TECNIS SIMPLICITY recallZ-1634-2026 · AMO Puerto Rico Manufacturing, Inc. · initiated Feb 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AMO Puerto Rico Manufacturing, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
FDA root cause category: Process control
What the company is doing
On 02/19/2026, the firm send via FedEx an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that in a limited quantity of TECNIS Eyhance IOL (intraocular lens) with TECNIS Simplicity" Delivery System (DIB00) products exhibit an occurrence of haptics sticking to the optic, which can prevent the lens from unfolding as expected following lens insertion during the surgical procedure. Customers are instructed to: 1. Identify if any inventory contains TECNIS Eyhance IOL (intraocular lens) with TECNIS Simplicity" Delivery System (DIB00). 2. Do not use and remove all affected TECNIS Eyhance IOL (intraocular lens) with TECNIS Simplicity" Delivery System (DIB00) from inventory. 3. Complete Customer Reply Form. It is required information for reconciliation purposes with regulatory agencies, even if you have no inventory. For products to be returned: " Complete the Customer Reply Form, noting the serial number of the TECNIS Eyhance IOL (intraocular lens) with TECNIS Simplicity" Delivery System (DIB00). " Use the provided return label or contact the Global Post Market Safety Team at 1-877-266-4543, Option 2 to obtain an RGA number and arrange product return. " Email Customer Reply Form to [email protected] within 3 business days of receipt of this letter. " If issues were encountered, inform JJSV by calling the Global Post Market Safety Team 1-877-266-4543, Option 2 or send an email to [email protected] . If reporting a complaint, provide the serial number, the date of surgery, a description of the event, and patient outcome. " Return affected product as soon as possible. The firm will work directly with customers to replace the affected unit and provide any necessary next steps. If you do not have product to be returned: " Review your patient records to confirm implantation of the impacted lens. No further action is required after the Customer Reply Form has been returned.
Recall details
- Recall number
- Z-1634-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 19, 2026
- Posted by the FDA
- Mar 24, 2026
- Recalling firm
- AMO Puerto Rico Manufacturing, Inc.
- Quantity
- 361
- Product code
- HQL
- FDA submissions
- P980040, P980040S117
- Distribution
- U.S Nationwide distribution in the states of CA,FL, IL, ME,MS, NJ, NY, OK, OR, TX, and WI.
Product description
Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No
Code information (models, lots, serial numbers)
Lot Code: GTIN: 05050474731776 SKU: DIB00U0210-12 Serial Numbers: 2409342403 2410072403 2409612403 2409692403 2409752403 2891682403 2891132403 2892032403 2531882403 2531902403 2531212403 2530842403 2531242403 2531892403 2531622403
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page