620024 260605-M19C01620033 260605-M19C01620004 260605-M19C01620037 260605-M19C01620036 260605-M19C01620013 260605-M19C01620011 260605-M19C01620008 260605-M19C01620027 260605-M19C01620002 260605-M19C01620026 260605-M19C01620012 260605-M19C01620009 260605-M19C01620040 260605-M19C01620030 260605-M19C01620029 260605-M19C01620010 260605-M19C01620014 260605-M19C01620022 260605-M19C01620017 260605-M19C01620015 260605-M19C01620035 260605-M19C01620001 260605-M19C01620003 260605-M20206350005 260645-M20206370004 260605-M20206350009 260605-M20206350001 260645-M20206370010 260645-M20206370003 260605-M20206350004 260605-M20206350003 260645-M20206370009 260645-M20206370005 260605-M20206350008 260645-M20206370006 260605-M20206350002 260605-M20206350010 260645-M20206370001 260645-M20206370007 260645-M20206370002 260605-M20206350007 260605-M20206350006 260645-M20206370008 260605-M20310140002 260645-M20310150011 260645-M20310150015 260605-M20310140003 260605-M20310140001 260645-M20310150025 260645-M20310150019 260645-M20310150018 260645-M20310150016 260645-M20310150010 260645-M20310150005 260605-M20310140008 260605-M20310140006 260605-M20310140004 260645-M20310150021 260605-M20310140007 260645-M20310150001 260645-M20310150014 260605-M20310140010 260605-M20310140005 260645-M20310150022 260645-M20310150020 260645-M20310150007 260645-M20310150002 260645-M20310150024 260645-M20310150023 260645-M20310150009 260645-M20310150008 260645-M20310150006 260645-M20310150004 260605-M20310140009 260645-M20310150017 260645-M20310150003 260645-M20310150012 260645-M20310150013 260605-M20317900019 260605-M20317900001 260605-M20317900002 260605-M20317900007 260605-M20317900016 260605-M20317900012 260605-M20317900011 260605-M20317900010 260605-M20317900018 260605-M20317900015 260605-M20317900014 260605-M20317900008 260605-M20317900005 260605-M20317900004 260605-M20317900003 260605-M20317900013 260605-M20317900020 260605-M20317900009 260605-M20317900017 260605-M20317900006 260645-M20317910019 260645-M20317910020 260645-M20317910011 260645-M20317910004 260645-M20317910009 260645-M20317910003 260645-M20317910013 260645-M20317910012 260645-M20317910007 260645-M20317910015 260645-M20317910016 260645-M20317910002 260645-M20317910014 260645-M20317910010 260645-M20317910008 260645-M20317910005 260645-M20317910017 260645-M20317910001 260645-M20317910018 260645-M20317910006 260605-M20411630018 260605-M20403670007 260605-M20411620030 260605-M20403670018 260605-M20411620007 260605-M20411660004 260605-M20403670016 260605-M20411630025 260605-M20403670006 260605-M20411620029 260605-M20411660006 260605-M20411660002 260605-M20411660008 260605-M20411620028 260605-M20411620026 260605-M20411620016 260605-M20411630009 260605-M20411660010 260605-M20403670002 260605-M20411620013 260605-M20411630026 260605-M20403670021 260605-M20411630013 260605-M20411620006 260605-M20411620009 260605-M20411620011 260605-M20411620014 260605-M20411620020 260605-M20403670012 260605-M20403670025 260605-M20411630017 260605-M20411630023 260605-M20411630024 260605-M20411620002 260605-M20411620012 260605-M20403670023 260605-M20411630003 260605-M20411630004 260605-M20411630030 260605-M20403670010 260605-M20403670015 260605-M20403670005 260605-M20411620004 260605-M20411620003 260605-M20411620001 260605-M20411630016 260605-M20411630011 260605-M20411630008 260605-M20411630002 260605-M20411630014 260605-M20403670028 260605-M20411620010 260605-M20403670022 260605-M20403670014 260605-M20411660005 260605-M20411620022 260605-M20411620019 260605-M20411620018 260605-M20411620005 260605-M20411630028 260605-M20411630020 260605-M20411630007 260605-M20403670017 260605-M20411620025 260605-M20411660001 260605-M20403670019 260605-M20403670001 260605-M20411660003 260605-M20411620015 260605-M20411630027 260605-M20411630019 260605-M20411630012 260605-M20411630010 260605-M20411630006 260605-M20411630001 260605-M20411620021 260605-M20403670020 260605-M20403670004 260605-M20411660009 260605-M20411620027 260605-M20411630029 260605-M20411630005 260605-M20411620008 260605-M20403670008 260605-M20403670030 260605-M20403670027 260605-M20403670024 260605-M20403670009 260605-M20403670029 260605-M20411620023 260605-M20403710018 260605-M20403710017 260605-M20413830003 260605-M20403670011 260605-M20411640006 260605-M20411640013 260605-M20411640016 260605-M20411640001 260605-M20403710007 260605-M20411640030 260605-M20403670013 260605-M20403710014 260605-M20403710012 260605-M20411640024 260605-M20411640019 260605-M20411640007 260605-M20413830025 260605-M20403710004 260605-M20411630022 260605-M20403670026 260605-M20411640028 260605-M20411640011 260605-M20403710016 260605-M20403710009 260605-M20403710006 260605-M20403710002 260605-M20413830013 260605-M20411620017 260605-M20413830008 260605-M20411660007 260605-M20413830050 260605-M20413830045 260605-M20413830035 260605-M20413830034 260605-M20413830021 260605-M20413830006 260605-M20413830007 260605-M20413830049 260605-M20413830047 260605-M20413830046 260605-M20413830040 260605-M20413830038 260605-M20413830037 260605-M20411630021 260605-M20413830009 260605-M20413830048 260605-M20413830001 260605-M20413830042 260605-M20413830024 260605-M20411640010 260605-M20403710019 260605-M20403710003 260605-M20403710001 260605-M20413830014 260605-M20411640020 260605-M20411640017 260605-M20411640014 260605-M20411640005 260605-M20403710010 260605-M20411640029 260605-M20411640012 260605-M20411640009 260605-M20413830028 260605-M20411640027 260605-M20411640022 260605-M20413830041 