Vortex Disposable Forceps and Cannula recallZ-1247-2026 · Vortex Surgical Inc. · initiated Dec 16, 2025
Class II Open, Classified All recalls posted in February 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Vortex Surgical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
FDA root cause category: Employee error
What the company is doing
On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: [email protected]. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: [email protected] from 8am to 5pm.
Recall details
- Recall number
- Z-1247-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Dec 16, 2025
- Posted by the FDA
- Feb 4, 2026
- Recalling firm
- Vortex Surgical Inc.
- Quantity
- 14,789 (8651 US; 6138 OUS)
- Product code
- HNR
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Product description
Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps Adaptive VS0740.25 3. 25GA ACTU8 Forceps ILM VS0744.25 4. 25GA ACTU8 Forceps ILM -Myopic VS0744.25-35 5. 25GA ACTU8 Forceps Pro-Grip VS0713.25 6. 25GA ACTU8 Talon Forceps ILM VS0752.25 7. 25GA ACTU8 Talon Forceps Pro-Grip VS0751.25 8. 27GA ACTU8 Forceps ILM VS0744.27 9. 23GA Soft Tip Cannula VS0200.23 10. 23GA Soft Tip Cannula - 0.5mm VS0205.23 11. 25GA Soft Tip Cannula VS0200.25 12. 25GA Soft Tip Cannula - 0.5mm VS0205.25 13. 25GA Soft Tip Cannula - 0.5mm - Bulk VS0205.25B 14. 25GA Soft Tip Cannula-Bulk VS0200.25B 15. 27GA Soft Tip Cannula VS0200.27
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1. 23GA ACTU8 Forceps Adaptive Pouch UDI 810123481125 Box UDI 810123481538 Lot 2411009 2. 25GA ACTU8 Forceps Adaptive Pouch UDI 810123481132 Box UDI 810123481545 Lot 2410051 3. 25GA ACTU8 Forceps ILM Pouch UDI 810123481231 Box UDI 810123481644 Lot 2410021 2410053 2411006 4. 25GA ACTU8 Forceps ILM -Myopic Pouch UDI 810123481248 Box UDI 810123481651 Lot 2410059 5. 25GA ACTU8 Forceps Pro-Grip Pouch UDI 810123481101 Box UDI 810123481514 Lot 2410050 2411011 6. 25GA ACTU8 Talon Forceps ILM Pouch UDI 810123481835 Box UDI 810123481859 Lot 2411003 7. 25GA ACTU8 Talon Forceps Pro-Grip Pouch UDI 810123482375 Box UDI 810123482405 Lot 2410060 8. 27GA ACTU8 Forceps ILM Pouch UDI 810123481255 Box UDI 810123481668 Lot 2410054 9. 23GA Soft Tip Cannula UDI Pouch 810123480302 Box UDI 810123480562 Lot 2410037 10. 23GA Soft Tip Cannula - 0.5mm UDI Pouch 810123480388 Box UDI 810123480661 Lot 2411032 11. 25GA Soft Tip Cannula UDI Pouch 810123480319 Box UDI 810123480586 Lot 2410039 12. 25GA Soft Tip Cannula - 0.5mm UDI Pouch 810123480395 Box UDI 810123480678 Lot 2410041 13. 25GA Soft Tip Cannula - 0.5mm - Bulk UDI Pouch 810123480395 Box UDI 810123481927 Lot 2410042 2411018 14. 25GA Soft Tip Cannula-Bulk UDI Pouch 810123480319 Box UDI 810123480593 Lot 2410040 15. 27GA Soft Tip Cannula UDI Pouch 810123480326 Box UDI 810123480609 Lot 2412026
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- Vortex Surgical Inc VS0740.25Vortex Surgical, Inc. · Forceps, Ophthalmic2 recalls0 reports
- Vortex Surgical Inc VS0200.23Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0200.25Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0200.25BVortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0200.27Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0205.23Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0205.25Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0713.25Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0740.23Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0744.25Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0744.25-35Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0744.27Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
- Vortex Surgical Inc. VS0752.25Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0205.25BVortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc. VS0751.25Vortex Surgical, Inc. · Forceps, Ophthalmic1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98255).
- Z-1245-2026 · Open, ClassifiedVortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Ill…
- Z-1248-2026 · Open, Classified25GA Subretinal Injection Cannula VS0220.25
- Z-1249-2026 · Open, ClassifiedConvenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.25GA Convenience Kit - XES VS0644.25-XES
- Z-1250-2026 · Open, ClassifiedVortex Surgical 25GA I.D.D. Internal Delivery Device, VS0250.25
- Z-1252-2026 · Open, ClassifiedVortex Surgical I2 Injection Kit, VS0500
- Z-1254-2026 · Open, ClassifiedVortex Surgical TID Pharos Illuminated Depressor, VS0801B
- Z-1255-2026 · Open, ClassifiedVortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27
- Z-1256-2026 · Open, ClassifiedVortex Surgical Lindsell Sutured IOL Marker, VS0390 - Scleral Marker
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page