Raz-AT (Attendant Tilt) Mobile Shower Commode Chair… recallZ-1498-2026 · Raz Design Inc · initiated Jan 5, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raz Design Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in place.
FDA root cause category: Use error
What the company is doing
On 01/05/2025, the firm email an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that an incident was reported to Raz involving a Raz Mobile Shower Commode Chair with a certain Front and Rear Open (FAR) Seat. This incident in Canada involved an unauthorized modification of a FAR seat by a vendor, followed by reinstallation without properly engaging the outside rear seat brackets that keep the seat in place. This is not a supported configuration, and it resulted in one side of the seat halves slipping off the seat frame, which caused an injury to the user. Customers are instructed to: 1 Notify impacted customers using this Correction Notice and request that they return the Acknowledgement and Receipt Form to you. 2 Raz will provide items to facilitate Correction. The items will be provided directly to the Sales Rep or to Dealer, whichever will execute the Correction action: 3 Sales Rep or Dealer shall conduct a service visit with the impacted end-user. For users/caregivers: 1 Be sure that the FAR seat is properly installed and latched before a user is transferred onto the chair. The rear seat attachment bracket must be fully hooked under the rear seat support tube(s). The front seat attachment brackets must be securely latched around the front seat support tubes. Both front and rear seat attachment brackets should be secured in place by tightening the 8 screws on the underside of the seat. 2 if seat needs to be removed for cleaning, follow instructions as per User & Dealer Manual proper removal and installation of the seat to ensure safe use of the chair prior to the installation of the FAR Seat Retrofit Kit. 3 If you do not have a User and Dealer Manual, download a copy from www.razdesigninc.com, or contact Sales Rep or Dealer. For questions or further assistance - contact Raz Design Inc by email at [email protected]
Recall details
- Recall number
- Z-1498-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 5, 2026
- Posted by the FDA
- Mar 6, 2026
- Recalling firm
- Raz Design Inc
- Quantity
- 66
- Product code
- INN
- Distribution
- U.S. Nationwide distribution in the states of CO, FL, IL, MA, MI, MN, MO, NC, NE, NH, NJ, NY, OR, PA, TX, WA, and WI.
Product description
Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007 Raz-ART (Attendant Rotational Tilt) Mobile Shower Commode Chair, Catalog Number Z333 UDI-DI code: (01)00628758003335 Raz-AT600 (Bariatric Tilt) Mobile Shower Commode Chair, Catalog Number Z360 UDI-DI code: (01)00628758003601 Raz-SP (Self Propel) Mobile Shower Commode Chair, Catalog Number Z200 UDI-DI code: (01)00628758001003 Raz-AP (Attendant Propel) Mobile Shower Commode Chair, Catalog Number Z100 UDI-DI code: (01)00628758001003 The Raz Mobile Shower Commode Chair is intended to be used by people with mobility impairments, who may have difficulty in standing or keeping their balance while completing activities such as showering, toileting, or personal hygiene.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Raz-AT (Attendant Tilt) Mobile Shower Commode Chair Catalog Number Z300 Seat Catalog Numbers: ZVFIFAR17 ZVFIFAR20 ZCSEAT (FAR) UDI-DI code: (01)00628758003007 Serial Numbers: RAT218789 RAT116906 RAT200337 RAT208010 RAT208987 RAT210426 RAT211344 RAT211837 RAT211998 RAT212001 RAT212066 RAT212450 RAT213190 RAT213227 RAT214334 RAT217987 RAT218213 RAT218552 RAT218570 RAT218612 RAT219142 RAT220066 RAT222210 RAT223140 RAT224902 RAT206923 RAT217768 RAT221724 RAT213036 RAT218862 RAT221044 RAT216437 RAT217073 RAT218555 RAT218557 RAT219817 RAT121348 RAT216055 RAT217379 RAT220648 RAT124264 RAT200864 Raz-ART (Attendant Rotational Tilt) Mobile Shower Commode Chair Catalog Number Z333 Seat Catalog Numbers: ZVFIFAR17 ZVFIFAR20 ZCSEAT (FAR) UDI-DI code: (01)00628758003335 Serial Numbers: ART219556 ART222544 ART211605 ART223825 ART212722 ART212675 Raz-AT600 (Bariatric Tilt) Mobile Shower Commode Chair Catalog Number Z360 Seat Catalog Numbers: ZCSEATHD (FAR) UDI-DI code: (01)00628758003601 Serial Numbers: BAT216618 BAT204112 Raz-SP (Self Propel) Mobile Shower Commode Chair Catalog Number Z200 Seat Catalog Numbers: ZVFIFAR17 ZVFIFAR20 ZCSEAT (FAR) UDI-DI code: (01)00628758001003 Serial Numbers: RAS210872 RAS214903 RAS217818 RAS220760 RAS209166 RAS218239 RAS107030 RAS214470 Raz-AP (Attendant Propel) Mobile Shower Commode Chair Catalog Number Z100 Seat Catalog Numbers: ZVFIFAR17 ZVFIFAR 20 ZCSEAT (FAR) UDI-DI code: (01)00628758001003 Serial Numbers: RAS217210 RAS224711 RAS214364 FAR Seat (Demo) Catalog Numbers: ZVFI17FAR-A ZCSEAT-A UDI-DI code: N/A Serial Numbers: N/A
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Raz Raz-APRaz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Raz Raz-ATRaz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Raz Raz-ARTRaz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Raz Raz-AT600Raz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page