260605-M20403710015 260605-M20411640021 260605-M20413830002 260605-M20411630015 260605-M20403670003 260605-M20411640023 260605-M20411640015 260605-M20411640002 260605-M20411640025 260605-M20411640003 260605-M20411640008 260605-M20411640004 260605-M20411640026 260605-M20403710020 260605-M20403710013 260605-M20403710011 260605-M20403710008 260605-M20413830012 260605-M20403710005 260605-M20413830010 260605-M20413830016 260605-M20413830017 260605-M20413830030 260605-M20413830020 260605-M20413830011 260605-M20413830005 260605-M20413830004 260605-M20413830044 260605-M20413830043 260605-M20413830036 260605-M20413830033 260605-M20413830015 260605-M20411620024 260605-M20413830032 260605-M20413830031 260605-M20413830023 260605-M20413830019 260605-M20413830018 260605-M20413830022 260605-M20413830039 260605-M20413830026 260605-M20413830027 260605-M20413830029 260605-M20411640018 260645-M20507030019 260645-M20507030009 260645-M20507030005 260645-M20411650030 260645-M20411650023 260645-M20411650019 260645-M20411650011 260645-M20411650003 260645-M20507030015 260645-M20507030004 260645-M20507030006 260645-M20507030003 260645-M20411650025 260645-M20411650022 260645-M20411650013 260645-M20411650009 260645-M20411650029 260645-M20411650021 260645-M20411650016 260645-M20411650001 260645-M20507030028 260645-M20507030026 260645-M20507030024 260645-M20507030021 260645-M20507030020 260645-M20507030017 260645-M20507030013 260645-M20411650027 260645-M20411650026 260645-M20411650005 260645-M20411650002 260645-M20411650028 260645-M20507030007 260645-M20411650024 260645-M20507030014 260645-M20507030022 260645-M20411650018 260645-M20411650014 260645-M20411650012 260645-M20507030027 260645-M20507030023 260645-M20507030012 260645-M20507030002 260645-M20507030018 260645-M20507030011 260645-M20507030010 260645-M20507030008 260645-M20507030001 260645-M20507030030 260645-M20507030029 260645-M20507030025 260645-M20411650015 260645-M20411650007 260645-M20411650006 260645-M20411650017 260645-M20411650010 260645-M20507030016 260645-M20411650020 260645-M20411650004 260645-M20411650008 260605-M20801500007 260605-M20801500008 260605-M20801500022 260605-M20801500024 260605-M20801500043 260605-M20801500025 260645-M20804320003 260645-M20804320047 260645-M20804320005 260645-M20804320013 260645-M20804320036 260605-M20801500028 260645-M20804320014 260645-M20804320010 260645-M20804320004 260605-M20801500049 260605-M20801500039 260605-M20801500027 260605-M20801500020 260605-M20801500018 260645-M20804320015 260605-M20801500017 260645-M20804320024 260645-M20804320023 260645-M20804320034 260645-M20804320050 260645-M20804320048 260645-M20804320037 260645-M20804320025 260645-M20804320044 260605-M20801500042 260645-M20804320039 260645-M20804320032 260645-M20804320046 260605-M20801500009 260605-M20801500030 260605-M20801500046 260605-M20801500036 260605-M20801500023 260605-M20801500013 260645-M20804320029 260605-M20801500037 260605-M20801500038 260605-M20801500040 260605-M20801500012 260605-M20801500014 260605-M20801500015 260605-M20801500029 260605-M20801500033 260605-M20801500041 260645-M20804320021 260645-M20804320028 260645-M20804320035 260645-M20804320041 260645-M20804320042 260605-M20801500006 260605-M20801500003 260605-M20801500021 260605-M20801500032 260605-M20801500034 260605-M20801500035 260605-M20801500044 260605-M20801500045 260605-M20801500050 260645-M20804320007 260645-M20804320017 260645-M20804320020 260645-M20804320026 260605-M20801500011 260645-M20804320045 260645-M20804320043 260645-M20804320030 260605-M20801500047 260645-M20804320006 260645-M20804320016 260645-M20804320031 260645-M20804320049 260605-M20801500002 260605-M20801500004 260605-M20801500005 260605-M20801500019 260605-M20801500031 260605-M20801500048 260645-M20804320002 260645-M20804320011 260645-M20804320012 260645-M20804320022 260645-M20804320033 260645-M20804320038 260605-M20801500001 260605-M20801500010 260605-M20801500016 260605-M20801500026 260645-M20804320001 260645-M20804320008 260645-M20804320018 260645-M20804320027 260645-M20804320040 260645-M20804320009 260645-M20804320019 260605-M20808580042 260605-M20808580033 260605-M20808580039 260605-M20808580040 260605-M20808580044 260605-M20808580049 260605-M20808580022 260605-M20808580024 260605-M20808580043 260605-M20808580004 260605-M20808580021 260605-M20808580025 260605-M20808580056 260605-M20808580032 260605-M20808580030 260605-M20808580034 260605-M20808580002 260605-M20808580027 260605-M20808580010 260605-M20808580016 260605-M20808580058 260605-M20808580035 260605-M20808580051 260605-M20808580055 260605-M20808580007 260605-M20808580037 260605-M20808580041 260605-M20808580045 260605-M20808580047 260605-M20808580008 260605-M20808580014 260605-M20808580019 260605-M20808580038 260605-M20808580046 260605-M20808580048 260605-M20808580003 260605-M20808580009 260605-M20808580018 260605-M20808580057 260605-M20808580031 260605-M20808580060 260605-M20808580028 260605-M20808580013 260605-M20808580026 260605-M20808580006 260605-M20808580012 260605-M20808580036 260605-M20808580050 260605-M20808580052 260605-M20808580054 260605-M20808580011 260605-M20808580015 260605-M20808580017 260605-M20808580020 260605-M20808580023 260605-M20808580029 260605-M20808580053 260605-M20808580059 260605-M20808580001 260605-M20808580005 360001-M20A11130006 360001-M20A11130003 360001-M20A11130002 360001-M20A11130027 360001-M20A11130025 360001-M20A11130015 360001-M20A13140030 360001-M20A13140022 360001-M20A13140011 360001-M20A11130016 360001-M20A11130012 