Skip to content

IBA Proton Therapy System - PROTEUS 235 recallZ-1497-2026 · Ion Beam Applications S.A. · initiated Feb 3, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ion Beam Applications S.A. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

FDA root cause category: Software design

What the company is doing

On February 3, 2026 Urgent Medical Device Correction letters were sent to customers. Letter provides awareness of the Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. On-site modification of the devices will be performed by IBA with a TSS Configuration update. The update will be deployed at the impacted sites by June 2026.

Recall details

Recall number
Z-1497-2026
Status
Open, Classified
Initiated by the company
Feb 3, 2026
Posted by the FDA
Mar 6, 2026
Recalling firm
Ion Beam Applications S.A.
Quantity
4 units (1 US, 3 OUS)
Product code
LHN
FDA submissions
K163500
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.

Product description

IBA Proton Therapy System - PROTEUS 235

Code information (models, lots, serial numbers)

PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates