Medline Convenience Kits recallZ-1433-2026 · Medline Industries, LP · initiated Jan 5, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 1/6/26 was sent to customers regarding Saline Wound Wash. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-267 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/5/26 was sent to customers regarding Meijer Saline Wound Wash item number 347792. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Your account will receive credit once the response form is submitted, if applicable. 4. If you are a distributor, or have resold or transferred this prod

Recall details

Recall number
Z-1433-2026
Status
Open, Classified
Initiated by the company
Jan 5, 2026
Posted by the FDA
Feb 20, 2026
Recalling firm
Medline Industries, LP
Quantity
10550 units (495 additional units 3/27/26)
Product code
LRO
FDA submissions
K112666
Distribution
Worldwide distribution - US Nationwide and the countries of Bahamas, Bermuda, Cayman Islands.

Product description

Medline Convenience Kits: 1) TRUNK KIT W/EXPIRATION, Model Number: DYKM1361A; 2) TRUNK KIT 1EA, Model Number: DYKM2013; 3) TRUNK KIT 1EA, Model Number: DYKM2013A; 4) RN TRUNK KIT, Model Number: DYKM2699; 5) TRUNK KIT, Model Number: DYKTRUNK1

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

1) DYKM1361A, UDI-DI: 10193489586862(each), 40193489586863(case), Lot Number: 25JBG340; 2) DYKM1361A, UDI-DI: 10193489586862(each), 40193489586863(case), Lot Number: 25EBV829; 3) DYKM1361A, UDI-DI: 10193489586862(each), 40193489586863(case), Lot Number: 25CBM029; 4) DYKM1361A, UDI-DI: 10193489586862(each), 40193489586863(case), Lot Number: 25BBI797; 5) DYKM1361A, UDI-DI: 10193489586862(each), 40193489586863(case), Lot Number: 25ABW687; 6) DYKM2013, UDI-DI: 10193489796582(each), 40193489796583(case), Lot Number: 23EBN045; 7) DYKM2013, UDI-DI: 10193489796582(each), 40193489796583(case), Lot Number: 23DBM064; 8) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 25HBB845; 9) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 25GBS259; 10) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 25DBR145; 11) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 25BBS769; 12) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 24KBJ466; 13) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 24JBQ815; 14) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 24IBU526; 15) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 23IBJ540; 16) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 23HBP844; 17) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 23HBE308; 18) DYKM2013A, UDI-DI: 10195327419868(each), 40195327419869(case), Lot Number: 23GBT102; 19) DYKM2699, UDI-DI: 10195327610654(each), 40195327610655(case), Lot Number: 25DLA344; 20) DYKM2699, UDI-DI: 10195327610654(each), 40195327610655(case), Lot Number: 25CLA657; 21) DYKM2699, UDI-DI: 10195327610654(each), 40195327610655(case), Lot Number: 24GLB208; 22) DYKM2699, UDI-DI: 10195327610654(each), 40195327610655(case), Lot Number: 24ELA332; 23) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25HBS451; 24) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25HBS450; 25) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25HBE908; 26) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25HBE909; 27) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25GBM037; 28) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25GBK565; 29) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25FBL177; 30) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25EBP727; 31) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25EBO749; 32) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25DBQ021; 33) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25DBD942; 34) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25DBA345; 35) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25BBU730; 36) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25BBS503; 37) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25ABM247; 38) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24KBN725; 39) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24KBE264; 40) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24IBH956; 41) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24IBH955; 42) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24IBH957; 43) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24HBQ854; 44) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24HBO200; 45) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24GBN624; 46) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24FBH929; 47) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24FBG790; 48) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24FBD096; 49) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24DBS507; 50) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24DBC075; 51) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24CBG460; 52) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 24ABO402; 53) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23LBU172; 54) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23KBK857; 55) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23KBJ794; 56) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23JBM608; 57) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23IBQ747; 58) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23IBO843; 59) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23IBO844; 60) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23HBT781; 61) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23FBK991; 62) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23FBC676; 63) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23EBG472; 64) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23CBW700; 65) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 23CBM546; Additional Lot numbers as of 3/27/26: 1) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25IBL812; 2) DYKTRUNK1, UDI-DI: 10888277572195(each), 40888277572196(case), Lot Number: 25IBH378

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K112666 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98319).