360001-M20A11130007 360001-M20A11130005 360001-M20A13140021 360001-M20A13140020 360001-M20A13140014 360001-M20A13140012 360001-M20A13140002 360001-M20A11130030 360001-M20A11130021 360001-M20A11130017 360001-M20A13140025 360001-M20A11130009 360001-M20A11130028 360001-M20A11130011 360001-M20A11130018 360001-M20A11130024 360001-M20A13140003 360001-M20A13140007 360001-M20A13140008 360001-M20A13140009 360001-M20A13140017 360001-M20A11130004 360001-M20A11130008 360001-M20A11130010 360001-M20A11130013 360001-M20A11130020 360001-M20A11130022 360001-M20A11130023 360001-M20A11130026 360001-M20A13140004 360001-M20A13140026 360001-M20A13140027 360001-M20A13140028 360001-M20A13140029 360001-M20A13140005 360001-M20A13140013 360001-M20A13140015 360001-M20A13140024 360001-M20A11130001 360001-M20A11130014 360001-M20A11130029 360001-M20A11130019 360001-M20A13140019 360001-M20A13140023 360001-M20A13140018 360001-M20A13140010 360001-M20A13140006 360001-M20A13140001 360001-M20A13140016 260605-M20C13200006 260605-M20C13200024 260605-M20C16040006 260605-M20C13200028 260605-M20C13200003 260605-M20C13200009 260605-M20C13200015 260605-M20C13200025 260605-M20C16040001 260605-M20C16040002 260605-M20C16040007 260605-M20C16040009 260605-M20C13200004 260605-M20C13200008 260605-M20C16040010 260605-M20C13200026 260605-M20C16040014 260605-M20C16040020 260605-M20C13200012 260605-M20C13200017 260605-M20C13200022 260605-M20C13200027 260605-M20C16040011 260605-M20C16040017 260605-M20C16040019 260605-M20C16040008 260605-M20C16040004 260605-M20C13200007 260605-M20C16040013 260605-M20C13200018 260605-M20C13200021 260605-M20C13200030 260605-M20C16040018 260605-M20C13200005 260605-M20C13200019 260605-M20C13200029 260605-M20C13200002 260605-M20C13200014 260605-M20C16040015 260605-M20C13200010 260605-M20C13200013 260605-M20C13200016 260605-M20C13200020 260605-M20C13200023 260605-M20C16040003 260605-M20C16040005 260605-M20C16040012 260605-M20C16040016 260605-M20C13200001 260605-M20C13200011 260605-M20C13190026 260605-M20C13190015 260605-M20C13190006 260605-M20C13190002 260605-M20C13190001 260605-M20C13190070 260605-M20C13190065 260605-M20C13190053 260605-M20C13190047 260605-M20C13190043 260605-M20C13190029 260605-M20C13190017 260605-M20C13190013 260605-M20C13190009 260605-M20C13190067 260605-M20C13190063 260605-M20C13190056 260605-M20C13190052 260605-M20C13190051 260605-M20C13190050 260605-M20C13190046 260605-M20C13190042 260605-M20C13190020 260605-M20C13190021 260605-M20C13190038 260605-M20C13190055 260605-M20C13190037 260605-M20C13190061 260605-M20C13190060 260605-M20C13190059 260605-M20C13190025 260605-M20C13190007 260605-M20C13190003 260605-M20C13190068 260605-M20C13190033 260605-M20C13190024 260605-M20C13190023 260605-M20C13190019 260605-M20C13190014 260605-M20C13190057 260605-M20C13190045 260605-M20C13190004 260605-M20C13190027 260605-M20C13190039 260605-M20C13190044 260605-M20C13190049 260605-M20C13190016 260605-M20C13190012 260605-M20C13190011 260605-M20C13190005 260605-M20C13190066 260605-M20C13190062 260605-M20C13190041 260605-M20C13190040 260605-M20C13190036 260605-M20C13190032 260605-M20C13190028 260605-M20C13190031 260605-M20C13190010 260605-M20C13190008 260605-M20C13190069 260605-M20C13190064 260605-M20C13190048 260605-M20C13190034 260605-M20C13190022 260605-M20C13190018 260605-M20C13190058 260605-M20C13190054 260605-M20C13190030 260605-M20C13190035 260605-M21108540015 260605-M21108540009 260605-M21113090030 260605-M21113090040 260605-M21113090050 260605-M21113090028 260605-M21113090049 260605-M21113090039 260605-M21113090035 260605-M21113090021 260605-M21113090014 260645-M21108550007 260605-M21108540024 260605-M21108540023 260605-M21108540020 260605-M21108540017 260605-M21113090045 260605-M21113090022 260645-M21108550010 260605-M21113090031 260605-M21113090036 260605-M21113090041 260605-M21113090046 260605-M21108540012 260605-M21108540018 260605-M21108540029 260645-M21108550001 260645-M21108550020 260605-M21113090017 260645-M21108550006 260605-M21113090013 260645-M21108550012 260605-M21108540001 260605-M21108540022 260605-M21108540028 260605-M21113090009 260605-M21113090010 260605-M21113090025 260605-M21113090029 260605-M21113090001 260605-M21113090012 260605-M21113090018 260605-M21113090002 260605-M21113090008 260605-M21113090015 260605-M21113090019 260605-M21113090020 260605-M21113090034 260605-M21113090044 260605-M21108540016 260605-M21113090027 260605-M21113090026 260605-M21113090023 260645-M21108550004 260605-M21108540021 260605-M21108540004 260605-M21108540003 260605-M21113090047 260605-M21113090005 260645-M21108550019 260645-M21108550011 260645-M21108550005 260605-M21113090032 260605-M21113090038 260605-M21113090042 260605-M21113090048 260605-M21108540007 260605-M21108540014 260605-M21108540019 260605-M21108540025 260605-M21108540030 260645-M21108550002 260645-M21108550009 260605-M21113090043 260645-M21108550013 260645-M21108550018 260605-M21108540006 260645-M21108550003 260645-M21108550008 260645-M21108550014 260645-M21108550017 260605-M21108540005 260605-M21108540010 260605-M21108540011 260605-M21108540013 260605-M21108540026 260645-M21108550016 260605-M21113090037 260605-M21113090004 260605-M21113090033 260605-M21113090024 260605-M21113090016 260605-M21113090011 260605-M21113090007 260605-M21113090006 260645-M21108550015 260605-M21108540027 260605-M21113090003 260605-M21108540002 260605-M21108540008 360001-M21301990012 360001-M21301990016 360001-M21301990014 360001-M21301990007 360001-M21301990006 360001-M21301990004 