Spectral CT recallZ-1317-2026 · Philips North America Llc · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
FDA root cause category: Process change control
What the company is doing
On or about January 7, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. As a precaution, monitor the gantry for any unusual sounds or vibrations during operation. If unusual sounds are heard or vibrations are observed, immediately call your service contact. In the absence of unusual sounds or vibrations, you may continue to use your system(s) according to the intended use. 2. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until Philips has inspected your system and has provided any correction, if needed. Ensure the letter is in a place likely to be seen and read. Actions planned by Philips: Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and to inspect the fasteners on your system (reference FCO72800845). If the inspection indicates that your system is affected, Philips will perform the correction to resolve the issue (reference FCO72800846)
Recall details
- Recall number
- Z-1317-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- Feb 6, 2026
- Recalling firm
- Philips North America Llc
- Quantity
- 283 units
- Product code
- JAK
- FDA submissions
- K244008
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.
Product description
Spectral CT; Product Code (REF): 728333;
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Product Code (REF): 728333; UDI-DI: 00884838101111; Serial Numbers: 10055, 10075, 10086, 10096, 10106, 10116, 10126, 10136, 10146, 10156, 10166, 10176, 10186, 10196, 10206, 10217, 10227, 10237, 10247, 10257, 10267, 10277, 10289, 10302, 10314, 12009, 12019, 12029, 396013, 10056, 10076, 10087, 10097, 10107, 10117, 10127, 10137, 10147, 10157, 10167, 10177, 10187, 10197, 10207, 10218, 10228, 10238, 10248, 10258, 10268, 10278, 10290, 10303, 12000, 12010, 12020, 12030, 398003, 10057, 10077, 10088, 10098, 10108, 10118, 10128, 10138, 10148, 10158, 10168, 10178, 10188, 10198, 10208, 10219, 10229, 10239, 10249, 10259, 10269, 10279, 10291, 10304, 12001, 12011, 12021, 12031, 398004, 10058, 10078, 10089, 10099, 10109, 10119, 10129, 10139, 10149, 10159, 10169, 10179, 10189, 10199, 10209, 10220, 10230, 10240, 10250, 10260, 10270, 10280, 10292, 10305, 12002, 12012, 12022, 396005, 10059, 10079, 10090, 10100, 10110, 10120, 10130, 10140, 10150, 10160, 10170, 10180, 10190, 10200, 10210, 10221, 10231, 10241, 10251, 10261, 10271, 10281, 10293, 10306, 12003, 12013, 12023, 396006, 10060, 10080, 10091, 10101, 10111, 10121, 10131, 10141, 10151, 10161, 10171, 10181, 10191, 10201, 10211, 10222, 10232, 10242, 10252, 10262, 10272, 10284, 10294, 10307, 12004, 12014, 12024, 396008, 10061, 10081, 10092, 10102, 10112, 10122, 10132, 10142, 10152, 10162, 10172, 10182, 10192, 10202, 10212, 10223, 10233, 10243, 10253, 10263, 10273, 10285, 10298, 10308, 12005, 12015, 12025, 396009, 10062, 10082, 10093, 10103, 10113, 10123, 10133, 10143, 10153, 10163, 10173, 10183, 10193, 10203, 10213, 10224, 10234, 10244, 10254, 10264, 10274, 10286, 10299, 10310, 12006, 12016, 12026, 396010, 10070, 10083, 10094, 10104, 10114, 10124, 10134, 10144, 10154, 10164, 10174, 10184, 10194, 10204, 10214, 10225, 10235, 10245, 10255, 10265, 10275, 10287, 10300, 10311, 12007, 12017, 12027, 396011, 10074, 10085, 10095, 10105, 10115, 10125, 10135, 10145, 10155, 10165, 10175, 10185, 10195, 10205, 10215, 10226, 10236, 10246, 10256, 10266, 10276, 10288, 10301, 10312, 12008, 12018, 12028, 396012.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98248).
- Z-1315-2026 · Open, ClassifiedBrilliance iCT; Product Code (REF): 728306;
- Z-1316-2026 · Open, ClassifiedIQon Spectral CT; Product Code (REF): 728332;
- Z-1318-2026 · Open, ClassifiedSpectral CT 7500 on Rails; Product Code (REF): 728334;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page