Brand Name: STA Liatest D-Di recallZ-1585-2026 · Diagnostica Stago, Inc. · initiated Feb 5, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
FDA root cause category: Process change control
What the company is doing
On February 5, 2026 QUALITY ALERT letters were emailed to customers. Actions required: 1. As the intended purpose of the kit is to exclude PE and DVT, and its sensitivity and negative predictive value remain unchanged, it is appropriate to continue utilizing the reagents. 2. Please complete, sign and return the attached Acknowledgement Form by email, as directed on the form. In the event that a patient experiences an adverse event with any medical device, you may report it to FDA; see the guidance document Medical Device Reporting for User Facilities, which is available on the FDA.gov website, for instructions. For additional information, please contact Stago s 24/7 Hotline at 1-800-725-0607.
Recall details
- Recall number
- Z-1585-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 5, 2026
- Posted by the FDA
- Mar 19, 2026
- Recalling firm
- Diagnostica Stago, Inc.
- Quantity
- 34000 units (28446 US, 5554 OUS)
- Product code
- DAP
- FDA submissions
- K162227
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Singapore, Guam, Japan, South Korea, and Russia.
Product description
Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: in vitro diagnostic. The STA - Liatest D-Di kit is an assay for the quantitative determination of D-dimer in venous plasma for use on STA-R, STA Compact and STA Satellite analyzers by professional laboratory personnel.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model/Catalog Number: REF 00515; UDIs: (1) (01)03607450005158(11)240630(17)250930(10)273015(241)00515, (2) (01)03607450005158(11)250325(17)260630(10)273127(241)00515, (3)(01)03607450005158(11)250416(17)260731(10)273275(241)00515, (4)(01)03607450005158(11)250527(17)260831(10)273567(241)00515, (5)(01)03607450005158(11)250618(17)260930(10)273681(241)00515, (6)(01)03607450005158(11)250716(17)261031(10)273808(241)00515, (7)(01)03607450005158(11)250716(17)261231(10)273808(241)00515, (8)(01)03607450005158(11)250909(17)261231(10)274108(241)00515 (9) (01)03607450005158(11)250326(17)260630(10)273128(241)00515, (10) (01)03607450005158(11)250429(17)260731(10)273606(241)00515, (11) (01)03607450005158(11)250709(17)261031(10)273805(241)00515, (12) (01)03607450005158(11)251007(17)270131(10)274219(241)00515; Lot numbers: (1)273015, (2)273127, (3)273275, (4)273567, (5)273681, (6)273808, (7)273995, (8)274108, (9)273128, (10)273606, (11)273805, (12)274219;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- STA® - Liatest® D-Di 00515Diagnostica Stago · FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page