360001-M21301990013 360001-M21301990011 360001-M21301990020 360001-M21301990015 360001-M21301990005 360001-M21301990003 360001-M21301990017 360001-M21301990001 360001-M21301990009 360001-M21301990018 360001-M21301990002 360001-M21301990019 360001-M21301990010 360001-M21301990008 360001-M21502500002 360001-M21502500008 360001-M21502500036 360001-M21502500038 360001-M21502510008 360001-M21502510009 360001-M21502510016 360001-M21502510026 360001-M21502500020 360001-M21502500024 360001-M21502500030 360001-M21502500031 360001-M21502500040 360001-M21502510003 360001-M21502510006 360001-M21502510029 360001-M21502500022 360001-M21502510023 360001-M21502510019 360001-M21502510010 360001-M21502510014 360001-M21502510015 360001-M21502510020 360001-M21502500001 360001-M21502500003 360001-M21502500016 360001-M21502500018 360001-M21502500027 360001-M21502510017 360001-M21502510018 360001-M21502500004 360001-M21502500015 360001-M21502500019 360001-M21502500035 360001-M21502500039 360001-M21502510004 360001-M21502500010 360001-M21502500012 360001-M21502500014 360001-M21502500037 360001-M21502510012 360001-M21502510011 360001-M21502510001 360001-M21502500032 360001-M21502500034 360001-M21502500025 360001-M21502500026 360001-M21502500029 360001-M21502510013 360001-M21502510021 360001-M21502510024 360001-M21502510030 360001-M21502500007 360001-M21502500028 360001-M21502510002 360001-M21502510005 360001-M21502510007 360001-M21502500005 360001-M21502500006 360001-M21502500009 360001-M21502500017 360001-M21502500021 360001-M21502500023 360001-M21502500013 360001-M21502500011 360001-M21502500033 360001-M21502510027 360001-M21502510028 360001-M21502510025 360001-M21502510022 360001-M21608830028 360001-M21608830029 360001-M21608830020 360001-M21608820028 360001-M21608830010 360001-M21608830025 360001-M21608830030 360001-M21608820014 360001-M21608820018 360001-M21608820025 360001-M21608820036 360001-M21608820037 360001-M21608830003 360001-M21608830008 360001-M21608830019 360001-M21608830009 360001-M21608830026 360001-M21608820002 360001-M21608820019 360001-M21608820038 360001-M21608820005 360001-M21608820011 360001-M21608820017 360001-M21608820024 360001-M21608820032 360001-M21608820033 360001-M21608830016 360001-M21608820003 360001-M21608830022 360001-M21608830015 360001-M21608820023 360001-M21608830012 360001-M21608820035 360001-M21608830001 360001-M21608830004 360001-M21608830006 360001-M21608830017 360001-M21608830027 360001-M21608820004 360001-M21608820008 360001-M21608820010 360001-M21608820020 360001-M21608820039 360001-M21608830007 360001-M21608830013 360001-M21608830018 360001-M21608830023 360001-M21608820001 360001-M21608820006 360001-M21608820007 360001-M21608820012 360001-M21608820013 360001-M21608820022 360001-M21608830024 360001-M21608830021 360001-M21608830011 360001-M21608830014 360001-M21608820016 360001-M21608820009 360001-M21608820026 360001-M21608820027 360001-M21608820030 360001-M21608830002 360001-M21608830005 360001-M21608820015 360001-M21608820029 360001-M21608820031 360001-M21608820034 360001-M21608820040 360001-M21608820021 360001-M21714530029 360001-M21714530030 360001-M21714300006 360001-M21714300010 360001-M21714300013 360001-M21714300020 360001-M21714530019 360001-M21714530008 360001-M21714530006 360001-M21714530007 360001-M21714300019 360001-M21714530009 360001-M21714300004 360001-M21714300011 360001-M21714530010 360001-M21714530011 360001-M21714530020 360001-M21714530021 360001-M21714300014 360001-M21714300018 360001-M21714300007 360001-M21714300015 360001-M21714300016 360001-M21714530012 360001-M21714530013 360001-M21714530017 360001-M21714530018 360001-M21714530022 360001-M21714530026 360001-M21714530027 360001-M21714530028 360001-M21714300005 360001-M21714300009 360001-M21714300012 360001-M21714530002 360001-M21714530001 360001-M21714300001 360001-M21714300008 360001-M21714530005 360001-M21714530016 360001-M21714300003 360001-M21714530015 360001-M21714530004 360001-M21714530003 360001-M21714530025 360001-M21714300002 360001-M21714530024 360001-M21714530023 360001-M21714530014 360001-M21714300017 360001-M21816690006 360001-M21816690005 360001-M21816690009 360001-M21816690002 360001-M21816690010 360001-M21816690008 360001-M21816690007 360001-M21816690003 360001-M21816690001 360001-M21816690004 360001-M21A04750015 360001-M21A02510003 360001-M21A02510017 360001-M21A04470020 360001-M21A04470022 360001-M21A04470026 360001-M21A04470027 360001-M21A04470028 360001-M21A04470029 360001-M21A04750010 360001-M21A02510007 360001-M21A04750014 360001-M21A02510001 360001-M21A04470021 360001-M21A04470030 360001-M21A02510004 360001-M21A04470025 360001-M21A04470024 360001-M21A04470023 360001-M21A04470010 360001-M21A02510015 360001-M21A02510013 360001-M21A02510006 360001-M21A02510002 360001-M21A04750007 360001-M21A04750003 360001-M21A04750002 360001-M21A02510008 360001-M21A04470004 360001-M21A04470011 360001-M21A04470005 360001-M21A04750012 360001-M21A04470003 360001-M21A02510019 360001-M21A02510010 360001-M21A04750011 360001-M21A04750009 360001-M21A04750005 360001-M21A02510016 360001-M21A04470015 360001-M21A04470014 360001-M21A04470013 360001-M21A04470009 360001-M21A04470012 360001-M21A04470016 360001-M21A04470017 360001-M21A04470018 360001-M21A04470019 360001-M21A04750004 360001-M21A04750006 360001-M21A04750008 360001-M21A02510018 360001-M21A02510012 360001-M21A02510020 360001-M21A02510009 360001-M21A02510011 360001-M21A02510014 360001-M21A04470001 360001-M21A04470002 360001-M21A04470006 360001-M21A04470007 