MOLLI recallZ-1427-2026 · Stryker Corporation · initiated Jan 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.
FDA root cause category: Under Investigation by firm
What the company is doing
On 01/21/2026, the firm sent via USPS certified mail an "URGENT: Medical Device Recall" Letter to customers informing them that Stryker has updated the MOLLI 2 System Instructions for Use (IFU) following two Medical Device Reports (MDRs) submitted to the FDA. These reports involved patient injuries associated with MOLLI marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement. Customers are instructed to: 1. Inform individuals within their organization who need to be aware of this warning statement. 2. Review Attachment A regarding the warning statement that will be added to the IFU. Stryker estimates the updated IFUs to be released by 28FEB2026. No product return is required. 3. Complete Attachment B, Business Reply Form (page 4), as an acknowledgement of this notice and return the completed form via email to [email protected]. Response is required. 4. If customers have further distributed product, complete the table in Attachment B, Business Reply Form (page 4) and return the completed form via email to [email protected]. 5. Maintain awareness of this communication internally. For questions - contact [email protected] or Stryker Customer Service at 800-624-4422.
Recall details
- Recall number
- Z-1427-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 21, 2026
- Posted by the FDA
- Feb 17, 2026
- Recalling firm
- Stryker Corporation
- Quantity
- 5,864
- Product code
- NEU
- FDA submissions
- K231579, K253888
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, HI ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY. The countries of Canada, Singapore, Panama, Jamaica, and Cayman Islands.
Product description
MOLLI 2 System System includes: MM1000 (Pack of 10) - MOLLI Introducer (8cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195028 MS-00009 (Pack of 10) - MOLLI Introducer (12cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195134 MP2001 - MOLLI 2 Wand UDI (GTIN): 00850024195141 MP2015A - MOLLI OncoPen UDI (GTIN): 00850024195219 MT2001 - MOLLI 2 Tablet UDI (GTIN): 00850024195158 The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. the MOLLI Marker is located and surgically removed with the target tissue.
Code information (models, lots, serial numbers)
Show the full code information (8k characters)
Catalog Number: MM1000 (Pack of 10) UDI (GTIN): 00850024195028 Serial Numbers: 23-0350 24-0118 24-0289 24-0290 24-0291 24-0472 24-0473 24-0474 24-0534 24-0535 24-0536 24-0658 24-0660 24-0697 25-0020 25-0021 25-0022 25-0160 25-0161 25-0162 25-0353 25-0354 25-1057 25-1058 25-1060 25-1087 23-0003 23-0046 23-0057 23-0062 23-0125 23-0126 23-0129 23-0176 23-0177 23-0179 23-0413 23-0414 24-0413 Catalog Number: MS-00009 (Pack of 10) UDI (GTIN):00850024195134 Serial Numbers: 22-0130 23-0022 23-0130 23-0178 24-0035 24-0117 24-0225 24-0235 24-0236 24-0659 24-0696 24-0719 25-0251 25-0252 Catalog Number: MP2001 UDI (GTIN): 00850024195141 Serial Numbers: 16 18 29 30 32 43 49 54 60 62 66 77 80 81 84 85 92 93 100 101 114 115 138 155 158 160 163 164 174 180 181 182 184 186 190 199 202 203 216 217 218 220 221 222 225 226 227 228 229 230 231 232 234 235 236 237 238 239 240 241 242 243 244 245 246 249 250 251 252 253 254 255 256 257 258 259 260 261 262 263 264 265 266 267 268 269 270 271 273 274 275 276 277 278 279 280 281 282 283 284 285 286 287 288 290 291 292 293 294 295 296 297 298 299 300 301 302 303 304 305 307 310 311 