Skip to content

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number:… recallZ-1703-2026 · Synthes (USA) Products LLC · initiated Feb 25, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

FDA root cause category: Process change control

What the company is doing

Depuy Synthes initially notified Sales Consultants and Regional Managers on about 02/25/2026, followed by notification of consignees. Consignees were instructed to examine inventory for affected units, quarantine them immediately, and contact Depuy Synthes sales consultant or customer support services to arrange the return/credits for affected units. Additionally, consignees were requested to complete and return the provided Business Response Form as well as notify all applicable personnel within the facility or any consignees if the affected units were further distributed. Consignees were also requested to post the notification in a visible area for awareness.

Recall details

Recall number
Z-1703-2026
Status
Open, Classified
Initiated by the company
Feb 25, 2026
Posted by the FDA
Mar 30, 2026
Recalling firm
Synthes (USA) Products LLC
Quantity
33 units
Product code
HWX
Distribution
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.

Product description

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Code information (models, lots, serial numbers)

Part Number: 03.424.126-US. GTIN: 10886982335951. Lot Number: 68982P8.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98461).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page