360001-M21A04470008 360001-M21A04750001 360001-M21A02510005 360001-M21A04750013 360001-M21C04420007 360001-M21C04420043 360001-M21C04430010 360001-M21C04430025 360001-M21C04420006 360001-M21C04420005 360001-M21C04420038 360001-M21C04430027 360001-M21C04420010 360001-M21C04420003 360001-M21C04430020 360001-M21C04430007 360001-M21C04420042 360001-M21C04420037 360001-M21C04420036 360001-M21C04420035 360001-M21C04420017 360001-M21C04420018 360001-M21C04420016 360001-M21C04420001 360001-M21C04430015 360001-M21C04420050 360001-M21C04420048 360001-M21C04420021 360001-M21C04430028 360001-M21C04430024 360001-M21C04430016 360001-M21C04430013 360001-M21C04420015 360001-M21C04420024 360001-M21C04420029 360001-M21C04420030 360001-M21C04430021 360001-M21C04430018 360001-M21C04430014 360001-M21C04430006 360001-M21C04430003 360001-M21C04420046 360001-M21C04420045 360001-M21C04420040 360001-M21C04420033 360001-M21C04420028 360001-M21C04420020 360001-M21C04420013 360001-M21C04420002 360001-M21C04430022 360001-M21C04430019 360001-M21C04430017 360001-M21C04430011 360001-M21C04430008 360001-M21C04430004 360001-M21C04420044 360001-M21C04430029 360001-M21C04430002 360001-M21C04430005 360001-M21C04420049 360001-M21C04420047 360001-M21C04420031 360001-M21C04420027 360001-M21C04420026 360001-M21C04420025 360001-M21C04420041 360001-M21C04420032 360001-M21C04420023 360001-M21C04420012 360001-M21C04420009 360001-M21C04420004 360001-M21C04420011 360001-M21C04430030 360001-M21C04430026 360001-M21C04430023 360001-M21C04430012 360001-M21C04430009 360001-M21C04430001 360001-M21C04420039 360001-M21C04420034 360001-M21C04420022 360001-M21C04420019 360001-M21C04420014 360001-M21C04420008 360001-M22105670015 360001-M22105670012 360001-M22105660046 360001-M22105660042 360001-M22105660041 360001-M22105660036 360001-M22105660032 360001-M22105660009 360001-M22105660008 360001-M22105660007 360001-M22105660004 360001-M22105660003 360001-M22105670001 360001-M22105660045 360001-M22105660034 360001-M22105660010 360001-M22105660013 360001-M22105660014 360001-M22105660017 360001-M22105660005 360001-M22105660026 360001-M22105660022 360001-M22105660019 360001-M22105660018 360001-M22105670011 360001-M22105670004 360001-M22105660025 360001-M22105660035 360001-M22105660049 360001-M22105660044 360001-M22105660031 360001-M22105670014 360001-M22105660029 360001-M22105660028 360001-M22105660027 360001-M22105660024 360001-M22105660023 360001-M22105660020 360001-M22105660015 360001-M22105660006 360001-M22105660002 360001-M22105660001 360001-M22105670013 360001-M22105670006 360001-M22105660048 360001-M22105670003 360001-M22105670007 360001-M22105660011 360001-M22105660043 360001-M22105660040 360001-M22105660039 360001-M22105660038 360001-M22105660037 360001-M22105660033 360001-M22105660030 360001-M22105660047 360001-M22105670010 360001-M22105670002 360001-M22105670009 360001-M22105670005 360001-M22105660050 360001-M22105660021 360001-M22105660016 360001-M22105660012 360001-M22105670008 360001-M22405360025 360001-M22405360013 360001-M22405360020 360001-M22405360031 360001-M22405360033 360001-M22405360042 360001-M22405360050 360001-M22405360018 360001-M22405360034 360001-M22405360001 360001-M22405360041 360001-M22405360002 360001-M22405360008 360001-M22405360015 360001-M22405360019 360001-M22405360022 360001-M22405360028 360001-M22405360037 360001-M22405360046 360001-M22405360036 360001-M22405360010 360001-M22405360048 360001-M22405360003 360001-M22405360004 360001-M22405360006 360001-M22405360011 360001-M22405360024 360001-M22405360026 360001-M22405360030 360001-M22405360005 360001-M22405360012 360001-M22405360016 360001-M22405360035 360001-M22405360040 360001-M22405360043 360001-M22405360009 360001-M22405360014 360001-M22405360021 360001-M22405360023 360001-M22405360047 360001-M22405360007 360001-M22405360032 360001-M22405360038 360001-M22405360045 360001-M22405360049 360001-M22405360017 360001-M22405360027 360001-M22405360039 360001-M22405360044 360001-M22405360029 360001-M22511270003 360001-M22511270031 360001-M22511270040 360001-M22511270037 360001-M22511270039 360001-M22511270007 360001-M22511270009 360001-M22511270017 360001-M22511270028 360001-M22511270030 360001-M22511270035 360001-M22511270036 360001-M22511270038 360001-M22511270026 360001-M22511270024 360001-M22511270020 360001-M22511270014 360001-M22511270004 360001-M22511270033 360001-M22511270032 360001-M22511270018 360001-M22511270008 360001-M22511270005 360001-M22511270006 360001-M22511270010 360001-M22511270034 360001-M22511270029 360001-M22511270027 360001-M22511270021 360001-M22511270019 360001-M22511270011 360001-M22511270001 360001-M22511270025 360001-M22511270023 360001-M22511270022 360001-M22511270016 360001-M22511270015 360001-M22511270013 360001-M22511270012 360001-M22511270002 360001-M22615420002 360001-M22615420007 360001-M22615420003 360001-M22615420004 360001-M22615420001 360001-M22615420005 360001-M22615420008 360001-M22615420010 360001-M22615420006 360001-M22615420009 360001-M22808130026 360001-M22808130006 360001-M22808130014 360001-M22808130001 360001-M22808130027 360001-M22808130023 360001-M22808130020 360001-M22808130018 360001-M22808130017 360001-M22808130013 360001-M22808130010 360001-M22808130028 360001-M22808130022 360001-M22808130019 360001-M22808130015 360001-M22808130041 360001-M22808130031 360001-M22808130029 360001-M22808130025 360001-M22808130024 360001-M22808130012 360001-M22808130016 360001-M22808130036 360001-M22808130046 360001-M22808130044 360001-M22808130040 