Brand Name: AESCULAP recallZ-1485-2026 · Aesculap Inc · initiated Jan 15, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aesculap Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is the potential for the length of the trocar shaft to be too long.
FDA root cause category: Under Investigation by firm
What the company is doing
on January 28, 2026 URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Review the attachment Aesculap AG Urgent Field Safety Notice in its entirety. - If it is determined that a portion of the visual field is obscured, the manufacturer recommends rotating the device to compensate for the restricted view (refer to the attached Urgent Field Safety Notice from the manufacturer). Note: Devices are not being removed from the market, this is a field notification. - Return the completed Urgent Medical Device Correction Acknowledgement Form to Aesculap, Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt. Action planned by Aesculap AG to correct the problem: To address this issue, the manufacturer is implementing additional inspections of the optical system and introducing clearly defined acceptance criteria, along with other corrective actions. Should there be any questions, comments, or concerns regarding the product you have received, please feel free to contact (844-903-6417) or [email protected]
Recall details
- Recall number
- Z-1485-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 15, 2026
- Posted by the FDA
- Mar 2, 2026
- Recalling firm
- Aesculap Inc
- Quantity
- 126 units
- Product code
- GWG
- FDA submissions
- K983365
- Distribution
- US Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA, HI, IA, ID, IL, KS, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, WA.
Product description
Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Mode No. FF399R; UDI-DI 04038653065547; Serial Numbers: 4595, 4619, 4657, 4764, 4592, 4596, 4598, 4601, 4606, 4607, 4608, 4612, 4616, 4618, 4620, 4621, 4623, 4659, 4661, 4662, 4667, 4670, 4673, 4674, 4688, 4689, 4746, 4748, 4757, 4758, 4760, 4761, 4762, 4767, 4768, 4771, 4773, 4774, 4779, 4780, 4784, 4786, 4789, 4791, 4792, 4795, 4799, 4811, 4812, 4818, 4819, 4821, 4822, 4824, 4825, 4826, 4827, 4836, 4838, 4841, 4842, 4843, 4854, 4856, 4858, 4864, 4868, 4869, 4880, 4882, 4886, 4890, 4891, 4892, 4893, 4899, 4901, 4902, 4907, 4914, 4921, 4927, 4940, 4951, 4959, 4967, 4968, 4970, 4971, 4972, 4973, 4974, 4975, 4976, 4990, 4994, 4995, 5000, 5004, 5015, 5024, 5035, 5036, 5042, 5057, 5058, 5059, 5062, 5063, 5064, 5072, 5076, 5081, 5084, 5106, 5112, 5115, 5117, 5118, 5119, 5121, 5123, 5165, 5166, 5168, 5172;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AESCULAP FF399RAesculap AG · ENDOSCOPE, NEUROLOGICAL2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page