Skip to content

Aptima recallZ-1586-2026 · Hologic, Inc. · initiated Feb 18, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to product exhibiting potential to generate either invalid or false negative results.

FDA root cause category: Under Investigation by firm

What the company is doing

On 02/18/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter to customers informing them that Hologic has identified that the Aptima HPV Assay kits from ML 929785 exhibit the potential to generate either invalid results for samples and calibrators or false negative results for low analyte samples when reagents are not used on the day of reconstitution. Customers are instructed: 1. Discontinue use of the Aptima HPV Assay kits from ML 929785. 2. Segregate and count the number of affected product on-hand and dispose according t the local guidelines and regulations. 3.Fill out the Customer Response Form using the link https://iqvia-response.my.site.com/mt/fca?cid=hpv-929785 4. Hologic will process and ship the replacement kits For Questions or concerns, contact Hologic Technical Support at +1 888-484-4747 or +1 858-410-8511 during hours 5 am to 5 pm PT or by email at [email protected]

Recall details

Recall number
Z-1586-2026
Status
Open, Classified
Initiated by the company
Feb 18, 2026
Posted by the FDA
Mar 19, 2026
Recalling firm
Hologic, Inc.
Quantity
306
Product code
OYB
FDA submissions
P100042
Distribution
U.S. Nationwide distribution in the states of CA, FL. IA, IL, MO, NC, NM, PR, and TN.

Product description

Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No

Code information (models, lots, serial numbers)

Lot Code: The Aptima HPV Assay Kit Part Number 303585, Master Lot 929785 contains one Refrigerated Box (PN 302883) and one Room Temperature Box (PN 302885). The following are the UDI number for each box: UDI Number for PN 302883: 15420045500051 UDI Number for PN 302885: 15420045500068

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page