360001-M22808130037 360001-M22808130033 360001-M22808130030 360001-M22808130008 360001-M22808130002 360001-M22808130050 360001-M22808130038 360001-M22808130042 360001-M22808130032 360001-M22808130007 360001-M22808130003 360001-M22808130049 360001-M22808130045 360001-M22808130039 360001-M22808130035 360001-M22808130034 360001-M22808130021 360001-M22808130011 360001-M22808130043 360001-M22808130047 360001-M22808130048 360001-M22808130009 360001-M22808130005 360001-M22808130004 360001-M22C13200010 360001-M22C07420003 360001-M22C13200003 360001-M22C13200004 360001-M22C07420010 360001-M22C13200005 360001-M22C07420009 360001-M22C13200006 360001-M22C07420007 360001-M22C13200007 360001-M22C13200008 360001-M22C13200009 360001-M22C07420002 360001-M22C07420006 360001-M22C13200001 360001-M22C07420004 360001-M22C07420001 360001-M22C13200002 360001-M22C07420005 360001-M22C07420008 360001-M23226930011 360001-M23226930005 360001-M23226930004 360001-M23226920005 360001-M23226920013 360001-M23226920001 360001-M23226930002 360001-M23226920004 360001-M23226920007 360001-M23226920009 360001-M23226920018 360001-M23226930006 360001-M23226930007 360001-M23226930021 360001-M23226920008 360001-M23226920014 360001-M23226920017 360001-M23226920019 360001-M23226930001 360001-M23226930016 360001-M23226930017 360001-M23226930020 360001-M23226930026 360001-M23226930027 360001-M23226920015 360001-M23226920010 360001-M23226930008 360001-M23226930019 360001-M23226930029 360001-M23226920003 360001-M23226920011 360001-M23226930010 360001-M23226930012 360001-M23226930014 360001-M23226930015 360001-M23226930022 360001-M23226930024 360001-M23226930025 360001-M23226930030 360001-M23226920002 360001-M23226930009 360001-M23226930018 360001-M23226930028 360001-M23226930023 360001-M23226920012 360001-M23226930013 360001-M23226920016 360001-M23226930003 360001-M23226920006 360001-M23226920020 360001-M23402100003 360001-M23402100004 360001-M23402100010 360001-M23402100013 360001-M23402100017 360001-M23402100018 360001-M23402100019 360001-M23402100009 360001-M23402100014 360001-M23402100002 360001-M23402100016 360001-M23402100006 360001-M23402100012 360001-M23402100020 360001-M23402100008 360001-M23402100007 360001-M23402100015 360001-M23402100011 360001-M23402100001 360001-M23402100005 360001-M23713080014 360001-M23713080020 360001-M23713080019 360001-M23713080003 360001-M23713080010 360001-M23713080007 360001-M23713080018 360001-M23713080016 360001-M23713080008 360001-M23713080002 360001-M23713080017 360001-M23713080015 360001-M23713080012 360001-M23713080009 360001-M23713080004 360001-M23713080011 360001-M23713080006 360001-M23713080005 360001-M23713080001 360001-M23713080013 360001-M23813100012 360001-M23813100010 360001-M23813100013 360001-M23813100003 360001-M23813100006 360001-M23813100009 360001-M23813100007 360001-M23813100002 360001-M23813100014 360001-M23813100011 360001-M23813100008 360001-M23813100004 360001-M23813100001 360001-M23813100015 360001-M23813100005 360001-M23915460009 360001-M23915460015 360001-M23915460020 360001-M23915450002 360001-M23915450004 360001-M23915460003 360001-M23915460006 360001-M23915460014 360001-M23915460019 360001-M23915460007 360001-M23915450003 360001-M23915450001 360001-M23915450015 360001-M23915460002 360001-M23915460010 360001-M23915460011 360001-M23915450006 360001-M23915450008 360001-M23915450011 360001-M23915450013 360001-M23915460008 360001-M23915460017 360001-M23915450009 360001-M23915450010 360001-M23915450012 360001-M23915450014 360001-M23915460001 360001-M23915460013 360001-M23915460016 360001-M23915450005 360001-M23915460005 360001-M23915460018 360001-M23915450007 360001-M23915460004 360001-M23915460012 360001-M23C15850005 360001-M23C15860008 360001-M23C15850007 360001-M23C15860005 360001-M23C15860007 360001-M23C15850001 360001-M23C15850004 360001-M23C15860002 360001-M23C15860010 360001-M23C15850002 360001-M23C15850003 360001-M23C15850010 360001-M23C15860001 360001-M23C15860004 360001-M23C15850009 360001-M23C15850006 360001-M23C15860009 360001-M23C15860003 360001-M23C15850008 360001-M23C15860006 360001-M24104240001 360001-M24104240007 360001-M24104040008 360001-M24104040009 360001-M24104040011 360001-M24104040012 360001-M24104240003 360001-M24104240004 360001-M24104240006 360001-M24104240012 360001-M24104040003 360001-M24104040004 360001-M24104040005 360001-M24104040006 360001-M24104240008 360001-M24104240010 360001-M24104240013 360001-M24104040007 360001-M24104040001 360001-M24104040002 360001-M24104040010 360001-M24104040014 360001-M24104240009 360001-M24104240005 360001-M24104240011 360001-M24104040015 360001-M24104040013 360001-M24104240014 360001-M24104240002 360001-M24104240015 360001-M24306000006 360001-M24306000008 360001-M24306000009 360001-M24306000001 360001-M24306000004 360001-M24306000010 360001-M24306000005 360001-M24306000003 360001-M24306000002 360001-M24306000007 360001-M24707820007 360001-M24707820004 360001-M24707820020 360001-M24707820016 360001-M24707820008 360001-M24707820011 360001-M24707820003 360001-M24707820002 360001-M24707820013 360001-M24707820024 360001-M24707820017 360001-M24707820015 360001-M24707820019 360001-M24707820025 360001-M24707820012 360001-M24707820009 360001-M24707820001 360001-M24707820023 360001-M24707820005 360001-M24707820022 360001-M24707820006 360001-M24707820010 360001-M24707820021 360001-M24707820018 360001-M24707820014 360001-M24716370013 360001-M24716370011 360001-M24716370003 360001-M24716370001 360001-M24716370014 360001-M24716370012 