SUCTION VALVE recallZ-1610-2026 · Olympus Corporation of the Americas · initiated Feb 12, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

FDA root cause category: Reprocessing Controls

What the company is doing

Olympus notified consignees on about 02/12/2026 via URGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed to immediately cease reprocessing the MAJ-1443 and MAJ-1444 valves in an OER-Pro or OER-Elite and ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the full extent of the MAJ-1443 and MAJ-1444 reprocessing steps as detailed in the provided Instructions for Use (IFU). The updated IFU can also be accessed electronically through OlympusConnect.com. Additionally, consignees were instructed to replace any existing copies of the MAJ-1443/MAJ-1444 IFU with the updated IFU provided in this letter, identify and notify all customers if affected units were further distributed and complete and confirm receipt of the notification via the Olympus web portal.

Recall details

Recall number
Z-1610-2026
Status
Open, Classified
Initiated by the company
Feb 12, 2026
Posted by the FDA
Mar 20, 2026
Recalling firm
Olympus Corporation of the Americas
Quantity
47,383 units
Product code
ODC
FDA submissions
K250949
Distribution
US Nationwide distribution.

Product description

Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.

Code information (models, lots, serial numbers)

Model Number: MAJ-1443. UDI-DI: 04953170355912. Catalog Number: N5363430. All Lot/Serial Numbers.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K250949 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98371).