Olympus recallZ-1639-2026 · Olympus Corporation of the Americas · initiated Feb 25, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this letter through the Olymps portal 5. Please forward this notice to other users who may have the affected products if you have further distributed it.
Recall details
- Recall number
- Z-1639-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 25, 2026
- Posted by the FDA
- Mar 25, 2026
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 164 units
- Product code
- GEI
- FDA submissions
- K023492
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Product description
Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Code information (models, lots, serial numbers)
Model: 3006; UDI: 00821925035898; Lot Numbers: All unexpired;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- OLYMPUS 3006Gyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 recalls16 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K023492 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- OLYMPUS 3005Gyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 recalls6 reports
- PK TECHNOLOGY 920005PKGyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall9 reports
- PK TECHNOLOGY 3005PKGyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall2 reports
- GYRUS ACMI, INC 920000PKGyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall0 reports
- PK TECHNOLOGY 3000PKGyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98454).
- Z-1638-2026 · Open, ClassifiedBrand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bi…
- Z-1640-2026 · Open, ClassifiedBrand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate a…
- Z-1641-2026 · Open, ClassifiedBrand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are i…
- Z-1642-2026 · Open, ClassifiedBrand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be pas…
- Z-1643-2026 · Open, ClassifiedBrand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended …
- Z-1644-2026 · Open, ClassifiedBrand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be p…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page