Skip to content

CUSA Clarity recallZ-1828-2026 · Integra LifeSciences Corp. (NeuroSciences) · initiated Mar 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. (NeuroSciences) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue that renders the touch screen unresponsive.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice dated 3/6/26 was sent to consignees. The notice instructs consignees to restart systems when freeze to restore functionality of the touchscreen capabilities for 8 minutes; this action may be used multiple times. Integra's Service and Repair can provide remote support to disable and re-enable the write filter with Integra employees and/or Authorized service personnel. This will reset the console to provide approximately 28 days with no touchscreen issue. Integra Lifesciences Service Team will contact consignees within 5 to 10 business days to schedule correction of affected devices at no cost. Consignees with any questions can direct them to Customer Service by emailing [email protected] or by phone at 1-800-654-2873.

Recall details

Recall number
Z-1828-2026
Status
Open, Classified
Initiated by the company
Mar 6, 2026
Posted by the FDA
Apr 17, 2026
Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Quantity
78 units
Product code
LFL
FDA submissions
K161882
Distribution
Worldwide - US Nationwide distribution in the states of MI & NY. The countries of Belgium, China, Czechia, Dubai, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands. Poland, Spain, Switzerland, Tunisia & United Kingdom.

Product description

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model: C7000; UDI: 10381780126232; Serial Numbers: CHA1902101IE, CHA2001301IE, CHA2001401IE, CHA2002301IE, CHA2202301IE, CHA2305101IE, CHA2305401IE, CHA2307001IE, CHA2402201IE, CHB1711001IE, CHB1723601IE, CHB2000801IE, CHB2100901IE, CHB2301401IE, CHB2500101IE, CHC1800601IE, CHC1902601IE, CHC1902701IE, CHC2001201IE, CHD1705801IE, CHD1707101IE, CHD1801001IE, CHD1801101IE, CHD1902501IE , CHE1707201IE, CHE1707301IE, CHE2203301IE, CHE2204001IE, CHE2301001IE , CHE2305001IE, CHE2502901IE, CHE2503101IE, CHE2503201IE, CHE2503401IE, CHE2503701IE, CHF1701401IE, CHF1702701IE, CHF2100701IE, CHF2303701IE, CHF2305401IE, CHG1700101IE, CHG1700801IE, CHG1902401IE, CHG2000201IE, CHG2000301IE, CHG2204101IE, CHG2402301IE, CHG2502801IE, CHG2506001IE, CHH1802901IE, CHH2402801IE, CHH2500601IE, CHJ2001101IE, CHJ2001301IE , CHJ2301201IE, CHJ2304601IE, CHK1700801IE, CHK1702301IE, CHK1703001IE, CHK2402601IE, CHK2403401IE, CHK2404101IE, CHL1701701IE, CHL1900201IE, CHL2102901IE , CHL2103001IE, CHL2301301IE, CHL2302001IE, CHL2302101IE, CHL2403301IE, CHM1901901IE, CHM2100101IE, CHM2101601IE, CHM2200201IE, CHM2200301IE, CHM2200501IE, CHM2300701IE, CHM2402101IE.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.