Skip to content

Olympus Resection Sheath. Model No. A22014A and A22014T.… recallZ-1438-2026 · Olympus Corporation of the Americas · initiated Nov 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Complaints of the ceramic tip of the resection sheath breaking have been received.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any questions may contact Cynthia Ow at [email protected] or by phone at 947-999-3203.

Recall details

Recall number
Z-1438-2026
Status
Open, Classified
Initiated by the company
Nov 11, 2025
Posted by the FDA
Feb 20, 2026
Recalling firm
Olympus Corporation of the Americas
Quantity
29 units
Product code
FJL
FDA submissions
K931995
Distribution
US Nationwide Distribution.

Product description

Olympus Resection Sheath. Model No. A22014A and A22014T. for endoscopic diagnosis and treatment in urological applications.

Code information (models, lots, serial numbers)

Model: A22014A (UDI: 04042761020893) and A22014T (UDI: 04042761020909). All lots.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K931995 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98288).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page