360001-M24716370004 360001-M24716370002 360001-M24716370010 360001-M24716370005 360001-M24716370007 360001-M24716370015 360001-M24716370017 360001-M24716370006 360001-M24716370016 360001-M24716370008 360001-M24716370009 360001-M24716370018 360001-M24716370019 360001-M24716370020 360001-M24A07710007 360001-M24A07710002 360001-M24A07710006 360001-M24A07710010 360001-M24A07710001 360001-M24A07710004 360001-M24A07710009 360001-M24A07710005 360001-M24A07710008 360001-M24A07710003 360147-M17615170005 360147-M17615170003 360147-M17615170002 301243-M17615280049 301243-M17615280047 301243-M17615280046 301243-M17615280033 301243-M17615280027 301243-M17615280026 301243-M17615280018 301243-M17615280009 301243-M17615280007 301243-M17615280006 301235-M17615270016 301235-M17615270010 301235-M17615270005 260168-M17615180008 301243-M17615280035 301243-M17615280005 301243-M17615280016 301243-M17615280017 301243-M17615280023 301243-M17615280031 301243-M17615280034 301243-M17615280043 301243-M17615280045 301243-M17615280050 301243-M17615280053 360147-M17615170012 360147-M17615170013 360147-M17615170015 360147-M17615170011 301243-M17615280058 301243-M17615280044 301243-M17615280041 301243-M17615280032 301243-M17615280021 301243-M17615280014 301235-M17615270002 301235-M17615270001 260168-M17615180006 301243-M17615280011 301243-M17615280002 301235-M17615270025 301235-M17615270024 301235-M17615270023 301243-M17615280028 301243-M17615280003 301243-M17615280036 301243-M17615280039 360147-M17615170018 360147-M17615170016 360147-M17615170010 360147-M17615170004 301243-M17615280048 301243-M17615280037 301243-M17615280059 301235-M17615270015 301235-M17615270013 260168-M17615180007 301235-M17615270007 301235-M17615270008 301235-M17615270009 301243-M17615280015 301243-M17615280024 301243-M17615280030 301243-M17615280051 301243-M17615280054 360147-M17615170001 360147-M17615170014 260168-M17615180010 301235-M17615270006 301235-M17615270017 301235-M17615270021 301235-M17615270022 301243-M17615280019 301243-M17615280025 301243-M17615280055 301243-M17615280060 360147-M17615170007 360147-M17615170009 360147-M17615170017 260168-M17615180002 301235-M17615270014 360147-M17615170008 301243-M17615280052 260168-M17615180003 260168-M17615180005 260168-M17615180009 301235-M17615270011 301235-M17615270012 301235-M17615270020 301243-M17615280001 301243-M17615280004 301243-M17615280013 301243-M17615280020 301243-M17615280022 301243-M17615280038 301243-M17615280040 360147-M17615170006 301243-M17615280057 301243-M17615280029 301243-M17615280056 301243-M17615280012 301243-M17615280010 301243-M17615280008 301235-M17615270019 301235-M17615270018 301235-M17615270003 260168-M17615180004 260168-M17615180001 301235-M17615270004 301243-M17615280042 260087-M15A04410004 260087-M15A04410001 301448-M15B11140055 301448-M15B11140036 301448-M16410840050 301448-M16410840064 301448-M17302790006 301448-M17302790005 301448-M17302790007 301448-M17603360035 301448-M17706010024 301448-M17706010098 301448-M18112380026 301448-M18112380024 301448-M18112550055 301448-M18407200003 301448-M18C06300039 301448-M19108500012 301448-M19505960005 301448-M19505960001 301448-M19505960003 301448-M19505960004 360001-M24A19090019 360001-M24A19090006 360001-M24A19090001 360001-M24A19090016 360001-M24A19090014 360001-M24A19090011 360001-M24A19090009 360001-M24A19090017 360001-M24A19090018 360001-M24A19080006 360001-M24A19080007 360001-M24A19080008 360001-M24A19080009 360001-M24A19090003 360001-M24A19090004 360001-M24A19090012 360001-M24A19090020 360001-M24A19080001 360001-M24A19080002 360001-M24A19090015 360001-M24A19090013 360001-M24A19090002 360001-M24A19080005 360001-M24A19080004 360001-M24A19080003 360001-M24A19090010 360001-M24A19090008 360001-M24A19090007 360001-M24A19080010 360001-M24A19090005 360001-M24B19850004 360001-M24B19850009 360001-M24B19850002 360001-M24B19850005 360001-M24B19850001 360001-M24B19850008 360001-M24B19850006 360001-M24B19850010 360001-M24B19850003 360001-M24B19850007 360001-M24C20530010 360001-M24C20530006 360001-M24C20530004 360001-M24C20530008 360001-M24C20530002 360001-M24C20530007 360001-M24C20530005 360001-M24C20530003 360001-M24C20530009 360001-M24C20530001 360001-M25627280009 360001-M25627280008 360001-M25627280004 360001-M25627280005 360001-M25627280003 360001-M25627280006 360001-M25627280007 360001-M25627280002 360001-M25627280001 360001-M25627280010 EDAN Model: iM8A UDI-DI code: 06944413801608 Serial/Lot Numbers: 262028-M20411730007 262028-M20411730006 262028-M20411730024 262028-M20411730023 262028-M20411730019 262028-M20411730005 262028-M20411730034 262028-M20411730010 262028-M20411730008 262028-M20411730035 262028-M20411730022 262028-M20411730013 262028-M20411730014 262028-M20411730020 262028-M20411730030 262028-M20411730031 262028-M20411730018 262028-M20411730021 262028-M20411730026 262028-M20411730028 262028-M20411730033 262028-M20411730002 262028-M20411730032 262028-M20411730003 262028-M20411730004 262028-M20411730009 262028-M20411730015 262028-M20411730025 262028-M20411730029 262028-M20411730017 262028-M20411730016 262028-M20411730012 262029-M20411740005 262029-M20411740002 262029-M20411740014 262028-M20411730027 262029-M20411740024 262029-M20411740026 262029-M20411740007 262029-M20411740009 262029-M20411740011 262029-M20411740021 262029-M20411740001 262029-M20411740010 262029-M20411740017 262029-M20411740019 262029-M20411740020 262029-M20411740016 262029-M20411740027 262029-M20411740022 262028-M20411730001 262029-M20411740025 