Skip to content

PROBEAT recallZ-1430-2026 · Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha · initiated Feb 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software anomaly in the patient positioning system may result in positional discrepancy.

FDA root cause category: Software design

What the company is doing

Between the dates of February 5-11, 2026, Hitachi, Ltd, issued an "URGENT: MEDICAL DEVICE RECALL" notification dated January 21, 2026, via Service Team Manager to affected consignees. Hitachi is asking consignees to take the following actions: 1. Until the PPS software has been updated with the corrected version, users are instructed to follow the System Operation instruction in the document. 2. Users are also asked to post the document in a prominent space and sign its cover. 3. Complete and return the acknowledgment and receipt form. Action by the firm: Hitachi plans to update the PPS software to a corrected version that prevents the mismatch between the commanded error correction parameter and the internal variable. starting in late March 2026 and the users will be contacted after the update. If you have any questions, please call QA Manager at Tel: (408) 986-6300. Monday through Friday, 8:00 AM to 5:00 PM, Central Time.

Recall details

Recall number
Z-1430-2026
Status
Open, Classified
Initiated by the company
Feb 6, 2026
Posted by the FDA
Feb 20, 2026
Recalling firm
Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha
Quantity
1 unit
Product code
LHN
FDA submissions
K191801, K201042
Distribution
US distribution to states of: TX, Washington DC

Product description

Brand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CR

Code information (models, lots, serial numbers)

Serial Number: CRUS01/ UDI: (01)04560333350006(21)CRUS01(11)190920

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98330).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page