313 315 316 317 318 319 320 321 322 323 324 326 327 328 329 330 331 332 333 334 335 336 337 338 339 340 341 342 343 344 345 346 347 348 349 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366 367 368 369 370 371 372 373 374 375 376 377 378 379 380 381 382 383 384 385 386 387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 405 406 408 409 410 411 412 413 414 415 416 417 418 419 420 421 422 423 424 425 426 427 428 429 430 431 432 433 434 435 436 437 438 439 440 441 442 443 444 446 447 448 449 450 451 453 454 455 456 457 458 460 461 464 465 466 467 468 469 470 471 472 473 474 475 476 477 478 479 480 481 482 483 484 485 486 487 488 489 490 491 492 493 494 495 496 497 498 499 500 501 502 503 504 505 506 507 508 509 510 511 512 513 514 515 516 517 518 519 520 521 522 523 524 525 526 527 528 529 530 531 532 533 534 535 536 537 538 539 540 541 542 543 544 545 546 547 548 549 550 551 552 553 554 555 556 558 559 560 561 562 563 564 565 566 567 568 569 570 571 573 574 575 576 577 578 580 581 583 584 585 586 587 588 589 590 592 601 605 606 607 608 609 610 611 612 613 614 615 616 617 618 619 620 621 622 623 624 625 626 627 628 629 630 631 632 633 634 635 636 637 638 639 640 641 644 645 646 647 651 652 653 654 655 656 657 658 659 660 661 662 663 666 670 674 675 730 742 743 745 747 750 751 752 753 754 755 756 757 758 763 764 765 766 767 768 769 772 773 775 776 777 780 782 783 784 785 786 787 788 929 930 940 959 973 976 981 982 984 986 988 990 995 997 998 1001 1003 1004 1005 1006 1008 1009 1014 1015 1016 1092 Catalog Number: MP2015A UDI (GTIN): 00850024195219 Serial Numbers: 0010 0023 0030 0031 0032 0033 0034 0035 0036 0038 0039 0041 0042 0043 0044 0045 0046 0049 0050 0052 0054 0055 0056 0057 0059 0063 0065 0066 0070 0073 0074 0075 0076 0078 0079 0080 0081 0082 0085 0086 0087 0088 0089 0090 0091 0092 0094 0095 0096 0097 0099 0100 0101 0103 0104 0105 0106 0107 0108 0109 0110 0111 0112 0113 0114 0115 0116 0117 0118 0121 0122 0123 0124 0125 0126 0127 0128 0129 0131 0132 0133 0135 0136 0137 0138 0139 0140 0142 0143 0148 0149 0150 0151 0152 0153 0154 0155 0156 0157 0158 0159 0160 0161 0162 0164 0165 0166 0167 0168 0169 0170 0171 0172 0173 0174 0175 0176 0177 0179 0180 0181 0182 0183 0184 0186 0190 0192 0194 0197 0198 0199 0200 0201 0202 0203 0204 0205 0206 0207 0208 0209 0210 0211 0213 0215 0216 0217 0218 0219 0220 0221 0222 0223 0225 0226 0227 0229 0230 0231 0232 0234 0235 0236 0237 0238 0241 0242 0243 0245 0246 0247 0248 0249 0250 0251 0252 0253 0254 0255 0256 0257 0258 0259 0260 0262 0264 0265 0266 0267 0268 0270 0272 0273 0279 0280 0282 0283 0286 0289 0290 0291 0292 0293 0294 0296 0297 0298 0299 0300 0301 0302 0303 0304 0305 0306 0307 0308 0309 0310 0311 0312 0313 0314 0315 0316 0317 0318 0319 0321 0323 0324 0325 0326 0327 0328 0329 0330 0332 0333 0334 0335 0336 0337 0338 0339 0340 0341 0343 0344 0345 0346 0347 0348 0349 0350 0353 0355 0357 0359 0361 0362 0363 0364 0366 0368 0369 0371 0372 0373 0375 0376 0377 0379 0381 0382 0383 0384 0385 0386 0388 0389 0390 0391 0392 0393 0394 0395 0396 0397 0398 0399 0400 0401 0402 0403 0404 0405 0406 0407 0408 0409 0410 0411 0412 0413 0414 0415 0416 0417 0418 0419 0420 0421 0422 0423 0424 0425 0426 0427 0428 0429 0430 0431 0432 0433 0434 0435 0436 0437 0438 0439 0441 0443 0444 0445 0446 0449 0453 0454 0459 0466 0482 0508 0509 0510 0511 0512 0513 0514 0515 0516 0517 0518 0520 0521 0522 0523 0525 0527 0530 0531 0532 0533 0534 0535 0536 0537 0538 0539 0540 0541 0543 0544 0545 0548 0549 0550 0551 0553 0554 0555 0557 0558 0560 0563 0566 0568 0569 0570 0571 0572 0573 0575 0576 0578 0579 0602 0603 