262029-M20411740015 262029-M20411740030 262029-M20411740028 262029-M20411740023 262028-M20411730011 262029-M20411740013 262029-M20411740006 262029-M20411740004 262029-M20411740012 262029-M20411740008 262029-M20411740003 262029-M20411740018 262029-M20411740029 360001-M20415950001 262030-M20411750014 262030-M20411750001 262030-M20411750012 262030-M20411750017 262030-M20411750018 262030-M20411750022 262030-M20411750006 262030-M20411750019 262030-M20411750028 262030-M20411750029 262030-M20411750004 262030-M20411750005 262030-M20411750010 262030-M20411750013 262030-M20411750020 262030-M20411750023 262030-M20411750032 262030-M20411750027 262030-M20411750003 262030-M20411750016 262030-M20411750007 262030-M20411750035 262030-M20411750034 262030-M20411750021 262030-M20411750011 262030-M20411750008 262030-M20411750002 262030-M20411750033 262030-M20411750009 262030-M20411750030 262030-M20411750031 262030-M20411750026 262030-M20411750015 262030-M20411750024 262030-M20411750025 262029-M21702070013 262029-M21702070012 262029-M21702070014 262029-M21702070008 262029-M21702070011 262029-M21702070010 262029-M21702070005 262029-M21702070018 262029-M21702070007 262030-M21702080006 262030-M21702080024 262030-M21702080002 262030-M21702080008 262030-M21702080021 262030-M21702080023 262030-M21702080005 262029-M21702070016 262030-M21702080030 262030-M21702080013 262030-M21702080016 262030-M21702080027 262029-M21702070001 262029-M21702070004 262030-M21702080026 262029-M21702070015 262030-M21702080001 262030-M21702080019 262030-M21702080009 262030-M21702080017 262030-M21702080004 262030-M21702080010 262030-M21702080012 262030-M21702080018 262030-M21702080022 262030-M21702080003 262030-M21702080011 262030-M21702080015 262030-M21702080020 262030-M21702080028 262030-M21702080007 262030-M21702080014 262030-M21702080029 262029-M21702070017 262030-M21702080025 262029-K21700040001 262029-M21702070006 262029-M21702070009 262029-M20601160001 262029-M21702070002 262029-M21702070003 262030-M23913770001 262030-M23913770002 EDAN Model: iM8B UDI-DI code: 06944413801615 Serial/Lot Numbers: 260133-M16906960001 260133-M16906960002 260133-M16905680001 260133-M16905680002 360001-M20811210001 360001-M20907160001 360001-M20914640003 360001-M20914640008 360001-M20914640002 360001-M20914640001 360001-M20914640007 360001-M20914640009 360001-M20914640010 360001-M20914640006 360001-M20914640004 360001-M20914640005
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Patient Monitor iM8Edan Instruments, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 recall0 reports
- Patient Monitor iM8AEdan Instruments, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 recall0 reports
- Patient Monitor iM8BEdan Instruments, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97879).
- Z-1140-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Fetal Monitor Model/Catalog Number: F2, F3 Product Description: The devices are bedside fetal and maternal monitors, which are used to monitor the ph…
- Z-1141-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Fetal & Maternal Monitor Model/Catalog Number: F6, F9, F6 Express, F9 Express Product Description: The devices are bedside fetal and maternal monitor…
- Z-1142-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3 Product Description: The Vital Signs Monitor is a portable device intended for use by health care profes…
- Z-1143-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3 Product Description: The iM3 has three work modes, Monitor, Spot and Round, to measure physiological pa…
- Z-1144-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3s, iM3As, iM3Bs, iHM3s Product Description: The iM3s series vital signs monitors including iM3s/iM3As/iM…
- Z-1145-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70, iM80 Product Description: The iM series Patient Monitor including iM50, iM60, iM70 and iM80 c…
- Z-1146-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: elite V5, elite V6, elite V8 Product Description: V series Patient Monitor including elite V5, elite V6 and eli…
- Z-1147-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM20 Product Description: The iM20 Patient Monitor System (hereinafter called iM20) can perform long-time conti…
- Z-1148-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: X8, X10, X12 Product Description: The X8 X10 X12 Patient Monitor (hereinafter called X series) can perform long…
- Z-1149-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iX10, iX12, iX15 Product Description: The iX series Patient Monitors including iX10, iX12, iX15 can perform lon…
- Z-1150-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CMS Product Description: MFM-CMS is a central monitoring system product, which can connect and ma…
- Z-1151-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CNS, MFM-CNS Lite Product Description: The MFM-CNS and MFM-CNS Lite are clinical data managing so…
- Z-1152-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3B Product Description: M3B Vital Signs Monitor is a patient monitoring device providing the patient with …
- Z-1153-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3A Product Description: The Vital Signs Monitor is a portable device intended for use by health care profe…
- Z-1155-2026 · Open, ClassifiedBrand Name: EDAN Product Name: Telemetry Transmitter Model/Catalog Number: iT20 Product Description: iT20 collects physiological parameters by ECG cables and SpO2 sensors, then ach…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page