Skip to content

Medline Kits containing Tego Connectors recallZ-1463-2026 · Medline Industries, LP · initiated Jan 8, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

FDA root cause category: Process control

What the company is doing

On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.

Recall details

Recall number
Z-1463-2026
Status
Open, Classified
Initiated by the company
Jan 8, 2026
Posted by the FDA
Feb 24, 2026
Recalling firm
Medline Industries, LP
Quantity
516
Product code
LRO
Distribution
US Nationwide distribution.

Product description

Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200

Code information (models, lots, serial numbers)

IR PACK UDI-DI (ea) 10195327621780 UDI-DI (case) 40195327621781 Kit lots 24BBP627 24DBO029 24FBI148 24JBA858 24KBD925 24LBG978 25ABQ002 25EBV162; NO CATHETER DIALYSIS INSERTION UDI-DI (ea) 10195327675219 UDI-DI (case) 40195327675210 Kit Lots 24EBM915 24GBC287

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98333).

Sources and updates