0604 0605 0606 0607 0608 0609 0610 0611 0612 0613 Catalog Number: MT2001 UDI (GTIN): 00850024195158 Serial Numbers: 3 33 34 35 36 198 202 208 210 214 215 227 228 239 240 245 246 247 247 253 263 264 268 269 270 274 277 299 302 303 304 305 306 307 308 308 309 310 311 312 313 313 314 315 316 317 318 319 320 321 322 323 403 404 505 600 602 604 606 610 611 612 613 614 615 616 617 618 619 620 621 622 623 624 625 626 627 628 629 630 631 632 633 634 635 636 637 638 639 640 641 642 643 644 645 646 647 648 649 650 652 653 654 654 655 656 657 659 660 661 662 663 1000 1001 1002 1003 1004 1004 1005 1006 1007 1008 1009 1010 1011 1012 1013 1014 1015 1016 1017 1018 1019 1020 1021 1022 1023 1024 1025 1026 1028 1029 1030 1031 1032 1033 1034 1035 1036 1037 1039 1040 2013 2031 2032 2033 2034 2035 2036 2039 2200 2203 2204 2205 2208 2209 2210 2211 2211 2212 2213 2215 2216 2217 2218 2220 2221 2222 2224 2400 2401 2402 2403 2404 2405 2406 2407 2408 2409 2410 2411 2412 2413 2414 2415 2416 2416 2417 2418 2419 2420 2421 2422 2423 2423 2424 2425 2426 2427 2428 2429 2430 2431 2432 2433 2435 2436 2437 2438 2439 2440 2441 2442 2443 2444 2445 2446 2447 2449 2450 2451 2452 2453 3000 3001 3002 3002 3003 3005 3006 3007 3008 3009 3010 3011 3012 3013 3014 3015 3016 3017 3019 3020 3021 3022 3023 3024 3025 3026 3032 3033 3041 3042 3043 3044 3045 3046 3047 3048 3049 3050 3051 3052 3053 3054 3056 3057 3058 3061 3062 3063 3064 3065 3066 3067 3068 3069 3071 3072 3073 3075 3077 3078 3081 3085 3086 3092 3093 3094 3096 3097 3098 3099 3100 3101 3102 3103 3106 3107 3108 3109 3110 3111 3112 3113 3114 3115 3116 3117 3118 3119 3120 3121 3122 3123 3124 3125 3126 3127 3128 3129 3130 3131 3132 3133 3134 3135 3136 3138 3139 3140 3141 3142 3143 3144 3145 3146 3148 3149 3150 3152 3155 3157 3159 3161 3162 3164 3165 3167 3168 3169 3171 3172 3173 3174 3175 3176 3177 3179 3183 3184 3186 3187 3188 3190 3191 3192 3195 3196 3198 3199 3201 3202 3203 3205 3206 3209 3210 3211 3213 3214 3217 3218 3219 3220 3221 3222 3223 3224 3225 3226 3227 3228 3229 3230 3231 3232 3233 3238 3239 3240 3241 3242 3244 3245 3246 3247 3248 3249 3250 3252 3255 3256 3257 3260 3263 3264 3266 3267 3268 3269 3270 3272 3274 3275 3276 3277 3278 3279 3282 3283 3284 3285 3286 3287 3288 3289 3290 3291 3292 3293 3294 3295 3296 3297 3298 3299 3300 3301 3302 3303 3304 3305 3306 3307 3308 3309 3310 3311 3312 3313 3314 3315 3316 3317 3318 3319 3321 3322 3323 3324 3325 3326 3327 3328 3329 3330 3331 3332 3333 3334 3335 3336 3337 3338 3339 3340 3341 3342 3343 3344 3345 3346 3347 3348 3349 3350 3351 3352 3353 3354 3355 3356 3357 3359 3360 3361 3362 3363 3365 3369 3370 3371 3372 3373 3374 3375 3376 3377 3379 3380 3381 3382 3383 3384 3386 3387 3388 3389 3390 3392 3393 3394 3396 3397 3398 3399 3400 3401 3402 3403 3404 3405 3406 3408 3409 3410 3412 3413 3414 3415 3416 3419 3420 3422 3423 3425 3426 3427 3429 3433 3435 3439 3440 3441 3442 3445 3446 3448 3449 3450 3452 3453 3455 3456 3459 3463 3466 3476 3482 3485 3486 3487 3490 3491 3493 3494 3495 3498 3511 3514 3516 3520 3521 3523 3524 3530 3531 3532 3533 3536
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- MOLLI MM1000Stryker Corporation · Marker, Radiographic, Implantable1 recall0 reports
- MOLLI MS-00009Stryker Corporation · Marker, Radiographic, Implantable1 recall0 reports
- MOLLI MP2001Stryker Corporation · Marker, Radiographic, Implantable1 recall0 reports
- MOLLI MT2001Stryker Corporation · Marker, Radiographic, Implantable1 recall0 reports
- MOLLI MP2015AStryker Corporation · Marker, Radiographic, Implantable1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page