EndoWrist;daVinci Si recallZ-1699-2026 · Intuitive Surgical, Inc. · initiated Dec 9, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to increased complaints for broken/frayed grip cables for reusable instruments.
FDA root cause category: Process control
What the company is doing
On 12/29/2025, the firm sent via email an "Urgent: Medical Device Recall" Letter informing customers that intuitive has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci S and Si reusable instruments. Customers are instruct to: 1. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 2. Identify and quarantine any affected product(s). Affected product(s) listed above may be returned by sending an email with quantities and lot numbers to your regional customer service. 3. Ensure to include the FSCA number ISIFA2025-15-R in the return notes. 4. Credit will be provided based on the number of remaining lives. 5. Shared or further distributed these products with other sites, make sure appropriate staff at the site receive and understand this notification so they locate and return their affected product. For questions or concern, contact Clinical Sales Representative or contact Intuitive Customer Service at the numbers listed below: " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Recall details
- Recall number
- Z-1699-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 9, 2025
- Posted by the FDA
- Mar 30, 2026
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 2660
- Product code
- NAY
- FDA submissions
- K050369
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product description
Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Lot Code: 420278: Grasping Retractor GTIN: 00886874111772 FDA Medical Device Listing Number: D088920 Material Number: 420278-04 Batch / System Numbers: M10110613 SH1182 M10120127 SH1182 M10120217 SH1182 M10120829 SH1182 M10130103 SH1182 M10130114 SH1182 M11130404 SH1182 M10130527 SH1182 M10130606 SH1182 M10130909 SH1182 M10130925 SH1182 S10101229 SG704 S10110105 SG704 M10121214 SH0547 M10110118 SH0059 M10110613 SH0059 S10110819 SH0059 M10121122 SH0059 M10121214 SH0059 M10130114 SH0059 M10130220 SH0059 M10130319 SH0059 M11130404 SH0059 M10120912 SH1390 M10130730 USG721 M10130909 USG721 M10110307 SH0729 M10130214 SH0729 M11130404 SH0729 M10130617 SH0729 M10130628 SH0729 M10120127 SH1076 M10130307 SH1279 M10111028 SH0761 Material Number: 420278-06 Batch / System Numbers: M10140115 SH1182 N10140725 SH1182 N10140923 SH1182 N10150324 SH1182 N10150417 SH1182 N10150721 SH1182 N10150608 SH1182 N10150820 SH1182 N10151113 SH1182 N10151214 SH1182 N10151029 SH1182 N10160304 SH1182 N10151201 SH1182 N10141205 SH1182 N10160921 SH1182 N10161123 SH1182 N10170516 SH1182 N10170601 SH1182 N10170921 SH1182 N10171213 SH1182 N10171226 SH1182 N10180417 SH1182 N10140815 SG704 N10150210 SG704 N10150608 SG704 N10160203 SG704 N10170117 SG704 N10170921 SG704 N10170822 SG704 N10171002 SG704 N10171226 SG704 M10140306 SH0059 N10141205 SH0059 N10150605 SH0059 N10170213 SH0059 N10171213 SH0059 N10171117 SH1964 N10180417 SH1964 N10170310 USG294 M10140306 USG721 M10140319 USG721 N10170412 USG721 N10180117 USG721 N10150224 SH1378 N10160203 SH1378 M10140306 SH0729 N10140828 SH0729 N10150330 SH0729 N10150319 SH0729 N10150525 SH0729 N10151020 SH0729 N10160503 SH0729 N10160524 SH0729 N10180417 SH0729 N10141001 SH1076 N10141030 SH1076 N10151211 SH1604 N10160524 SH1604 N10160801 SH1604 N10160921 SH1604 N10171226 SH1604 N10160921 SH1385 N10161010 SH1385 N10161114 SH1385 N10171117 SH1385 Material Number: 420278-07 Batch / System Numbers: N10181219 SH1182 N10180906 SH1182 N10190423 SH1182 N10181114 SH1182 N10180906 SG704 N10180910 USG721 N10181114 SH1604 Material Number: 420278-09 Batch / System Numbers: N10190826 SH1182 N10190515 SH1182 N10190909 SH1182 N10190902 SH1964 N10190826 SH1964 N10190515 SH1964 N10190909 SH1964 N10190826 USG721 N10190902 SH1604
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EndoWrist;daVinci Si 420278Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall8 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98383).
- Z-1696-2026 · Open, ClassifiedBrand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: …
- Z-1697-2026 · Open, ClassifiedBrand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: N…
- Z-1698-2026 · Open, ClassifiedBrand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Descri…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.