Skip to content

SINGLE USE BIOPSY VALVE recallZ-1488-2026 · Olympus Corporation of the Americas · initiated Jan 29, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for rubber fragment detachment during use.

FDA root cause category: Under Investigation by firm

What the company is doing

Olympus notified consignees on about 01/29/2026 via letter titled URGENT: FIELD SAFETY NOTICE. Consignees were instructed to carefully read the content of the notification, ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, and that they may continue using the devices as per the notification and instructions for use. Olympus is highlighting the importance of strictly adhering to the Inspection of the Biopsy Valve and Inserting and Withdrawing the Endo-therapy accessories sections in the IFU related to the detachment of fragment from the slit part of rubber. Consignees were requested to acknowledge receipt of the notification via their portal and if product has been further distributed to notify those consignees.

Recall details

Recall number
Z-1488-2026
Status
Open, Classified
Initiated by the company
Jan 29, 2026
Posted by the FDA
Mar 3, 2026
Recalling firm
Olympus Corporation of the Americas
Quantity
95,882 boxes (20 per box)
Product code
EOQ
FDA submissions
K061313
Distribution
US Nationwide distribution.

Product description

Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent reflux of body fluids.

Code information (models, lots, serial numbers)

Model/Catalog Number: MAJ-210. UDI-DI: 14953170152433, 14953170452069. All Lot Numbers. 20 units per box.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98240).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page